INFINITY™ Occipitocervical Upper Thoracic System G3603524Recalls & reported problems

Medtronic Sofamor Danek, Inc. · MULTI AXIAL SCREW G3603524 3.5 X 24MM · Class 2 · Implant

All models of INFINITY Occipitocervical Upper Thoracic System (35)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169542297), received between Sep 9, 2024 and Jul 2, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

5 other recalls cover devices cleared under the same FDA submission (K163375); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Malposition of Device1
  • Use of Device Problem1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1
  • Loss of consciousness1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 2, 2025Use of Device Problem; Malposition of Device; Use of Device Problem; Malposition of Device

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER REGARDING A PATIENT HAVING SPINAL THERAPY FOR CERVICAL FRACTURE. IT WAS REPORTED THAT DURING A C2-C7 SPINAL FIXATION PROCEDURE, A POTENTIAL MISALIGNMENT OF THE NAVIGATION SYSTEM (O-ARM) RESULTED IN THE LEFT PEDICLE SCREW AT C4 BEING MISPLACED BY 5 MM OR MORE, POSING A RISK OF VERTEBRAL ARTERY (VA) DAMAGE. DUE TO THE HIGH RISK OF HEMORRHAGING IF REPOSITIONED, THE SURGICAL TEAM DECIDED TO LEAVE THE SCREW IN PLACE AND PROCEEDED WITH WOUND CLOSURE. POST-OPERATIVELY, THE PATIENT DID NOT REGAIN CONSCIOUSNESS AND WAS TRANSFERRED TO THE WARD. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; Loss of consciousness

    MAUDE report 22381682 ↗
  • InjurySep 9, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    INFORMATION RECEIVED FROM HEALTHCARE PROVIDER VIA MANUFACTURER REPRESENTATIVE REGARDING AN EVENT HAPPENED DURING POST-OP OF THE REPORTED PRODUCTS. IT WAS REPORTED THAT, THERE WAS LOOSENING OF THE RIGHT C6 LATERAL MASS SCREW (LMS) AND C7 PEDICLE SCREW (PS) AFTER THE INITIAL OPERATION. REVISION SURGERY WAS PERFORMED TO EXPLANT THE RIGHT C6 LMS AND C7 PEDICLE SCREW. DURING THE REVISION, C6 BILATERAL PEDICLE SCREWS WERE IMPLANTED, AND RIGHT C7 PEDICLE SCREW WAS SIZED UP. (F3.5/24 HAS BEEN CHANGED TO F4.5/26). PROCEDURE OR TECHNIQUE PERFORMED DURING INITIAL SURGERY WAS META C6-TH4 FIXATION. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20167594 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K163375. One submission often covers many models, so check the product and lot details in each record.

  • Z-1409-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 24mm, REF 3604024; for immobilization and stabilization of spinal segment as an adjunct to fusion
  • Z-1413-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 34mm, REF 3604034; for immobilization and stabilization of spinal segment as an adjunct to fusion
  • Z-1410-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 32mm, REF 3604532; for immobilization and stabilization of spinal segment as an adjunct to fusion
  • Z-1411-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, PARTIALLY THREADED MULTI AXIAL SCREW, Size: 4.0mm x 22mm, REF 360PT4022; for immobilization and stabilization of spinal segment as
  • Z-1412-2023 · Feb 16, 2023 · Open, Classified
    INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page