ViSAORecalls & reported problems across all models
Medtronic Xomed, Inc. · 16 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 412
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 0
We found 16 models of ViSAO by Medtronic Xomed, Inc. in the FDA's device registry (GUDID). Together they are named in 412 adverse event reports in the FDA's MAUDE database, received between Dec 6, 2016 and Sep 11, 2026 — classified as 3 injuries, 409 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (322 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported ViSAO problems
Device problems
- Overheating of Device322
- Vibration78
- Noise, Audible21
- Break10
- Connection Problem7
- Use of Device Problem7
- Material Fragmentation6
- Material Separation5
Patient problems
- No Clinical Signs, Symptoms or Conditions254
- No Consequences Or Impact To Patient105
- No Known Impact Or Consequence To Patient31
- No Patient Involvement28
- Insufficient Information2
- Radiation Exposure, Unintended2
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
ViSAO models
16 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Visao™ 3334800DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE262 reports
- VISAO® 3334800DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE65 reports
- VISAO® 3334800DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE55 reports
- Visao™ 3334800TDRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE12 reports
- ViSAO® 31181098EBUR, EAR, NOSE AND THROAT5 reports
- VISAO™ 3334635DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE2 reports
- ViSAO® 31181598EBUR, EAR, NOSE AND THROAT2 reports
- VISAO® 3334800TDRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 report
- VISAO™ 3334635DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 report
- ViSAO® 31152090EBUR, EAR, NOSE AND THROAT1 report
- ViSAO® 311910115EBUR, EAR, NOSE AND THROAT1 report
- ViSAO® 311920115EBUR, EAR, NOSE AND THROAT1 report
- ViSAO® 31202090EBUR, EAR, NOSE AND THROAT1 report
- ViSAO® 31151098EBUR, EAR, NOSE AND THROAT1 report
- ViSAO® 312120115EBUR, EAR, NOSE AND THROAT1 report
- Visao® 3318606DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 report
Recent ViSAO adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 11, 2026Vibration
It was reported that during pre-op, The doctor installed the bur and performed a test. During operation, the motor produced an abnormal noise, the bur was wobbling. The doctor tried reinstalling the bur, but the issue persisted. Therefore, the doctor replaced the handpiece with a new one to complete the surgery. The same bur worked well with another handpiece. When other bur was tried with this handpiece, bur wobbled. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50728046 ↗MalfunctionJan 2, 2026Overheating of Device
IT WAS REPORTED THAT PREOPERATIVE, HANDPIECE WAS INSTALLED ONTO THE IPC FOR TESTING, FOUND THAT THE BUR LOCK PLACE BECAME HOT IN LESS THAN 30 SECONDS WHILE THE MOTOR WAS RUNNING. THEY TRIED TO RECONNECT THE HANDPIECE AND BUR BUT THE PROBLEM STILL EXISTED. FINALLY, CHANGED HANDPIECE WITH BACKUP ONE AND THIS WAS FUNCTIONING NORMALLY. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23952327 ↗MalfunctionDec 1, 2025Overheating of Device; Use of Device Problem; Overheating of Device; Use of Device Problem
IT WAS REPORTED THAT INTRA-OP, THE VISAO DRILL WAS RUNNING HOT AND NOT RUNNING RIGHT. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23677720 ↗MalfunctionDec 1, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT INTRA OPERATIVE, HANDPIECE WAS RUNNING HOT AND NOT RUNNING RIGHT. THERE WAS NO PATIENT IMPACT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23677672 ↗MalfunctionNov 25, 2025Overheating of Device; Use of Device Problem; Overheating of Device; Use of Device Problem
IT WAS REPORTED THAT PRE-OP, THE VISAO DRILL MOTOR WAS NOT WORKING PROPERLY AND OVERHEATING. THE HANDPIECE WAS BEING RUN AT 80000 RPM IN FORWARD MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23644159 ↗MalfunctionNov 25, 2025Overheating of Device; Vibration; Overheating of Device; Vibration
IT WAS REPORTED THAT DURING SET UP, HANDPIECE HAD VIBRATION ISSUE. THERE WAS NO PATIENT IMPACT. DURING ANALYSIS HANDPIECE FOUND TO BE OVERHEATING
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23642349 ↗MalfunctionOct 15, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT BEFORE OPERATION, THE VISAO DRILL WAS OVERHEATING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23287312 ↗MalfunctionOct 14, 2025Vibration; Vibration
IT WAS REPORTED THAT PRE-OP, BUR WAS WOBBLING WHEN USED WITH VISAO. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23275974 ↗MalfunctionJul 23, 2025Vibration; Vibration
IT WAS REPORTED THAT BEFORE OPERATION THE BUR WAS WOBBLING IN VISAO. THE BUR WAS WORKING FINE IN ANOTHER HANDPIECE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22573045 ↗MalfunctionJul 18, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING. NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22540360 ↗MalfunctionJun 10, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DEVICE WAS NOT COOLING PROPERLY/HEATING ISSUE. THERE IS NO KNOWN PATIENT IMPACT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 22174862 ↗MalfunctionJun 3, 2025Vibration; Vibration
IT WAS REPORTED THAT DURING THE OPERATION THE VISAO BALL TOOL WHIPS AND THE BALL TOOL DIDN¿T STAY IN THE AXIS OF ROTATION. THE CUSTOMER THINK THAT THE PROBLEM COME FROM EXTENDED BUR GUARD WITH IRRIGATION. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22129679 ↗MalfunctionMay 29, 2025Overheating of Device; Use of Device Problem; Overheating of Device; Use of Device Problem
IT WAS REPORTED THAT DURING THE USE THE DEVICE HEATS UP AFTER USING IT FOR MORE THAN A MINUTE AND NOT WORKING PROPERLY. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22107586 ↗MalfunctionMay 26, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT PRE OP, THE DOCTOR INSTALLED THE HANDPIECE INTO THE IPC FOR TESTING AND FOUND THAT THE MOTOR OVERHEATED AFTER OPERATING FOR LESS THAN ONE MINUTE. THE DOCTOR TRIED TO TEST AGAIN BUT STILL OVERHEATED. FINALLY, THE DOCTOR CHANGED A HANDPIECE TO PROCEED WITH THE SURGICAL PROCEDURE. DEVICE BEING RUN AT A SPEED OF 80000 RPM AND IN FORWARD MODE AND IRRIGATION BEING USED. THERE IS NO PATIENT INVOLVED IN THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22083088 ↗MalfunctionMay 5, 2025Vibration; Vibration
IT WAS REPORTED THAT DURING PRE-OP, BUR WOBBLES IN THE VISAO DRILL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21958721 ↗
Other Medtronic Xomed, Inc. products
- NIM Vital9 models · 1,696 reports · 4 recalls
- NIM-Response13 models · 974 reports
- Indigo6 models · 916 reports
- NIM TriVantage18 models · 541 reports · 2 recalls
- StraightShot6 models · 546 reports
- NIM15 models · 410 reports · 2 recalls
- XPS Straightshot2 models · 356 reports
- RAD29 models · 183 reports · 1 recall
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page