Evera MRI™ XT DR SureScan® DDMB1D4 (UDI 00643169314092)Recalls & reported problems

Medtronic, Inc. · ICD DDMB1D4 EVERA MRI DR XT DF4 US · Class 3 · Implant

All models of Evera MRI XT DR SureScan (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic Evera MRI XT DR is an implanted defibrillator that treats heart rhythm problems. The 943 reports mostly describe inappropriate shocks or pacing, sensing problems, early battery drain, and patients fainting or feeling shocks from the leads. The reports include 24 deaths and 639 injuries. There are 2 recalls, 1 open and 1 closed, about early replacement warnings and possibly blocked shock therapy. Reports fell to 35 over the last year from 69 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
945
Deaths reported
24
Injuries reported
640
Recalls naming it
2

Summary

The FDA's MAUDE database contains 945 adverse event reports that list this exact device (DI 00643169314092), received between Dec 9, 2016 and Sep 29, 2026 — classified as 24 deaths, 640 injuries, 281 malfunctions.

The most frequently reported specific device problem is “Inappropriate/Inadequate Shock/Stimulation” (284 reports).

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 12 reportsDec 2023: 7 reportsJan 2024: 9 reportsFeb 2024: 6 reportsMar 2024: 8 reportsApr 2024: 13 reportsMay 2024: 8 reportsJun 2024: 11 reportsJul 2024: 7 reportsAug 2024: 11 reportsSep 2024: 6 reportsOct 2024: 8 reportsNov 2024: 8 reportsDec 2024: 2 reportsJan 2025: 11 reportsFeb 2025: 5 reportsMar 2025: 6 reportsApr 2025: 8 reportsMay 2025: 5 reportsJun 2025: 6 reportsJul 2025: 7 reportsAug 2025: 3 reportsSep 2025: 3 reportsOct 2025: 5 reportsNov 2025: 4 reportsDec 2025: 1 reportsJan 2026: 4 reportsFeb 2026: 7 reportsMar 2026: 1 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 4 reportsJul 2026: 4 reportsAug 2026: 3 reportsSep 2026: 5 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 3, 2021Class IOpen, Classified

    A small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation

    Z-1305-2021 · Root cause: Device Design · Recall details · FDA record ↗

  2. Dec 19, 2017Class ITerminated

    Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).

    Z-0585-2018 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inappropriate/Inadequate Shock/Stimulation284
  • Device Sensing Problem255
  • Adverse Event Without Identified Device or Use Problem178
  • Pacing Problem120
  • Premature Discharge of Battery102
  • Electrical /Electronic Property Problem72
  • Over-Sensing52
  • Invalid Sensing51

Patient problems

  • Shock from Patient Lead(s)285
  • No Clinical Signs, Symptoms or Conditions171
  • No Known Impact Or Consequence To Patient155
  • Unspecified Infection114
  • No Patient Involvement57
  • Bradycardia26
  • Pain17
  • Tachycardia17

Recent adverse event reports

The 25 most recent of 945 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    It was reported that the patient received possible inappropriate Anti-Tachycardia Pacing (ATP) and shocks for Supra Ventricular Tach ycardia (SVT) detected as Ventricular Fibrillation (VF). Additionally, the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events. The Implantable Cardioverter Defibrillator (ICD) and RA lead remain in use. No further patient complications have been reported as a result of this event.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50938082 ↗
  • InjurySep 14, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50763404 ↗
  • MalfunctionSep 10, 2026Electromagnetic Interference

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722933 ↗
  • InjurySep 9, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    It was reported that  the Implantable Cardioverter Defibrillator (ICD) delivered inappropriate Anti-Tachycardia Pacing (ATP) and shocked for Supra Ventricular Tachycardia (SVT)  which the device detected and treated as Ventricular Tachycardia (VT). The device remains in use. No further patient complications have been reported as a result of this event

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50689736 ↗
  • MalfunctionSep 1, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50543348 ↗
  • InjuryAug 25, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    It was reported that there was undersensing on the Right Ventricular (RV) lead and Far Field R-Wave (FFRW) oversensing on the Right Atrial (RA) lead. It was also noted the Implantable Cardioverter Defibrillator (ICD) possibly detected Supra Ventricular Tachycardia (SVT) as Ventricular Fibrillation (VF) and monitored Ventricular Tachycardia (VT). The VF detection resulted in Anti-Tachycardia Pacing (ATP) and high voltage therapy. The leads and device remain in use. No further patient complications have been reported as a result of this event.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 50431593 ↗
  • MalfunctionAug 20, 2026Battery Problem

    Patient problems: Confusion/ Disorientation

    MAUDE report 50357329 ↗
  • InjuryAug 4, 2026Failure to Deliver Shock/Stimulation

    It was reported by a nurse that the Implantable Cardioverter Defibrillator (ICD) failed to treat an episode of Ventricular Tachycard ia (VT). It was noted the patient was unresponsive during the episode and required external defibrillation. Upon technical review, it was noted that the ventricular rate during the episode fell into the monitor only zone, so it did not meeting rate zone criteria for therapy. The ICD remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 50132247 ↗
  • MalfunctionJul 27, 2026Premature Discharge of Battery

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50045509 ↗
  • InjuryJul 13, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED BY THE PATIENT THAT THEY HAD FAINTED AND REPORTED AND WERE ADMITTED TO THE HOSPITAL THAT HAD IMPLANTED THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DEVICE. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Syncope/Fainting

    MAUDE report 25798412 ↗
  • InjuryJul 8, 2026Failure to Deliver Shock/Stimulation

    IT WAS REPORTED BY THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) PATIENT THAT THEY HAD INSTANCES OF BLACKING OUT AND CHEST PAINS. THE PATIENT REPORTED BEING INFORMED BY THEIR DOCTOR THAT THE ICD SHOULD HAVE SHOCKED THEM. THE PATIENT WAS REPORTEDLY PRESCRIBED A HEART MONITOR WHERE "SKIPPING" WAS OBSERVED THREE TIMES. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Chest Pain; Syncope/Fainting

    MAUDE report 25749389 ↗
  • InjuryJul 6, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) MAY HAVE PROVIDED INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND HIGH VOLTAGE THERAPY FOR A SUPRA VENTRICULAR TACHYCARDIA (SVT) EPISODE DETECTED AS VENTRICULAR FIBRILLATION (VF). FOLLOWING THE ATP DELIVERY THE RATES WERE NOTED TO HAVE INCREASED IN ADDITION TO A MORPHOLOGY CHANGE, AND HIGH VOLTAGE THERAPY WAS APPROPRIATELY DELIVERED.  THE ICD REMAINS IN USE.  NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 25725233 ↗
  • InjuryJun 26, 2026Misconnection

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH, OUT-OF-RANGE, UNDEFINED PACING IMPEDANCE AND A LARGE NUMBER OF SHORT V-V INTERVALS. DURING THE REPLACEMENT OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DEVICE, THE PHYSICIAN OBSERVED TISSUE/DEBRIS WITHIN THE CONNECTOR PORT/HEADER AREA. AFTER CONNECTING THE RV LEAD TO A NEW ICD, IMPEDANCE, SENSING, AND PACING MEASUREMENTS NORMALIZED. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25637967 ↗
  • MalfunctionJun 18, 2026Pacing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED DIZZINESS AND SHORTNESS OF BREATH AND THE NURSE MENTIONED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) STARTED PACING THE PATIENT AFTER THE EMERGENCY MEDICAL SERVICES (EMS) TEAM WENT OVER A ROCK. ADDITIONALLY, RIGHT ATRIAL (RA) LEAD UNDERSENSING WAS OBSERVED RESULTING IN TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EPISODES IN PROGRESS AND UNNECESSARY ATRIAL PACING AT TIMES. A FAILED ATRIAL LEAD POSITION CHECK WAS OBSERVED AND IT WAS ALSO REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) MAY HAVE INAPPROPRIATELY CLASSIFIED ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EPISODES WITH FAST VENTRICULAR RESPONSE AS SUPRA VENTRICULAR TACHYCARDIA (SVT). THE ICD AND RA LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Dizziness; Dyspnea

    MAUDE report 25552590 ↗
  • InjuryJun 12, 2026Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED BY THE PATIENT THAT THERE WERE ISSUES WITH THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). THE PATIENT REPORTEDLY KEPT RECEIVING SHOCKS WHILE SLEEPING DURING THE NIGHT BUT THE SHOCKS WERE NOT SHOWING UP ON THE PATIENT'S REPORTS. THE PATIENT WENT IN TO THEIR CLINIC AND ADJUSTMENTS WERE MADE, MORE SPECIFICALLY THEY HAD SOMETHING TURNED DOWN; HOWEVER, THE SHOCKS CONTINUED. THE PATIENT BELIEVED SOMETHING WAS WRONG WITH THE DEVICE. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 25483616 ↗
  • MalfunctionJun 3, 2026Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED DETECTION ISSUE AS THE DEVICE DETECTED SUPRA VENTRICULAR TACHYCARDIA (SVT) AS VENTRICULAR TACHYCARDIA (VT). THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25382610 ↗
  • InjuryMay 12, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A WARNING FOR HIGH SUPERIOR VENA CAVA (SVC) DEFIBRILLATOR COIL IMPEDANCE. THE RV LEAD REMAINS IN USE. IT WAS ALSO REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE HIGH VOLTAGE THERAPY AND ANTI-TACHYCARDIA PACING (ATP) FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) VERSUS VENTRICULAR TACHYCARDIA (VT). THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 25150265 ↗
  • MalfunctionApr 21, 2026Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) POSSIBLY DETECTED FAST ATRIAL AND VENTRICULAR ARRHYTHMIA AS NO N-SUSTAINED VENTRICULAR TACHYCARDIA. THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24947968 ↗
  • InjuryApr 20, 2026Appropriate Device Problem Term/Code Not Available

    FOLLOW UP WITH THE CLINIC FOUND THAT THE PATIENT HAS NOT BEEN SEEN IN CLINIC FOR SEVERAL YEARS AND IS NOT SET UP WITH REMOTE MONITORING. THE CLINIC HAD NO FURTHER INFORMATION.

    Patient problems: Cardiac Arrest

    MAUDE report 24930745 ↗
  • MalfunctionApr 20, 2026Electromagnetic Interference

    IT WAS REPORTED THAT DURING USE OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) ELECTRICAL NOISE WAS NOTED, LIKELY FROM CONTACTING SOMETHING POORLY GROUNDED. THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24928120 ↗
  • InjuryMar 19, 2026Failure to Deliver Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT POSSIBLY EXPERIENCED VENTRICULAR TACHYCARDIA (VT) AS THE DEVICE DETECTED SUPRA VENTRICULAR TACHYCARDIA (SVT) INSTEAD. THE DEVICE WITHHELD THERAPY INAPPROPRIATELY. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Tachycardia

    MAUDE report 24641828 ↗
  • InjuryFeb 26, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) PATIENT THAT THEIR HEART RATE HAD BEEN RACING AND THEN SLOWING DOWN AND THEY HAD BAD ATRIAL FIBRILLATION (AF). THE PATIENT REPORTED BEING HOSPITALIZED ON AND OFF FOR THE LAST THREE MONTHS. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Atrial Fibrillation

    MAUDE report 24455359 ↗
  • InjuryFeb 24, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) PATIENT THAT THE EXPERIENCED AN ENDOCARDITIS INFECTION. THE PATIENT WAS HOSPITALIZED AND THE DEVICE SYSTEM WAS EXPLANTED AND THE PATIENT WAS GIVEN AN EXTERNAL DEFIBRILLATOR VEST. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Endocarditis

    MAUDE report 24432795 ↗
  • InjuryFeb 24, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED A STROKE AND WAS "NOT DOING WELL". THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Stroke/CVA

    MAUDE report 24429688 ↗
  • InjuryFeb 17, 2026Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO AN UNKNOWN REASON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Endocarditis; No Clinical Signs, Symptoms or Conditions; Unspecified Infection

    MAUDE report 24365850 ↗

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →