Skip to content

ZINGER® LIGHT ZNGRLS180HJ (UDI 00643169452992)Recalls & reported problems

Medtronic, Inc. · GDWR ZNGRLS180HJ ZINGERLS 5PK 180 J HY · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169452992), received between Sep 8, 2017 and Oct 9, 2023 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Operates Differently Than Expected3
  • Detachment of Device or Device Component1
  • Material Deformation1

Patient problems

  • Aortic Dissection3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 9, 2023Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation

    AN ATTEMPT WAS MADE TO USE ONE ZINGER GUIDEWIRE DURING AN UPGRADE TO CARDIAC RESYNCHRONIZATION THERAPY (CRT) PROCEDURE. THE UPGRADE WAS BEING DONE TO IMPLANT A MEDTRONIC ATTAIN LEAD IN THE MID LATERAL VASUS WHICH ORIGINATED FROM THE CORONARY SINUS IN ORDER TO OBTAIN A CRT. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES. THE WIRE TIP WAS NOT FORMED BY THE PHYSICIAN. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING INSERTION/DELIVERY. ACCESS WAS GAINED IN THE LEFT SUBCLAVIAN VEIN. IT WAS REPORTED THAT THE TIP OF THE GUIDEWIRE WAS DAMAGED WHEN ATTEMPTING TO REACH THE LESION. THE GUIDEWIRE WAS INSIDE THE ATTAIN LEAD. THERE WERE NO DIFFICULTIES LOADING THE DEVICE INTO THE LEAD, THE GUIDE WAS LOADED AND THEN BECAME DAMAGED. THE GUIDEWIRE TIP FRAYED. THE GUIDE HAD NOT BEEN USED WITH OTHER DEVICES PRIOR TO THE EVENT. THERE WAS RESISTANCE WHILE REMOVING THE GUIDEWIRE FROM THE LEAD. A REPLACEMENT NON-MDT GUIDEWIRE WAS USED WITH THE SAME LEAD TO COMPLETE THE PROCEDURE. THE LEAD WAS LATER SUCCESSFULLY IMPLANTED WITH

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17894597 ↗
  • InjurySep 8, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT IS REPORTED THAT THE WIRE IS HARD, BAD TORQUE, TENDS TO DISSECT THE ARTERY WITH EASE. PATIENT PRESENTED WITH AORTIC DISSECTION WHEN WIRE WAS USED.

    Patient problems: Aortic Dissection

    MAUDE report 6851100 ↗
  • InjurySep 8, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT IS REPORTED THAT THE WIRE IS HARD, BAD TORQUE, TENDS TO DISSECT THE ARTERY WITH EASE. PATIENT PRESENTED WITH AORTIC DISSECTION WHEN WIRE WAS USED.

    Patient problems: Aortic Dissection

    MAUDE report 6851091 ↗
  • InjurySep 8, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT IS REPORTED THAT THE WIRE IS HARD, BAD TORQUE, TENDS TO DISSECT THE ARTERY WITH EASE. PATIENT PRESENTED WITH AORTIC DISSECTION WHEN WIRE WAS USED. WIRE DISCARDED.

    Patient problems: Aortic Dissection

    MAUDE report 6851084 ↗

Other WIRE, GUIDE, CATHETER devices

All WIRE, GUIDE, CATHETER devices →

Sources and updates