CD HORIZON® Spinal System 25840019580Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 25840019580 BS CNMAS 9.5X80 T/C · Class 2 · Implant
All models of CD HORIZON Spinal System (342)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169413221), received between Feb 19, 2025 and Apr 18, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 18, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED POST A POSTERIOR THORACOLUMBAR FUSION IN A PATIENT. IT WAS REPORTED THAT THE SET SCREW BACKED OFF THE TULIP. A LAMINECTOMY WAS ALSO PERFORMED IN THE THORACIC REGION WHEN THE ILLIAC SCREWS WERE REPLACED. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21863172 ↗MalfunctionFeb 19, 2025Material Deformation; Material Deformation
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR UNKNOWN INDICATION. IT WAS REPORTED THAT THE SCREWS WERE COLD WELDED/STRIPPED DUE TO DRIVER MALFUNCTION INTRAOPERATIVELY. HAD TO HELICOPTER OUT AFTER NUMEROUS ATTEMPTS TO REMOVE USING A MULTITUDE OF DRIVERS. ONE OF THE DRIVERS WAS SEVERELY WARPED AFTER ATTEMPTING TO PLACE 10.5 SCREW. WHEN ATTEMPTING TO PLACE SECOND SCREW (10.5X80) THE TORX TIP OF THE DRIVER SNAPPED OFF AND LODGED INSIDE THE TULIP/TORX INTERFACE OF THE SCREW. FRAGMENTS COULD NOT BE REMOVED AND IT WAS LEFT IN THE PT. THE OTHER DRIVER WAS SEVERELY WARPED AFTER ATTEMPTING TO PLACE 9.5 SCREW. SURGEON WAS FEARFUL OF SNAPPING OFF TIP OF THE DRIVER AGAIN SO THE DECISION WAS MADE TO HELICOPTER THE SCREW OUT AND LEAVE UNILATERAL PELVIC FIXATION. PRODUCTS WILL BE RETURNED. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THE PROCEDURE INVOLVED WAS PSF L2-PELVIS WITH TLIF L3-5. PRE-OP DIAGNOSIS M43.16/M54.16/M40.30/M47.26. BOTH BALLAST DRIVERS WERE SEVERELY WARPED AND STRIPPED. THE PRODUCT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21417821 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page