NA ATLS02Recalls & reported problems
Medtronic, Inc. · ONE SOURCE PACK ATLS02 · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169108332), received between Nov 8, 2024 and Apr 23, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak5
- Crack1
- Device Alarm System1
- Material Puncture/Hole1
- Obstruction of Flow1
Patient problems
- Hemorrhage/Blood Loss/Bleeding3
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2025Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AUTOLOG ONE SOURCE PACK IN RIGHT HEPATECTOMY SURGERY, IT WAS REPORTED THAT AT THE TIME OF EQUIPMENT PURGE THE BOWL WAS FILLED WITH WATER OUTSIDE. THE USE OF THE DEVICE WAS UNKNOWN. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21887958 ↗MalfunctionMar 7, 2025Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
IT WAS REPORTED THAT DURING USE OF THE AUTOLOG ONE SOURCE PACK, THE CENTRIFUGE CORE SUDDENLY RUPTURED, CAUSING MASSIVE BLOOD LOSS AND THREATENING THE PATIENT'S HEALTH. THE USE OF THE DEVICE WAS NOT SPECIFIED.NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 21544637 ↗MalfunctionNov 28, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AUTOLOG ONE SOURCE PACK, IT WAS REPORTED THAT WATER LEAKED OUT OF THE BOWL AND DRAINED OUT FROM THE AUTOLOG MACHINE TO THE FLOOR. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE NOTED IN THE INSTRUMENT OR THE DISPOSABLE, AND THE ONLY VISUAL ABNORMALITIES OBSERVED WAS THE LEAKAGE. THE FILL LEVEL FOR THE RESERVOIR, HOLDING, SALINE, AND WASTE BAG AT THE TIME OF THE ISSUE WAS PRIMING STAGE, MIN VOLUME. NO ERROR WARNING MESSAGE APPEARED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20804083 ↗MalfunctionNov 26, 2024Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A ATLS02, IT WAS REPORTED THAT IT CRACKED DURING THE WASHING PROCESS. BLOOD/WATER LEAKED OUT OF THE BOWL AND DRAIN OUT FROM THE AUTOLOG MACHINE ONTO THE FLOOR. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WERE AUDIBLE ABNORMALITIES, STRANGE AND NOISY SOUND WHEN PROCESSING. THE LEAK O CCURRED DURING THE WASH CYCLE. THE BOWL WAS RE-SEATED TO THE SAME MACHINE, THE DEVICE WAS STILL LEAKING AND THE BOWL CRACKED. THE CUSTOMER CHANGED TO A NEW SET OF CONSUMABLES. THE LEVEL OF RESERVOIR WAS 1600ML, HOLDING 200ML, SALINE 2000ML AND WASTE BAG WAS 2000ML. NO ERROR WARNING MASSAGE APPEARED. PATIENT BLOOD LOSS AS A RESULT OF THIS LEAK WAS 300ML. NO TRANSFUSION WAS REQUIRED AS A RESULT OF THIS LEAK.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20788523 ↗InjuryNov 8, 2024Device Alarm System; Fluid/Blood Leak; Obstruction of Flow; Device Alarm System; Fluid/Blood Leak; Obstruction of Flow
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF AN AUTOLOG ONE SOURCE PACK, IT WAS REPORTED THAT THE CENTRIFUGE BOWL FAILED AND BLOOD SPILLED OUTSIDE THE BOWL. CUSTOMER COULD NOT COMPLETE THE CELL SALVAGE PROCEDURE. SEE ATTACHED PHOTOS. USE OF DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 20639231 ↗
Other APPARATUS, AUTOTRANSFUSION devices
- OASIS 3650-100Atrium Medical Corporation · APPARATUS, AUTOTRANSFUSION2 recalls22 reports
- EXPRESS 4050-100NAtrium Medical Corporation · APPARATUS, AUTOTRANSFUSION2 recalls2 reports
- NA ATL2001Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 death79 reports
- autoLog IQ™ ATLGIQMedtronic, Inc. · APPARATUS, AUTOTRANSFUSION58 reports
- NA BTC93Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 recall2 reports
- OCEAN 2050-000Atrium Medical Corporation · Apparatus, Autotransfusion1 recall2 reports
- NA BTC98Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 recall0 reports
- HAEMONETICS CELL SAVER ELITE SYSTEM CSE-P-125Haemonetics Corporation · Apparatus, Autotransfusion47 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.