ZINGER® LIGHT ZNGRLS180HSRecalls & reported problems
Medtronic, Inc. · GDWR ZNGRLS180HS ZINGERLS 5PK 180 S HY · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169452985), received between Aug 10, 2021 and Jan 2, 2024 — classified as 1 death, 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component2
- Adverse Event Without Identified Device or Use Problem1
- Flaked1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
DeathJan 2, 2024Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem
DURING A PROCEDURE AN ATTEMPT WAS MADE TO USE A MEDTRONIC ZINGER GUIDEWIRE AND AN INTUITION GUIDEWIRE. THERE WAS NO DAMAGE NOTED TO DEVICE PACKAGING. ONLY MEDTRONIC DEVICES WERE USED IN THE PROCEDURE. ONE OF THE GUIDEWIRES WAS DAMAGED. IT WAS REPORTED THAT THE DEVICE WAS DETACHED IN PIECES AND ONE OF THE PIECES WAS IN THE BODY OF THE PATIENT AND REMAINED IN THE PATIENT. THEY COULD NOT GET IT FROM THE VESSELS AND AS A RESULT THE PATIENT DIED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 18423861 ↗MalfunctionFeb 22, 2022Flaked; Flaked
IT WAS REPORTED THAT DURING THE PROCEDURE A ZINGER GUIDEWIRE WAS UNABLE TO MOVE AND BECAME STUCK INSIDE THE LEFT VENTRICULAR (LV) LEAD. THE LEAD AND GUIDEWIRE WERE REMOVED AND REPLACED. NO DAMAGE WAS NOTED TO THE PACKAGING OF THE ZINGER GUIDEWIRE AND NO ISSUES WERE NOTED WHEN REMOVING THE DEVICE FROM THE PACKAGING. THE ZINGER GUIDEWIRE WAS NOT INSPECTED. THE ZINGER GUIDEWIRE WAS FLUSHED BEFORE USE, WITH NO ISSUES NOTED. THE WIRE TIP WAS FORMED BY THE PHYSICIAN. NO DIFFICULTIES WERE NOTED WHEN LOADING THE GUIDEWIRE. THE ZINGER GUIDEWIRE WAS NOT USED SUCCESSFULLY WITH OTHER DEVICES PRIOR TO DIFFICULTIES OCCURRING AS THIS WAS THE FIRST TIME IT WAS ATTEMPTED TO BE USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13576435 ↗MalfunctionAug 10, 2021Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation
DURING A PROCEDURE AN ATTEMPT WAS MADE TO USE ONE ZINGER GUIDEWIRE TO TREAT A NON CALCIFIED, MODERATELY TORTUOUS LESION LOCATED IN THE PROXIMAL DIAGONAL ARTERY WITH 90% STENOSIS. THERE WAS NO DAMAGE NOTED TO THE DEVICE PACKAGING. THERE WERE NO ISSUES NOTED REMOVING THE DEVICE FROM THE PROTECTIVE HOOP. THE DEVICE WAS INSPECTED BEFORE USE WITH NO ISSUES NOTED. IT WAS REPORTED THAT DAMAGE WAS NOTED TO THE DEVICE AND THE DEVICE DETACHED IN THE PATIENT. COIL SEPARATION OCCURRED TO THE DISTAL EXTREME OF THE DEVICE. THE TIP OF THE DEVICE DETACHED DURING ADVANCEMENT OF THE DEVICE. THE PROCEDURE WAS COMPLETED USING A SAME SIZED ZINGER GUIDEWIRE. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12299467 ↗
Other WIRE, GUIDE, CATHETER devices
- HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II 1009664Abbott Vascular Inc. · WIRE, GUIDE, CATHETER321 reports
- SYNCHRO 14 M00313010Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls220 reports
- PressureWire™ C12059Lightlab Imaging, Inc. · WIRE, GUIDE, CATHETER1 death296 reports
- SYNCHRO 2 M00326410Stryker Corporation · WIRE, GUIDE, CATHETER2 recalls155 reports
- ENROUTE .014" Guidewire 11686-01Silk Road Medical, Inc. · Wire, Guide, Catheter2 deaths228 reports
- ARROW IPN035915Teleflex Incorporated · WIRE, GUIDE, CATHETER217 reports
- OmniWire Pressure guide wire 89185Philips Image Guided Therapy Corporation · Wire, Guide, Catheter1 recall1 death160 reports
- ARROW IPN055780Teleflex Incorporated · Wire, guide, catheter4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page