Skip to content

THUNDER® THNDR190JRecalls & reported problems

Medtronic, Inc. · GDWR THNDR190J THUNDER 5PK 190 J SIL · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169452275), received between Oct 7, 2022 and Oct 7, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 7, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING A PROCEDURE AN ATTEMPT WAS MADE TO USE ONE THUNDER GUIDEWIRE TO TREAT A LESION IN THE CORONARY SINUS. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES NOTED. THE WIRE TIP WAS NOT FORMED BY THE PHYSICIAN. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING INSERTION/DELIVERY. THE GUIDEWIRE WAS EXAMINED AND FLUSHED BEFORE USE WITH NO ISSUES/KINKS NOTED. NO DIFFICULTIES WERE NOTED WHEN LOADING THE GUIDEWIRE. THE GUIDEWIRE WAS NOT USED WITH OTHER DEVICES PRIOR TO DIFFICULTIES OCCURRING. THE WIRE DID NOT PASS THROUGH THE CELL OF A STENT OR A TIGHT CALCIFIED LESION DURING DELIVERY BUT IT IS BELIEVED THAT THERE WAS PROBABLY A VALVE IN THE CORONARY SINUS. IT WAS REPORTED THAT THE GUIDEWIRE KINKED WHILE IN THE PATIENT. IT WAS STATED THAT THE DISTAL PART OF THE ANGIOPLASTY GUIDEWIRE WAS STUCK IN THE DISTAL PART OF THE CATHETER AND IT COULD NOT BE SEPARATED. WHEN THE WHOLE SYSTEM WAS REMOVED IT WAS OBSERVED THAT THE GUIDEWIRE WAS KINKED AND THE RIGIDITY WAS LOST. IT WAS LATER REPORTED THAT A SECTION OF THE CORE WIRE OF THE THUNDER GUIDE WIRE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15559825 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →