RIALTO™ SI FUSION SYSTEM 7426004Recalls & reported problems
Medtronic Sofamor Danek, Inc. · DRIVER 7426004 RIALTO T27 THREADED · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169572584), received between Jan 29, 2025 and Feb 3, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 3, 2025Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT WHILE USING THE DRIVER TO REMOVE RIALTO SCREWS FROM THE PATIENT, ONE DRIVER STARTED TO TWIST AND THE OTHER DRIVER'S TIP BROKE OFF. THE TIP WAS RETRIEVED FROM THE PATIENT. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21289590 ↗MalfunctionJan 31, 2025Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED DURING AN UNKNOWN SPINAL PROCEDURE. IT WAS REPORTED THAT THE SCREWDRIVER TIP BROKE OFF. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21280969 ↗MalfunctionJan 29, 2025Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT PRESENTED WITH BACK PAIN. IT WAS REPORTED THAT WHILE ATTEMPTING TO REMOVE RIALTO SCREWS/CAGES FROM THE PATIENTS SACRUM THE ADUSTMENT DRIVER WAS BROKEN. HE THEN ASKED FOR THE T27 THREADED DRIVER AND IT WAS ALSO BROKEN. THE SURGEON DECIDED TO REMOVE THE 6 X BILATERAL RIALTO SCREWS THAT HAD BEEN IMPLANTED ON (B)(6) 2017. THERE WAS NOTHING WRONG WITH THE RIALTO IMPLANTS AND THEY SEEMED SOLIDLY FUSED AND WERE FULL OF BONE. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21262466 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page