Skip to content

INFINITY™ Occipitocervical Upper Thoracic System G3603514Recalls & reported problems

Medtronic Sofamor Danek, Inc. · MULTI AXIAL SCREW G3603514 3.5 X 14MM · Class 2 · Implant

All models of INFINITY Occipitocervical Upper Thoracic System (35)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169542242), received between Aug 11, 2023 and Mar 12, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

5 other recalls cover devices cleared under the same FDA submission (K163375); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Material Twisted/Bent1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Bone Fracture(s)1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 12, 2025Material Twisted/Bent; Material Twisted/Bent

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING C3-4 LMS, C6-7 PS POSTERIOR FIXATION (2A-2B) FOR C5 FRACTURE DUE TO FALL INJURY. IT WAS REPORTED THAT THE SCREW WAS PREPARED FOR LMS PLACEMENT, AND WHEN IT WAS TRIED TO PLACE IT ON A SCREWDRIVER, THE SCREW OF THE PRODUCT DID NOT FIT STRAIGHT INTO THE SCREWDRIVER. IT DID NOT FIT AND WAS PLACED ASKEW/IN A SLANTED ANGLE NO MATTER HOW MANY ATTEMPTS WERE MADE, SO IT WAS REPLACED WITH A PRODUCT OF THE SAME MODEL NUMBER AND SIZE.  THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21581589 ↗
  • InjurySep 9, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    INFORMATION RECEIVED FROM HEALTHCARE PROVIDER VIA MANUFACTURER REPRESENTATIVE REGARDING AN EVENT HAPPENED DURING POST-OP OF THE REPORTED PRODUCTS. IT WAS REPORTED THAT, THERE WAS LOOSENING OF THE RIGHT C6 LATERAL MASS SCREW (LMS) AND C7 PEDICLE SCREW (PS) AFTER THE INITIAL OPERATION. REVISION SURGERY WAS PERFORMED TO EXPLANT THE RIGHT C6 LMS AND C7 PEDICLE SCREW. DURING THE REVISION, C6 BILATERAL PEDICLE SCREWS WERE IMPLANTED, AND RIGHT C7 PEDICLE SCREW WAS SIZED UP. (F3.5/24 HAS BEEN CHANGED TO F4.5/26). PROCEDURE OR TECHNIQUE PERFORMED DURING INITIAL SURGERY WAS META C6-TH4 FIXATION. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20167555 ↗
  • InjuryAug 11, 2023Use of Device Problem; Use of Device Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR CERVI CAL FUSION FOR CERVICAL SPONDYLOTIC MYELOPATHY. IT WAS REPORTED THAT, LATERAL MASS OF VERTEBRAE WAS BROKEN WHILE IMPLANTING SCREW. THE STATUS OF THE SCREW WAS EXPLANTED-COMPLETE. TO IMPROVE POSTOPERATIVE SCREW FIXATION, THE 3.5MM SCREW WAS REMOVED AND REPLACED WITH A 4.5MM SCREW. THERE IS A POSSIBILITY THAT AN IMMENSE FORCE WAS EXERTED ON THE SCREW WHICH MAY CAUSE THE LATERAL MASS BROKEN. THERE WERE NO PATIENT SYMPTOMS REPORTED. NO ADDITIONAL TREATMENT OR SURGERY PERFORMED AS A RESULT. NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: Bone Fracture(s)

    MAUDE report 17508305 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K163375. One submission often covers many models, so check the product and lot details in each record.

  • Z-1412-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion
  • Z-1411-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, PARTIALLY THREADED MULTI AXIAL SCREW, Size: 4.0mm x 22mm, REF 360PT4022; for immobilization and stabilization of spinal segment as
  • Z-1413-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 34mm, REF 3604034; for immobilization and stabilization of spinal segment as an adjunct to fusion
  • Z-1409-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 24mm, REF 3604024; for immobilization and stabilization of spinal segment as an adjunct to fusion
  • Z-1410-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 32mm, REF 3604532; for immobilization and stabilization of spinal segment as an adjunct to fusion

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page