Skip to content

N/A 9735016Recalls & reported problems

Medtronic Navigation, Inc. · SUCTION 9735016 MALLEABLE STDTIP 9-FR · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169443518), received between Jul 1, 2024 and Apr 4, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Device Sensing Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Sensing Problem3
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Unintended Application Program Shut Down1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 4, 2025Device Sensing Problem; Device Sensing Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS) PROCEDURE. IT WAS REPORTED THAT THE PATIENT TRACKER AND INSTRUMENT TRACKER WERE OPERATING PERFECT WITH ALL INSTRUMENTS EXCEPT THE 70 DEGREE SUCKER WOULDN'T VERIFY. ONCE PLUGGED IN, THE MALLEABLE SUCTION WOULDN'T APPEAR ON THE NAVIGATION SCREEN AND DID NOT APPEAR IN THE TRACKER/EMITTER DETAILS (BOTTOM RIGHT CORNER OF SCREEN). MOVING THE MALLEABLE SUCTION ALL AROUND THE PATIENT HEAD, IT APPEARED BRIEFLY IN RED AT EXTREMITIES OF THE MAGNETIC FIELD PRODUCED BY THE SIDE EMITTER. THE MALLEABLE SUCTION WAS PLUGGED INTO OTHER AXIEM PORTS WITH SAME BEHAVIOR CONTINUING. THE MALLEABLE SUCTION WAS ABANDONED AND COMPETED THE CASE WITH A METAL INSTRUMENT SET. THIS ISSUE CAUSED LESS THAN 1 HOUR SURGICAL DELAY. THERE WAS NO REPORTED IMPACT ON PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21777915 ↗
  • MalfunctionMar 10, 2025Device Sensing Problem; Device Sensing Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS) PROCEDURE. IT WAS REPORTED THAT THE MALLEABLE SUCTION WAS PLUGGED INTO AN ELECTROMAGNETIC NAVIGATION INTERFACE UNIT (AXIEM) PORT AND THE SYSTEM WAS UNABLE TO SEE SUCTION IN THE FIELD OF VIEW. A REPLACEMENT SUCTION WAS OPENED WITH NO ISSUE AND PROCEDURE CONTINUED. THERE WAS NO DELAY. THERE WAS NO IMPACT ON PATIENT OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21566363 ↗
  • MalfunctionNov 5, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    ADDITIONAL INFORMATION WAS RECEIVED. THE SITE NURSES PLUGGED THE INSTRUMENT INTO THE NAVIGATION SYSTEM, IT REGISTERED AS IT SHOULD BUT WHEN THE SURGEON TRIED TO VERIFY THE POINTS ON THE PATIENT, IT FAILED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20609088 ↗
  • MalfunctionJul 1, 2024Device Sensing Problem; Unintended Application Program Shut Down; Device Sensing Problem; Unintended Application Program Shut Down

    ADDITIONAL INFORMATION WAS RECEIVED. THE PROCEDURE BEING PERFORMED WAS A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS), THERE WAS A LESS THAN 30 MINUTE SURGICAL DELAY, AND THERE WAS NO IMPACT TO THE PATIENT'S OUTCOME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19646721 ↗

Other Ear, Nose, and Throat Stereotaxic Instrument devices

All Ear, Nose, and Throat Stereotaxic Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page