NA ATLS02Recalls & reported problems
Medtronic, Inc. · ONE SOURCE PACK ATLS02 · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169108332), received between Nov 8, 2024 and Apr 23, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak5
- Crack1
- Device Alarm System1
- Material Puncture/Hole1
- Obstruction of Flow1
Patient problems
- Hemorrhage/Blood Loss/Bleeding3
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2025Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AUTOLOG ONE SOURCE PACK IN RIGHT HEPATECTOMY SURGERY, IT WAS REPORTED THAT AT THE TIME OF EQUIPMENT PURGE THE BOWL WAS FILLED WITH WATER OUTSIDE. THE USE OF THE DEVICE WAS UNKNOWN. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21887958 ↗MalfunctionMar 7, 2025Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT DURING USE OF THE AUTOLOG ONE SOURCE PACK, THE CENTRIFUGE CORE SUDDENLY RUPTURED, CAUSING MASSIVE BLOOD LOSS AND THREATENING THE PATIENT'S HEALTH. THE USE OF THE DEVICE WAS NOT SPECIFIED.NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21544637 ↗MalfunctionNov 28, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AUTOLOG ONE SOURCE PACK, IT WAS REPORTED THAT WATER LEAKED OUT OF THE BOWL AND DRAINED OUT FROM THE AUTOLOG MACHINE TO THE FLOOR. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE NOTED IN THE INSTRUMENT OR THE DISPOSABLE, AND THE ONLY VISUAL ABNORMALITIES OBSERVED WAS THE LEAKAGE. THE FILL LEVEL FOR THE RESERVOIR, HOLDING, SALINE, AND WASTE BAG AT THE TIME OF THE ISSUE WAS PRIMING STAGE, MIN VOLUME. NO ERROR WARNING MESSAGE APPEARED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20804083 ↗MalfunctionNov 26, 2024Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A ATLS02, IT WAS REPORTED THAT IT CRACKED DURING THE WASHING PROCESS. BLOOD/WATER LEAKED OUT OF THE BOWL AND DRAIN OUT FROM THE AUTOLOG MACHINE ONTO THE FLOOR. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WERE AUDIBLE ABNORMALITIES, STRANGE AND NOISY SOUND WHEN PROCESSING. THE LEAK O CCURRED DURING THE WASH CYCLE. THE BOWL WAS RE-SEATED TO THE SAME MACHINE, THE DEVICE WAS STILL LEAKING AND THE BOWL CRACKED. THE CUSTOMER CHANGED TO A NEW SET OF CONSUMABLES. THE LEVEL OF RESERVOIR WAS 1600ML, HOLDING 200ML, SALINE 2000ML AND WASTE BAG WAS 2000ML. NO ERROR WARNING MASSAGE APPEARED. PATIENT BLOOD LOSS AS A RESULT OF THIS LEAK WAS 300ML. NO TRANSFUSION WAS REQUIRED AS A RESULT OF THIS LEAK.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20788523 ↗InjuryNov 8, 2024Device Alarm System; Fluid/Blood Leak; Obstruction of Flow; Device Alarm System; Fluid/Blood Leak; Obstruction of Flow
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF AN AUTOLOG ONE SOURCE PACK, IT WAS REPORTED THAT THE CENTRIFUGE BOWL FAILED AND BLOOD SPILLED OUTSIDE THE BOWL. CUSTOMER COULD NOT COMPLETE THE CELL SALVAGE PROCEDURE. SEE ATTACHED PHOTOS. USE OF DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 20639231 ↗
FDA clearance and classification
Cleared through 510(k) K830735 on May 2, 1983 (decision: Substantially Equivalent). Submitted by Electromedics, Inc. as “Bt 795/A Autotrans”. The FDA review took 55 days. FDA record ↗
The FDA regulates product code CAC (apparatus, autotransfusion) under 21 CFR 868.5830 as a Class II device. 21 CFR 868.5830 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA ATLS02
Has the NA ATLS02 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00643169108332), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA ATLS02?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Nov 8, 2024 and Apr 23, 2025. In those reports, the most common specific device problems are “Fluid/Blood Leak” (5), “Crack” (1) and “Device Alarm System” (1); the most common patient problem is “Hemorrhage/Blood Loss/Bleeding” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA ATLS02?
None of the reports are classified as deaths, 1 as an injury and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA ATLS02 FDA approved?
It was cleared through 510(k) K830735 on May 2, 1983. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the NA ATLS02, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand NA, model ATLS02). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other APPARATUS, AUTOTRANSFUSION devices
- OASIS 3650-100Atrium Medical Corporation · APPARATUS, AUTOTRANSFUSION2 recalls22 reports
- EXPRESS 4050-100NAtrium Medical Corporation · APPARATUS, AUTOTRANSFUSION2 recalls2 reports
- NA ATL2001Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 death79 reports
- autoLog IQ™ ATLGIQMedtronic, Inc. · APPARATUS, AUTOTRANSFUSION57 reports
- NA BTC93Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 recall2 reports
- OCEAN 2050-000Atrium Medical Corporation · Apparatus, Autotransfusion1 recall2 reports
- NA BTC98Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 recall0 reports
- HAEMONETICS CELL SAVER ELITE SYSTEM CSE-P-125Haemonetics Corporation · Apparatus, Autotransfusion47 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page