VISAO® 3334800 (UDI 00643169360938)Recalls & reported problems
Medtronic Xomed, Inc. · DRILL 3334800 VISAO 80K HANDPIECE ROHS · Class 2
All models of ViSAO (16)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This is a Medtronic Xomed VISAO surgical drill handpiece used in ENT procedures. The 65 reports mostly describe the handpiece overheating during use, along with vibration and a rattling or wobbling burr. No deaths and no injuries are reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 65
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 65 adverse event reports that list this exact device (DI 00643169360938), received between Mar 30, 2017 and Jun 3, 2025 — classified as 65 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (55 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device55
- Vibration10
- Noise, Audible3
- Connection Problem1
- Leak/Splash1
- Mechanical Jam1
- Mechanical Problem1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions39
- No Consequences Or Impact To Patient18
- No Known Impact Or Consequence To Patient4
- No Patient Involvement4
Recent adverse event reports
The 25 most recent of 65 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 3, 2025Vibration; Vibration
IT WAS REPORTED THAT DURING THE OPERATION THE VISAO BALL TOOL WHIPS AND THE BALL TOOL DIDN¿T STAY IN THE AXIS OF ROTATION. THE CUSTOMER THINK THAT THE PROBLEM COME FROM EXTENDED BUR GUARD WITH IRRIGATION. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22129679 ↗MalfunctionFeb 12, 2025Vibration; Vibration
IT WAS REPORTED THAT DURING PRE-OP, THE VISAO DRILL HAD BUR WOBBLING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21362690 ↗MalfunctionFeb 6, 2025Vibration; Vibration
IT WAS REPORTED THAT DURING PRE-OP, THE DEVICE HAD BUR WOBBLES. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21321446 ↗MalfunctionFeb 4, 2025Vibration; Vibration
IT WAS REPORTED THAT BEFORE OPERATION, THE BUR WOBBLES IN VISAO HANDPIECE. THERE IS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21304643 ↗MalfunctionNov 11, 2024Leak/Splash; Overheating of Device; Leak/Splash; Overheating of Device
IT WAS REPORTED THAT INTRA-OP, THE VISAO DRILL WAS HEATING UP WHEN USED AND OIL LEAKING OUT OF HANDPIECE. AS SOON AS HANDPIECE OVERHEATS, USER STOPPED USING IT IMMEDIATELY. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20649544 ↗MalfunctionNov 8, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING TESTING, THE LAB TECH STATED THAT THE NOSE BEARING BECAME HOT EXTREMELY FAST. THERE IS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20638134 ↗MalfunctionJul 25, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT PRIOR TO USE, THE DEVICE WAS OVERHEATING. OVERHEATING WAS NOTICED WITH IN ONE MINUTE OF USAGE. THE DEVICE WAS RUN AT THE SPEED OF 80000 RPM IN FORWARD MODE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE PRE-TEMPERATURE TEST RESULT OF THE HANDPIECE FOR ONE MINUTE AT 80,000 RPMS, THE RESULT WAS 50.2 DEGREE CELSIUS(EXCEED 47.78DEGREE). THERE WAS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19834560 ↗MalfunctionJun 17, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT PRE-OP, VISAO HANDPIECE WAS OVERHEATING WHEN RUNNING ABOUT 1 MINUTE. PROCEDURE WAS COMPLETED WITH BACK-UP PRODUCT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19545843 ↗MalfunctionJun 17, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING PRE-OP, THE VISAO DRILL WAS OVERHEATING AT ABOUT 1 MINUTE OF USAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19545660 ↗MalfunctionMay 21, 2024Overheating of Device; Overheating of Device
PRE-REPAIR TEMPERATURE TEST OF THE DEVICE PERFORMED FOR 1 MIN, FOUND THAT AT NOSE - 114.9F, MIDDLE - 115.9F AND REAR - 84.4F WHICH HAVE NOT EXCEEDED 118.4 F HENCE DOES NOT MEET THE REPORTING CRITERIA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19361530 ↗MalfunctionMar 25, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED DURING INTRA-OP THAT THE DEVICE HAD HOT BEARING AND NOT WORKING. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18966813 ↗MalfunctionJan 12, 2024Vibration; Vibration
IT WAS REPORTED THAT DURING THE OPERATION, THE BUR WOBBLES. THE BEARING BEAD FELL OFF AND SHOOK. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18500750 ↗MalfunctionDec 4, 2023Overheating of Device; Use of Device Problem; Overheating of Device; Use of Device Problem
ADDITIONAL INFORMATION RECEIVED, THE EVENT OCCURRED INTRA-OP AND THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18253859 ↗MalfunctionAug 29, 2023Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING SETUP, THE VISAO DRILL HAD NO POWER. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17643050 ↗MalfunctionAug 17, 2023Overheating of Device; Overheating of Device
IT WAS REPORTED THAT PRE-OP, THE VISAO DRILL HANDPIECE OVERHEATING WAS NOTICED IN LESS THAN A MINUTE, WITH THE SPEED OF 8000RPM IN THE FORWARD MODE AND IRRIGATION WAS USED WITH 40CC/MIN. PROCEDURE WAS COMPLETED WITHOUT HANDPIECE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17563538 ↗MalfunctionMar 10, 2023Overheating of Device; Mechanical Jam; Overheating of Device; Mechanical Jam
IT WAS REPORTED THAT DURING PROCESS, VISAO DRILL WAS OVERHEATING AND JAMMED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16522957 ↗MalfunctionSep 29, 2022Overheating of Device; Overheating of Device
A HEALTH CARE PROFESSIONAL REPORTED THAT, DURING SETUP HANDPIECE WAS OVERHEATING. NO PATIENT INVOLVED IN THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15505728 ↗MalfunctionSep 26, 2022Overheating of Device; Overheating of Device
A HEALTH CARE PROFESSIONAL REPORTED THAT, DURING OPERATION HANDPIECE WAS OVERHEATING. NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED AFTER FOLLOW UP THAT DEVICE WAS OVERHEATING IN LESS THAN ONE MINUTE, PROCEDURE WAS COMPLETED WITH SPARE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15489296 ↗MalfunctionAug 10, 2022Vibration; Vibration
HEALTHCARE PROFESSIONAL REPORTED THAT THE DEVICE BUR WOBBLES DURING INTRA-OP. THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED , VIBRATING DEVICE USED ON THE PATIENT. THE PROCEDURE WAS COMPLETED WITH THE BACK UP DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15202350 ↗MalfunctionJul 18, 2022Overheating of Device; Overheating of Device
HCP REPORTED THAT DURING OPERATION THE DEVICE OVERHEATS. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15041330 ↗MalfunctionApr 21, 2022Vibration; Vibration
A HEALTHCARE PROFESSIONAL (HCP) REPORTED THAT INTRA OP, THE DEVICE WAS JUMPING. THERE WAS NO PATIENT IMPACT. THERE WAS NO DELAY IN PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14166960 ↗MalfunctionApr 19, 2022Overheating of Device; Overheating of Device
THE DISTRIBUTOR REPORTED THAT WHEN THE DOCTOR INSTALLED THE HANDPIECE INTO THE IPC FOR TESTING DURING FES SURGERY, HE FOUND THE MOTOR OF THE HANDPIECE WOULD GET HOT AND THE BUR COULDN'T INSTALL INTO THE HANDPIECE. SO, THE DOCTOR DECIDED TO USE A BACKUP HANDPIECE TO COMPLETE THE SURGERY. NO INTERVENTION PERFORMED. THE SURGERY DELAYED FOR 10 MINUTES. NO PATIENT IMPACT. THE HEATING ISSUE OCCURRED LESS THAN A MINUTE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14139763 ↗MalfunctionApr 2, 2022Overheating of Device; Overheating of Device
INFORMATION WAS RECEIVED FROM HCP. IT WAS REPORTED THAT DIAGNOSTIC WAS REQUIRED. NO POWER. NO PATIENT IMPACT. ADDITIONAL INFORMATION RECEIVED, CARTRIDGE WAS OVERHEATING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13991999 ↗MalfunctionDec 6, 2021Vibration; Vibration
IT WAS REPORTED THAT THE BUR WOBBLES IN THE DRILL DURING OPERATION. THERE WAS NO PATIENT IMPACT. ON FOLLOW UP, THE INFORMATION WAS UNKNOWN WHETHER THE DEVICE WAS USED ON PATIENT OR NOT. NO DELAY IN SURGERY, BACK UP DEVICE WAS USED TO COMPLETE THE SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12932948 ↗MalfunctionDec 2, 2021Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING PREPARATION FOR USE/PRE-OP THE DRILL GETTING VERY HOT. THERE WAS NO PATIENT INVOLVED. NO PATIENT / USER COMPLICATIONS /HARMS AS A RESULT OF THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12915180 ↗
Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices
- RemB 6400034000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,618 reports
- NA XMAXAnspach Effort, The · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,225 reports
- NA EG1ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE859 reports
- NA CRANI-ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall772 reports
- NA 5450400000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE775 reports
- NA 05.001.010Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE770 reports
- RemB 6400015000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE603 reports
- NA QD11Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE580 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.