CD HORIZON® Spinal System 25740019590Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 25740019590 BS CNMAS 9.5X90 TC S · Class 2 · Implant
All models of CD HORIZON Spinal System (342)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169415010), received between Nov 22, 2018 and Oct 30, 2024 — classified as 5 injuries.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth2
- Device Dislodged or Dislocated1
- Material Twisted/Bent1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Pain2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryOct 30, 2024Use of Device Problem; Use of Device Problem
ADDITIONAL INFORMATION WAS RECEIVED THAT PRODUCT MOVED INTO TO THE INSIDE OF THE ILIAC BONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20564066 ↗InjuryApr 3, 2023Material Twisted/Bent; Material Twisted/Bent
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR L5/S BONE FUSION FAILURE, S2AI SET SCREW DISLODGEMENT. DATE OF INITIAL SURGERY FIRST TIME: (B)(6) 2021 , SECOND TIME: (B)(6) 2022 INITIAL SURGERY OR THE PROCEDURE TAKEN: TH4-S2AI L4/5/S PLIF (ADULT SPINAL DEFORMITY) IT WAS REPORTED THAT THE PATIENT WAS INITIALLY FIXED WITH TH4-S2AI, THEN IT WAS ONCE REPLACED WITH S2AI BY REOPERATION AND CONNECTED WITH MRC WITH DOUBLE S2AI ON BOTH SIDES AND PARALLEL ROD EXTENSION TO THE CRANIAL SIDE. RIGHT SET SCREW CAME OFF AND REPLACEMENT RE-OPERATION WAS PERFORMED. THE INNER ROD WAS EXTENDED TO L2 AND IT WAS IN CONTACT WITH THE MRC, SO IT WAS DEPLOYED BELOW L2. WHEN ALL THE IMPLANTS WERE EXPOSED, IT WAS FOUND THAT THE TWO S2AI SET SCREWS ON THE RIGHT SIDE HAD COME OFF AS SHOWN IN THE XRAY, BUT THE SET SCREW ON THE LEFT SIDE HAD ALSO COME OFF ON THE CRANIAL SIDE. IN ADDITION, THE S2AI SCREWS ON BOTH CAUDAL SIDES HAD RODS DETACHED FROM THE HEAD ON BOTH SIDES. ON THE LEFT SIDE, THE SET SCREW WAS NOT DETACHED, BUT THE ROD WAS DETACHED FROM THE HEAD, WHICH SURPRISED THE DOCTOR. FIRST, THE PROLOCK BETWEEN L4/5 WAS REM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16661516 ↗InjuryApr 20, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT PREVIOUSLY IMPLANTED WITH A SCREW IN A THORACOLUMBAR POSTERIOR FIXATION FOR A FRACTURE AT S2AIS. IT WAS REPORTED THAT THE SET SCREW WAS DISLODGED. THE PATIENT COMPLAINED OF LOWER BACK PAIN AS A RESULT OF THIS EVENT. THE PATIENT NEEDED TO BE HOSPITALIZED FOR RE-SURGERY. THE SCREW WAS REPLACED WITH A NEW ONE DURING THE RE-FIXATION SURGERY. THE DATE OF IMPLANT WAS (B)(6) 2022 AND DATE OF REFIXATION WAS (B)(6) 2022. NO FURTHER COMPLICATIONS WERE REPORTED/ ANTICIPATED.
Patient problems: Pain
MAUDE report 14159167 ↗InjuryMar 3, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT THE PATIENT UNDERWENT POSTERIOR SPINAL FUSION (PSF) AND POSTERIOR LUMBAR INTERBODY FUSION (PLIF) AT T10-S2AI DUE TO SPINAL KYPHOSCOLIOSIS. POST OP, SCREW LOOSENING OCCURRED AT LEFT S2AI AND DUE TO THIS RE-OPERATION HAD BEEN DONE. SCREW IN S2AI WAS REMOVED AND REPLACED WITH ILIAC SCREW.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9778806 ↗InjuryNov 22, 2018Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH LUMBAR CANAL STENOSIS AND UNDERWENT SACRAL-ILIAC FUSION AT S2 AND EXTENDING TOWARDS CAUDAL SIDE. POST-OP, THE SET SCREW OF S2AI SCREW BACKED OUT. PAIN OCCURRED DUE TO THE LOOSENING OF S1 SCREW. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2018 TO REPLACE THE SCREW.
Patient problems: Pain
MAUDE report 8098501 ↗
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page