Scopis 8000-021-002Recalls & reported problems
Stryker Leibinger GmbH & Co. KG · ENT Software with TGS · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 07613327514179), received between Jun 5, 2023 and Apr 22, 2026 — classified as 19 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (19 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Apr 4, 2024Class IIOpen, Classified
Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.
Z-1885-2024 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect Measurement19
- Device Displays Incorrect Message2
Patient problems
- No Clinical Signs, Symptoms or Conditions19
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2026Incorrect Measurement
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1 ¿ MARCH 31 2026. THIS REPORT SUMMARIZES 1 EVENT FOR THE FAILURE: INCORRECT MEASUREMENT. 1 EVENT HAD NO HEALTH CONSEQUENCES OR IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24954034 ↗MalfunctionJan 23, 2026Incorrect Measurement
MalfunctionJul 29, 2025Incorrect Measurement; Incorrect Measurement
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN APRIL 1 ¿ JUNE 30 2025. THIS REPORT SUMMARIZES 6 EVENTS FOR THE FAILURE: INCORRECT MEASUREMENT. - 5 EVENTS HAD NO HEALTH CONSEQUENCES OR IMPACT. - 1 EVENT HAD PROLONGED SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22647405 ↗MalfunctionApr 21, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCORRECT MEASUREMENT. 2 EVENTS HAD PATIENT INVOLVEMENT: NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21875917 ↗MalfunctionJan 21, 2025Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCORRECT MEASUREMENT. 2 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21200737 ↗MalfunctionOct 21, 2024Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 12 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCORRECT MEASUREMENT. - 11 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT. - 1 EVENT LED TO A PROLONGED SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20497730 ↗MalfunctionJul 24, 2024Incorrect Measurement; Incorrect Measurement
THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS IN WHICH THE DEVICE REPORTEDLY HAD AN INCORRECT MEASUREMENT. - 6 EVENTS HAD PATIENT INVOLVEMENT; NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19818520 ↗MalfunctionFeb 27, 2024Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER THE SCANS EITHER WILL NOT LOAD ON THE UNIT OR WILL LOAD THE IMAGES BACKWARDS SO THAT WHEN THE SURGEON IS OPERATING ON THE PATIENT LEFT, THE NAVIGATION/CT SCAN IS SHOWING THAT THEY ARE ON THE PATIENT RIGHT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18790429 ↗MalfunctionFeb 27, 2024Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER THE SCANS EITHER WILL NOT LOAD ON THE UNIT OR WILL LOAD THE IMAGES BACKWARDS SO THAT WHEN THE SURGEON IS OPERATING ON THE PATIENT LEFT, THE NAVIGATION/CT SCAN IS SHOWING THAT THEY ARE ON THE PATIENT RIGHT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18790425 ↗MalfunctionNov 15, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE DEVICE HAS ACCURACY ISSUES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18144615 ↗MalfunctionNov 15, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE DEVICE HAS ACCURACY ISSUES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18144497 ↗MalfunctionNov 1, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE DEVICE WAS INACCURATE DURING A PROCEDURE, SHOWING THE POINTS TO BE OFF FROM THE PHYSICAL REFERENCE POINT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18046618 ↗MalfunctionOct 19, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE DEVICE WOULD BE ACCURATE AT SOME POINTS DURING THE PROCEDURE, THEN BOTH WOULD SUDDENLY BE INACCURATE AT TIMES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17967970 ↗MalfunctionOct 19, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE DEVICE WOULD BE ACCURATE AT SOME POINTS DURING THE PROCEDURE, THEN BOTH WOULD SUDDENLY BE INACCURATE AT TIMES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17967636 ↗MalfunctionOct 19, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE DEVICE WOULD BE ACCURATE AT SOME POINTS DURING THE PROCEDURE, THEN BOTH WOULD SUDDENLY BE INACCURATE AT TIMES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17967075 ↗MalfunctionOct 19, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE DEVICE WOULD BE ACCURATE AT SOME POINTS DURING THE PROCEDURE, THEN BOTH WOULD SUDDENLY BE INACCURATE AT TIMES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17965910 ↗MalfunctionOct 19, 2023Incorrect Measurement; Incorrect Measurement
PER THE CUSTOMER, THE DEVICE WOULD BE ACCURATE AT SOME POINTS DURING THE PROCEDURE, THEN BOTH WOULD SUDDENLY BE INACCURATE AT TIMES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17965908 ↗MalfunctionJun 5, 2023Incorrect Measurement; Device Displays Incorrect Message; Incorrect Measurement; Device Displays Incorrect Message
PER THE CUSTOMER THE VIDEO LAGGED DURING THE CASE WHICH CAN LEAD TO INACCURACIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17063787 ↗MalfunctionJun 5, 2023Incorrect Measurement; Device Displays Incorrect Message; Incorrect Measurement; Device Displays Incorrect Message
PER THE CUSTOMER THE VIDEO LAGGED DURING THE CASE WHICH CAN LEAD TO INACCURACIES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SURGICAL DELAY; NO MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17059677 ↗
Other Ear, Nose, and Throat Stereotaxic Instrument devices
- TruDi NAV Suction TDNS000ZAcclarent, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall61 reports
- TruDi™ Navigation System 1Biosense Webster (Israel) Ltd. · Ear, Nose, and Throat Stereotaxic Instrument1 recall14 reports
- FUSION Compact ™ 9735602Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument64 reports
- StealthStation 9735762Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall0 reports
- Medtronic Navigation 9735762Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall0 reports
- StealthStation Cranial Software 9735762Medtronic Navigation, Inc. · Ear, Nose, and Throat Stereotaxic Instrument1 recall0 reports
- Scopis ENT 8000-020-002Stryker Leibinger GmbH & Co. KG · Ear, Nose, and Throat Stereotaxic Instrument33 reports
- TruDi NAV Suction TDNS070ZAcclarent, Inc. · Ear, Nose, and Throat Stereotaxic Instrument25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page