Skip to content

Ascenda® 8784 (UDI 00643169359178)Recalls & reported problems

Medtronic, Inc. · CATH 8784 PROXIMAL REV KIT OLD CATH eMAN · Class 3

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169359178), received between Sep 4, 2018 and Feb 3, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

39 other recalls cover devices cleared under the same FDA submission (P860004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. May 10, 2024Open, Classified

    Design update to the Ascenda Intrathecal Catheter Models 8780, 8781, and 8784. The intent of the design update is to reduce the potential for tissue growth into the Ascenda catheter connector which may potentially lead to catheter occlusion.

    Z-2352-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information2

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 3, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    INFORMATION WAS RECEIVED FROM A HEALTH CARE PROFESSIONAL (HCP) REGARDING A CATHETER, MODEL 8784. IT WAS REPORTED THAT THE CATHETER WAS OPENED BUT NOT USED IN THE CASE BECAUSE THE COLLET WAS BROKEN. NO PATIENT SYMPTOMS WERE REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9659494 ↗
  • MalfunctionSep 4, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2018, INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE (REP) REGARDING A PATIENT RECEIVING AN UNKNOWN DRUG (UNKNOWN DOSE AND CONCENTRATION) VIA AN IMPLANTABLE PUMP FOR CHRONIC PAIN. ON (B)(6) 2018, DURING A PROCEDURE, THE COLLET ON THE CONNECTOR PIN WAS MISSING. IT WAS NOT CLEAR WHETHER THE COLLET WAS MISSING FROM THE KIT OR WHETHER IT FELL OF WHEN THE KIT WAS OPENED. THEY COULD NOT FIND THE COLLET WHEN THEY SEARCHED FOR IT, BUT "IT HAD NOT BEEN NEAR THE INCISION WHEN WE DISCOVERED IT WAS LOST". THERE WERE NO ENVIRONMENTAL, EXTERNAL, OR PATIENT FACTORS THAT MAY HAVE LED OR CONTRIBUTED TO THE ISSUE. A CONNECTOR FROM A CATHETER (8785) KIT WAS USED. THE ISSUE WAS RESOLVED AT THE TIME OF THIS REPORT. THE PATIENT'S STATUS WAS ALIVE - NO INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7842320 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P860004. One submission often covers many models, so check the product and lot details in each record.

  • Z-2201-2026 · Apr 8, 2026 · Open, Classified
    Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900A CLINICIAN TABLET US, Model Number CT900
  • Z-1506-2025 · Jan 9, 2025 · Open, Classified
    A820 myPTM Software Application associated with Medtronic SynchroMed Pump and Infusion System.
  • Z-0615-2024 · Nov 13, 2023 · Open, Classified
    Medtronic SynchroMed II, Model 8637-40, Programmable pump
  • Z-0614-2024 · Nov 13, 2023 · Open, Classified
    Medtronic SynchroMed II, Model 8637-20, Programmable pump
  • Z-0522-2024 · Jul 13, 2020 · Completed
    Medtronic SynchroMed, Model A10
  • Z-0508-2020 · Oct 11, 2019 · Terminated
    Synchromed II programmable pump, Model numbers 8637-20 and 8637-40
  • Z-0537-2018 · Sep 21, 2017 · Terminated
    Medtronic Ascenda(R) Intrathecal Catheter, Models: (a) 8780, UDI: 00643169202122, 00643169783027 (b) 8781, UDI: 00643169202139 (c) 8784, UDI: 00643169202153 The Ascenda Intrath
  • Z-0896-2018 · Aug 9, 2017 · Open, Classified
    Medtronic Synchromed II Model 8637 supplied in 20mL and 40 mL reservoir size and a catheter. Product Usage: The implantable Model 8637 SynchroMed II programmable pump is part of a
  • Z-0788-2017 · Oct 3, 2016 · Terminated
    Medtronic SynchroMed II Implantable Drug Infusion System. This system includes: Model 8870 Software Application Card, Model 8840 N'Vision Clinician Programmer, and Model 8637 Sy
  • Z-1271-2016 · Feb 9, 2016 · Terminated
    Medtronic, Model 8781 Ascenda" Intrathecal Catheter kit, Rx Only, Sterile Ethylene Oxide.
  • Z-1681-2015 · Apr 10, 2015 · Terminated
    Medtronic SynchroMed II Implantable Infusion Pumps, models 8637-20 and 8637-40. Sterile. Product Usage: The implantable Medtronic SynchroMed II programmable pumps are part of an
  • Z-1060-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed II (Model No: 8637) implantable drug pump The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the follow
  • Z-1061-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed implantable drug pump Refill Kits (Model No: 8551, 8555, 8561, 8562, 8564, 8565, 8566 The Model 8551, 8555, 8561, 8562, 8564, 8565, and 8566 Refill Kits ar
  • Z-1062-2011 · Jan 14, 2011 · Terminated
    Medtronic SynchroMed EL (Model No: 8626 and 8627) implantable drug pump The SynchroMed EL Infusion System is indicated for use when patient therapy requires the chronic infusion
  • Z-2276-2009 · Jul 1, 2009 · Terminated
    Medtronic SynchroMed II Programmable Pump, 8637, Rx Only, Sterilized by ethylene oxide gas, Medtronic inc., Minneapolis, MN 55432-5604. (The SynchroMed II Pump is supplied in 20 m

…and 24 more.

Other PUMP, INFUSION, IMPLANTED, PROGRAMMABLE devices

All PUMP, INFUSION, IMPLANTED, PROGRAMMABLE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.