HANCOCK II ULTRA® T505U225 (UDI 00643169594593)Recalls & reported problems

Medtronic, Inc. · VLV T505U225 HKII AORULT CII NNA US PEEK · Class 3 · Implant

All models of HANCOCK II ULTRA (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00643169594593), received between Nov 2, 2017 and Sep 4, 2025 — classified as 7 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

4 other recalls cover devices cleared under the same FDA submission (P980043); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information3
  • Intravalvular regurgitation2
  • Adverse Event Without Identified Device or Use Problem1
  • Device Stenosis1
  • Gradient Increase1
  • Material Split, Cut or Torn1
  • Patient Device Interaction Problem1

Patient problems

  • Insufficient Information3
  • Aortic Valve Stenosis2
  • Non specific EKG/ECG Changes2
  • Aortic Valve Insufficiency/ Regurgitation1
  • Cusp Tear1
  • High Blood Pressure/ Hypertension1
  • Occlusion1
  • Pulmonary Valve Insufficiency/ Regurgitation1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjurySep 4, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT ON THE SAME DAY AS THE IMPLANT OF THIS HANCOCK II VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 22962096 ↗
  • InjuryAug 18, 2025Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 3 MONTHS POST IMPLANT OF THIS HANCOCK II VALVE IN THE PULMONARY POSITION, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS REGURGITATION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Pulmonary Valve Insufficiency/ Regurgitation

    MAUDE report 22825652 ↗
  • InjuryJul 11, 2025Device Stenosis; Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 8 MONTHS POST IMPLANT OF THIS HANCOCK II AORTIC VALVE, THE PATIENT WAS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 2 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 22480011 ↗
  • InjuryJul 9, 2025Insufficient Device Problem Information; Material Split, Cut or Torn; Intravalvular regurgitation; Insufficient Device Problem Information; Material Split, Cut or Torn; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT THE 25MM HANCOCK II VALVE WAS EXPLANTED AND REPLACED WITH A NON MEDTRONIC VALVE. THE REASON FOR THE REPLACEMENT WAS REPORTED AS SEVERE AORTIC INSUFFICIENCY. IT WAS STATED THAT THERE WERE "SOME BURNED OUT DEBRIS ON THE AORTIC VALVE WITH ONE TORN LEAFLET". NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Cusp Tear; Insufficient Information

    MAUDE report 22455710 ↗
  • InjuryJul 20, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION FROM THE PATIENT THAT 2 YEARS AND 1 MONTH POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A NON-MEDTRONIC MECHANICAL VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 12199825 ↗
  • InjuryMay 19, 2021Gradient Increase; Insufficient Device Problem Information; Gradient Increase; Insufficient Device Problem Information

    \MEDTRONIC RECEIVED INFORMATION THAT 3 YEARS AND 5 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. PRIOR TO REPLACEMENT, THE PATIENT CALLED MEDTRONIC TO REPORT THAT THEIR BIOPROSTHETIC VALVE WAS "NOT WORKING PROPERLY". NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis; Insufficient Information

    MAUDE report 11849625 ↗
  • InjuryNov 2, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 25 MM BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH A 23 MM BIOPROSTHETIC VALVE OF A DIFFERENT MODEL. THE BLOOD PRESSURE WAS ELEVATED AND ELECTROCARDIOGRAM (ECG) SHOWED ELEVATED ST SEGMENTS. NO FAILURE MECHANISM OF THE 25 MM BIOPROSTHETIC VALVE WAS REPORTED. THE SURGEON REPORTED THE 25 MM VALVE OCCLUDED THE RIGHT CORONARY OSTIA AND THE SAME EFFECTIVE ORIFICE AREA (EOA) WAS ACCOMPLISHED WITH THE 23 MM VALVE. FOLLOWING THE IMPLANT OF THE 23 MM BIOPROSTHETIC VALVE, NO FURTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: High Blood Pressure/ Hypertension; Non specific EKG/ECG Changes; Occlusion

    MAUDE report 6997301 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980043. One submission often covers many models, so check the product and lot details in each record.

  • Z-1884-2019 · May 21, 2019 · Terminated
    Hancock II T510 Bioprosthesis, Mitral, REF T510C27 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1883-2019 · May 21, 2019 · Terminated
    Hancock II T510 Bioprosthesis, Mitral, REF T510C27 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1882-2019 · May 21, 2019 · Terminated
    Hancock II T510 Bioprosthesis, Mitral, REF T505U227 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1226-2009 · Jan 26, 2009 · Terminated
    Hancock II Bioprostheses with Cinch, Heart Valve, Model # T505 & T510; Catalog #s T505C2501, T505C2101, T505U25ID , T505C2301, T505C2301, T505C2501, T505C2101, T510C29ID, T510C25ID

Other replacement Heart-valve devices

All replacement Heart-valve devices →