ATTAIN HYBRID™ Guide Wire GWR419588Recalls & reported problems
Medtronic, Inc. · GUIDE WIRE GWR419588 108CM GLOBAL CE · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00643169452411), received between Dec 7, 2017 and May 13, 2024 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem4
- Material Fragmentation3
- Detachment of Device or Device Component2
- Material Integrity Problem2
- Activation, Positioning or Separation Problem1
- Break1
- Device Or Device Fragments Location Unknown1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions3
- No Patient Involvement1
- Vascular Dissection1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2024Break; Break
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE GUIDEWIRE BROKE OFF INSIDE THE LEFT VENTRICULAR (LV) LEAD. THE LV LEAD WAS REMOVED AND THE GUIDEWIRE WAS SUCCESSFULLY REMOVED FROM THE DISTAL LEAD TIP. A REPLACEMENT LEAD WAS SUCCESSFULLY IMPLANTED. IT WAS FURTHER REPORTED THAT THE GUIDEWIRE WAS RETURNED TO THE MANUFACTURER AFTER USE WITH NO REASON FOR THE RETURN. THE GUIDEWIRE WAS ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19300865 ↗InjuryApr 4, 2022Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, AFTER ACCESSING THE CORONARY SINUS AND TRYING TO PLACE THE LEFT VENTRICULAR (LV) LEAD IN THE POSTEROLATERAL VEIN, THE PHYSICIAN INJECTED CONTRAST AND EXTRAVASATION OF THE CONTRAST WAS OBSERVED IN THE AREA WHERE THE GUIDEWIRE HAD BEEN MANIPULATED. IT WAS NOTED THAT THE PATIENT HAD CHEST PAIN AND EXPERIENCED A VAGAL REACTION BUT RECOVERED WITHOUT ANY ISSUES AFTER SERUM THERAPY AND MEDICATION. AN ECHO REVEALED A SLIGHT DETACHMENT OF PERICARDIAL LEAVES AROUND THE RIGHT VENTRICLE AND THE RIGHT ATRIOVENTRICULAR GROOVE. DUE TO THIS ISSUE THE PROCEDURE WAS ABORTED. THE LEAD WAS NOT USED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Vascular Dissection
MAUDE report 13997454 ↗MalfunctionMar 4, 2022Material Fragmentation; Activation, Positioning or Separation Problem; Detachment of Device or Device Component; Material Integrity Problem; Material Fragmentation; Activation, Positioning or Separation Problem; Detachment of Device or Device Component; Material Integrity Problem
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE HYBRID GUIDEWIRE THAT WAS INSERTED IN THE LEAD HAD FORMED A KNOT AND WAS LIKELY STUCK IN A DISTAL CAPILLARY OF THE VESSEL. WHILE THE PHYSICIAN ATTEMPTED TO REMOVE THE GUIDEWIRE, IT BECAME FRAYED. A PORTION OF THE GUIDEWIRE WAS REMOVE, BUT A PIECE OF THE SOFT PART REMAINED STICK IN THE VESSEL. A REPLACEMENT GUIDEWIRE WAS USED TO FINISH THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13667707 ↗MalfunctionMar 26, 2021Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT PRIOR TO USE, THE TIP OF THE GUIDEWIRE WAS NOTED TO BE TWISTED/CURLED AND THE PHYSICIAN WAS UNABLE TO INSERT THE GUIDEWIRE INTO THE LEAD. THE GUIDEWIRE WAS NOT USED. SUBSEQUENTLY, THE GUIDEWIRE WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11578452 ↗MalfunctionJan 3, 2020Material Fragmentation; Detachment of Device or Device Component; Material Fragmentation; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, THE TIP OF THE GUIDEWIRE DETACHED AND REMAINED IN THE CORONARY SINUS. IT WAS NOTED TO BE THE THIN BLACK PART OF THE GUIDEWIRE. ANOTHER GUIDEWIRE WAS USED FOR IMPLANT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9548911 ↗MalfunctionApr 17, 2019Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT PRIOR TO THE IMPLANT PROCEDURE, THE DISTAL PORTION OF THE GUIDEWIRE WAS STRETCHED UPON REMOVAL FROM THE PACKAGE. IT WAS NOTED THAT THE PLASTIC CASSETTE PACKAGING WAS INTACT. THE GUIDEWIRE WAS NOT USED, AND THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8524678 ↗MalfunctionMay 19, 2018Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, AFTER REMOVAL OF THE GUIDEWIRE FROM THE LEAD, IT WAS NOTICED THAT THE DISTAL TIP OF THE GUIDEWIRE APPEARED TO HAVE UNRAVELED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7528491 ↗MalfunctionMay 18, 2018Material Fragmentation; Device Or Device Fragments Location Unknown; Material Fragmentation; Device Or Device Fragments Location Unknown
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE, PART OF THE GUIDEWIRE HAD BROKEN AND WAS VISIBLE AT THE END OF THE VEIN. IT WAS NOTED THAT WHEN PLACING THE LV LEAD, THE PHYSICIAN REMOVED THE GUIDEWIRE AND OBSERVED ON THE FLUOROSCOPIC IMAGE THAT PART OF THE GUIDEWIRE WAS BROKEN. THE PART OF THE GUIDEWIRE THAT WAS BROKEN WILL REMAIN INSIDE THE PATIENT AND THE REMAINING PART THAT WAS REMOVED WILL BE RETURNED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7526577 ↗MalfunctionDec 7, 2017Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT WHEN THE GUIDEWIRE WAS REMOVED FROM THE PACKAGING, THE DISTAL END OF THE HYBRIDWIRE SEPARATED. THE GUIDEWIRE WAS NOT USED WITH THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7090729 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page