Implantable Cardioverter Defibrillators (ICDs), Evera MRI.… recallZ-0585-2018 · Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) · initiated Dec 19, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).

FDA root cause category: Process design

What the company is doing

On 12/19/2017, Medtronic verbally notified consignees and retrieved eight devices at sites that had affected inventory on hand. On 01/22/2018, Medtronic provided a recall notification letter to physicians following 48 patients implanted with affected devices and risk managers of those medical facilities, recommending that physicians strongly consider prophylactic device replacement for patients implanted with an affected device. Medtronic Technical Services is available to assist physicians with questions at 800-723-4636. Medtronic Patient Service is available to assist patients at 800-551-5544 (Monday-Friday, 8am-5pm Central Time). The recall was expanded on 03/15/2018 and customers were notified by letter. Affected product, not yet implanted, was retrieved on about 03/08/2018. Instructions in the physicians following patients implanted with the additional affected devices included prophylactic device replacement for those patients at higher risk, including patients whose clinical history indicates prior need for high-voltage therapy and/or for pacemaker-dependent patients. The physicians should carefully weigh the risks and benefits of device replacement. If it is determined that replacement is not warranted, instructions included the following: programming changes to reduce the potential for high-voltage charges associated with arrhythmia detection and therapies, such as enabling ATP before charging for fast ventricular rhythms or programming a separate fast VT via VF zone with ATP; Continue three-month in-clinic or remote follow-ups to verify device functionality because inability to interrogate a device or a failed remote monitoring transmission may be an indication that internal arcing has occurred; and to advice patients to seek medical attention immediately if they experience new or unexpected symptoms suspicious for a ventricular arrhythmia. Medtronic Patient Services is available to assist patients at 800-551-5544 (Monday-Friday, 8am-5pm Centra

Recall details

Recall number
Z-0585-2018
Status
Terminated
Initiated by the company
Dec 19, 2017
Posted by the FDA
Feb 1, 2018
Recalling firm
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Quantity
951 devices total
Product code
LWS
FDA submissions
P980016S586, P980016S536
Distribution
US Nationwide Distribution.

Product description

Implantable Cardioverter Defibrillators (ICDs), Evera MRI. Labeled as the following: a. Evera MRI XT DR ICD DF1 (Product No. DDMB1D1); b. Evera MRI XT DR ICD DF4 (Product No. DDMB1D4); c. Evera MRI XT VR ICD DF4 (Product No. DVMB1D4); d. Evera MRI S DR ICD DF1 (Product No. DDMC3D1); EXPANSION: e. Evera MRI S DR ICD DF4 (Product No. DDMC3D4) Implantable devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia. Some devices can also provide resynchronization therapy, including sequential biventricular pacing (CRT-D devices).

Code information (models, lots, serial numbers)

Show the full code information (3k characters)

a. Product No. DDMB1D1: UDI 00643169929197 (Serial No. CWA207109H) (EXPANSION Serial No. CWA207132H, CWA200074H, CWA200576H, CWA200619H, CWA200837H, CWA200888H, CWA201175H, CWA201443H, CWA201760H, CWA201771H, CWA203003H, CWA203389H, CWA204111H, CWA204889H, CWA204890H, CWA205625H, CWA205694H, CWA205695H, CWA205703H, CWA206083H, CWA206177H, CWA206178H, CWA206179H, CWA206180H, CWA206330H, CWA206666H, CWA206667H, CWA206779H, CWA206780H, CWA207122H, CWA207123H, CWA207125H, CWA207126H, CWA207128H, CWA207131H, CWA207218H, CWA207220H, CWA206176H, CWA205686H, CWA205697H, CWA206175H, CWA206785H, CWA207120H), 00643169717282 (Serial No. CWA200012H, CWA202259H, CWA203498H);  b. Product No. DDMB1D4: UDI 00643169929203 (Serial No. PFZ228501H, PFZ228504H, PFZ228590H, PFZ228836H, PFZ228838H, PFZ229236H, PFZ229520H), 00643169314092 (Serial No. PFZ223779H, PFZ214605H) (EXPANSION: Serial No. PFZ203846H, PFZ200814H, PFZ202951H, PFZ203501H, PFZ203757H, PFZ204555H, PFZ206452H, PFZ206453H, PFZ206456H, PFZ206595H, PFZ206712H, PFZ206881H, PFZ206900H, PFZ209342H, PFZ209675H, PFZ210215H, PFZ210528H, PFZ210537H, PFZ210538H, PFZ210539H, PFZ210791H, PFZ214550H, PFZ214823H, PFZ214971H, PFZ215285H, PFZ215626H, PFZ215627H, PFZ215630H, PFZ215631H, PFZ215876H, PFZ216099H, PFZ216100H, PFZ216973H, PFZ216974H, PFZ217247H, PFZ218221H, PFZ218432H, PFZ218433H, PFZ218435H, PFZ218436H, PFZ218437H, PFZ218877H, PFZ219142H, PFZ219143H, PFZ219373H, PFZ220377H, PFZ221425H, PFZ221716H, PFZ221724H, PFZ221725H, PFZ221726H, PFZ222665H, PFZ223869H, PFZ225847H, PFZ225849H, PFZ225850H, PFZ225851H, PFZ225852H, PFZ225854H, PFZ226023H, PFZ226092H, PFZ226179H, PFZ226185H, PFZ226442H, PFZ226443H, PFZ226513H, PFZ226586H, PFZ226924H, PFZ227065H, PFZ227408H, PFZ227453H, PFZ227454H, PFZ227455H, PFZ227456H, PFZ227778H, PFZ227779H, PFZ227780H, PFZ227781H, PFZ228380H, PFZ228459H, PFZ228560H, PFZ229000H, PFZ229904H, PFZ230774H, PFZ210536H, PFZ215361H, PFZ222655H, PFZ218434H, PFZ225120H, PFZ225848H, PFZ226091H, PFZ236813H, PFZ236991H, PFZ230454H);  c. Product No. DVMB1D4: UDI 00643169314153 (Serial No. PKZ203327H, PKZ210673H) (EXPANSION: Serial No. PKZ202755H, PKZ203168H, PKZ204640H, PKZ205375H, PKZ205377H, PKZ206276H, PKZ206383H, PKZ208113H, PKZ208114H, PKZ208527H, PKZ208981H, PKZ210816H, PKZ210819H, PKZ210927H, PKZ210928H, PKZ210939H, PKZ211270H, PKZ211617H, PKZ211619H, PKZ211620H, PKZ211715H, PKZ210817H, PKZ211289H);  d. Product No. DDMC3D1: UDI 00643169717244 (Serial No. CWC200055H);   EXPANSION:  e. Product No. DDMC3D4: UDI 00643169314139 (Serial No. PHZ201438H)

Devices named in this recall

7 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 78888).