Evera MRI™ XT VR SureScan® DVMB1D4 (UDI 00643169314153)Recalls & reported problems

Medtronic, Inc. · ICD DVMB1D4 EVERA MRI VR XT DF4 US · Class 3 · Implant

All models of Evera MRI XT VR SureScan (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic Evera MRI XT VR is an implantable defibrillator for abnormal heart rhythms. Of its 272 reports, 212 are injury reports, 53 are malfunction reports, and 7 deaths are listed. Many describe inappropriate shocks, and some describe sensing problems and batteries draining early. Reports fell from 29 in the previous year to 9 in the latest year. 2 recalls name this device: 1 open, covering early battery replacement timing, and 1 closed, covering shock therapy.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
272
Deaths reported
7
Injuries reported
212
Recalls naming it
2

Summary

The FDA's MAUDE database contains 272 adverse event reports that list this exact device (DI 00643169314153), received between Dec 9, 2016 and Jun 19, 2026 — classified as 7 deaths, 212 injuries, 53 malfunctions.

The most frequently reported specific device problem is “Inappropriate/Inadequate Shock/Stimulation” (80 reports).

2 FDA recall records name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 1 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 2 reportsNov 2024: 6 reportsDec 2024: 1 reportsJan 2025: 5 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 4 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 3 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 3, 2021Class IOpen, Classified

    A small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation

    Z-1305-2021 · Root cause: Device Design · Recall details · FDA record ↗

  2. Dec 19, 2017Class ITerminated

    Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).

    Z-0585-2018 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Inappropriate/Inadequate Shock/Stimulation80
  • Device Sensing Problem75
  • Adverse Event Without Identified Device or Use Problem70
  • Premature Discharge of Battery34
  • Pacing Problem28
  • Over-Sensing20
  • Battery Problem13
  • Electromagnetic Interference13

Patient problems

  • Shock from Patient Lead(s)82
  • No Clinical Signs, Symptoms or Conditions64
  • Unspecified Infection44
  • No Known Impact Or Consequence To Patient38
  • Death7
  • Tachycardia7
  • Sepsis6
  • Bradycardia5

Recent adverse event reports

The 25 most recent of 272 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2026Electromagnetic Interference; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) DUE TO HIGH RATE NON-SUSTAINED EPISODES AND ELEVATED SENSING INTEGRITY COUNTER (SIC). THE LIA WAS A RESULT OF ELECTROMAGNETIC INTERFERENCE DURING A PROCEDURE WHERE CAUTERYWAS USED TO REPAIR A FISTULA. THE RV LEAD AND IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25565776 ↗
  • InjuryApr 6, 2026Device Alarm System; Failure to Interrogate

    IT WAS FURTHER REPORTED THAT THE DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24792754 ↗
  • InjuryFeb 18, 2026Defibrillation/Stimulation Problem; Low impedance

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED 10 HIGH-VOLTAGE SHOCKS FOR AN EPISODE DETECTED IN THE VENTRICULAR TACHYCARDIA (VT) ZONE. REVIEW OF THE EPISODE DATA SHOWED THAT AT LEAST FOUR OF THE HIGH-VOLTAGE THERAPIES EXPERIENCED SHORT-CIRCUIT PROTECTION (SCP) DURING THE FIRST PHASE OF DELIVERY, RESULTING IN LESS THAN THE PROGRAMMED HIGH-VOLTAGE ENERGY BEING DELIVERED. NOT ALL HIGH-VOLTAGE THERAPIES WERE ABLE TO BE REVIEWED DUE TO EPISODE DURATION AND DEVICE STORAGE LIMITATIONS, BUT THERAPIES FOR THE EVENT WERE CLASSIFIED OVERALL AS UNSUCCESSFUL AT TERMINATING THE RHYTHM. REVIEW OF THE DEVICE DATA ALSO SHOWED THAT FOLLOWING HIGH-VOLTAGE THERAPIES AND SCP EVENTS, ALERTS WERE TRIGGERED FOR LOW OUT OF RANGE (OOR) VENTRICULAR PACING LEAD IMPEDANCE ACROSS ALL PATHWAYS. THE ICD WAS LATER EXPLANTED AND REPLACED WITH A DUAL-CHAMBER SYSTEM. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Tachycardia

    MAUDE report 24381752 ↗
  • InjuryFeb 2, 2026Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED A POSSIBLE INAPPROPRIATE SHOCK FOR AN ATRIAL ARRHYTHMIA DETECTED AS VENTRICULAR FIBRILLATION (VF). THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24233926 ↗
  • InjuryJan 26, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND SHOCKED FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) WHICH THE DEVICE DETECTED AND TREATED AS VENTRICULAR TACHYCARDIA (FVT). THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 24178503 ↗
  • InjuryJan 5, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL EXPERIENCING CHEST PAIN, COMPLETE HEART BLOCK, AND BRADYCARDIA. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS REPROGRAMMED TO INCREASE THE PATIENT'S RATE. THE ICD WAS LATER REMOVED AND UPGRADED TO A BI-VENTRICULAR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bradycardia; Chest Pain; Heart Block

    MAUDE report 23966797 ↗
  • MalfunctionDec 10, 2025Pacing Problem; Device Sensing Problem; Pacing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) THERAPY FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) FOR A DEVICE-CLASSIFIED EPISODE OF VENTRICULAR FIBRILLATION (VF) DUE TO THE DEVICE DISCRIMINATOR NOT WITHHOLDING THERAPY. THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23770777 ↗
  • MalfunctionDec 8, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED AN ACUTE ISSUE. THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23743830 ↗
  • MalfunctionNov 25, 2025Pacing Problem; Device Sensing Problem; Pacing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DETECTED AS VENTRICULAR TACHYCARDIA (VT). THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23649048 ↗
  • InjuryOct 15, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND AN INAPPROPRIATE SHOCK FOR VENTRICULAR FIBRILLATION (VF) THAT WAS SUSPECTED TO BE SUPRA VENTRICULAR TACHYCARDIA (SVT). THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 23304508 ↗
  • InjurySep 5, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED ANTI-TACHYCARDIA PACING (ATP) AND SHOCKS THAT WERE POSSIBLY INAPPROPRIATE. THE EPISODES WERE INITIALLY DETECTED AS SUPRA VENTRICULAR TACHYCARDIA (SVT) AND THERAPY WAS APPROPRIATELY WITHHELD; HOWEVER, DUE TO THE PROGRAMMED HIGH-RATE TIME OUT FUNCTION, THERAPY WAS DELIVERED FOR DETECTED VENTRICULAR FIBRILLATION (VF). THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22978308 ↗
  • InjuryAug 29, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) PATIENT EXPERIENCED SYNCOPE. A REMOTE TRANSMISSION INDICATED THAT ICD DELIVERED INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND SHOCKED FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) WHICH THE DEVICE DETECTED AND TREATED AS VENTRICULAR FIBRILLATION (VF). THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s); Syncope/Fainting

    MAUDE report 22923275 ↗
  • InjuryAug 15, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED SYMPTOMATIC HEART BLOCK WITH ASYNCHRONOUS PACING WITH THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). THE ICD WAS EXPLANTED AND REPLACED. ADDITIONALLY, IT WAS NOTED THAT A VENOGRAM WAS PERFORMED ON THE LEFT WHICH REVEALED AN OCCLUDED LEFT SUBCLAVIAN. THE RIGHT VENTRICULAR (RV) LEAD REMAINS IN USE BUT THE NEW SYSTEM HAD TO BE PLACED ON THE OTHER SIDE. DURING THE IMPLANT PROCEDURE, THE LEFT VENTRICULAR (LV) LEAD DISLODGED AND THE PHYSICIAN WAS NOT ABLE TO GET IT BACK INTO PLACE SO THE LEAD WAS ABANDONED. THE LEFT BUNDLE LEAD WAS PLACED BUT THEN DISLODGED UPON SPLITTING THE SHEATH AND WAS REPLACED WITH A RIGHT VENTRICULAR (RV) LEAD. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Heart Block

    MAUDE report 22820849 ↗
  • InjuryAug 13, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE SHOCKS FOR ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) WITH RAPID VENTRICULAR RESPONSE EPISODE THAT WAS DETECTED AS VENTRICULAR FIBRILLATION (VF). THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22792843 ↗
  • InjuryJul 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT VENTRICULAR PACING FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) INITIATED A RUN OF VENTRICULAR TACHYCA RDIA (VT) ON TELEMETRY. PACING PARAMETERS WERE REPROGRAMMED AND THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Tachycardia

    MAUDE report 22574117 ↗
  • InjuryJul 14, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED BY THE PATIENT THAT THEY EXPERIENCED APPROXIMATELY THREE EPISODES OF SYNCOPE, DURING WHICH THEY FELL TO THE GROUND. THE PATIENT STATED THAT EMERGENCY MEDICAL TECHNICIAN (EMT) PERSONNEL NOTED AN ABNORMAL RHYTHM AND TRANSPORTED THEM TO THE EMERGENCY ROOM (ER). THE PATIENT RECALLED AN ADDITIONAL EPISODE THREE MONTHS AND SIX MONTHS PRIOR. THE PATIENT EXPRESSED CONCERN THAT ¿SOMETHING IS NOT CONNECTING¿ IN THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). THE PATIENT FURTHER REPORTED FEELING A SHOCK-LIKE SENSATION WHEN TOUCHING DOORKNOBS BUT WAS UNSURE IF THE ICD DELIVERED ANY THERAPY. THE PATIENT WAS UNABLE TO CONFIRM WHETHER THE THERAPY WAS DELIVERED APPROPRIATELY. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s); Syncope/Fainting; Tachycardia

    MAUDE report 22500412 ↗
  • MalfunctionJul 3, 2025Battery Problem; Battery Problem

    IT WAS FURTHER REPORTED THAT THE ICD WAS EXPLANTED AND REPLACED WITH A GENERATOR REPLACEMENT DUE TO THE DEVICE AT END OF BATTERY LIFE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22395990 ↗
  • InjuryJun 21, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE THERAPY FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) DETECTED AS VENTRICULAR FIBRILLATION (VF) DUE TO A FEATURE THAT ALLOWS THE DEVICE TO OVERRIDE SVT DISCRIMINATION AND DELIVER THERAPY FOR A TACHYARRHYTHMIA THAT CONTINUES BEYOND A PROGRAMMED LENGTH OF TIME. THE PATIENT WAS HOSPITALIZED AND THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22303818 ↗
  • InjuryJun 10, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT POTENTIALLY RECEIVED INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND SHOCK THERAPY FOR AN EPISODE DETECTED IN THE FAST VENTRICULAR TACHYCARDIA (FVT) ZONE THAT WAS SUSPECTED TO BE ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22179530 ↗
  • InjuryMay 27, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED DUE TO INFECTION. IT WAS NOTED THAT THE SOURCE OF THE INFECTION WAS RECENT COLON SURGERY AND THE INFECTION WAS CAUSED BY GRAM-POSITIVE COCCI. THE PATIENT ALSO EXPERIENCED FEVER, SEPSIS, SEPTIC SHOCK, BACTEREMIA, AND VEGETATION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bacteremia; Fever; Sepsis; Septic Shock; Unspecified Infection

    MAUDE report 22090865 ↗
  • InjuryMay 23, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND HIGH VOLTAGE THERAPY FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT THE DEVICE DETECTED AS VENTRICULAR FIBRILLATION (VF). THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22079166 ↗
  • InjuryMay 19, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE PATIENT RECEIVED POSSIBLE INAPPROPRIATE THERAPY FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT WAS DETECTED AS FAST VENTRICULAR TACHYCARDIA (FVT). IN ADDITION, THERE WAS POSSIBLE T-WAVE OVERSENSING (TWOS) OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND RV LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22045113 ↗
  • InjuryMay 16, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) POSSIBLY DELIVERED AN INAPPROPRIATE SHOCK FOR SUPRA VENTRICULAR TACHYCARDIA (SVT). THE ICD DETECTED THE EPISODE AS VENTRICULAR FIBRILLATION (VF). THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPL ICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 22040078 ↗
  • MalfunctionApr 3, 2025Device Sensing Problem; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) MISCLASSIFIED A VENTRICULAR TACHYCARDIA (VT) EPISODE AS SUPRA VENTRICULAR TACHYCARDIA (SVT). THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21766799 ↗
  • InjuryMar 18, 2025Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED INAPPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND SHOCKED FOR AN ARRHYTHMIA DETECTED IN THE VENTRICULAR TACHYCARDIA (VT) ZONE. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 21635021 ↗

Other Implantable cardioverter defibrillator (non-CRT) devices

All Implantable cardioverter defibrillator (non-CRT) devices →