Attain Performa™ MRI SureScan™ 429888Recalls & reported problems

Medtronic, Inc. · LEAD 429888 MRI CANT US · Class 3 · Implant

All models of Attain Performa MRI SureScan (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Attain Performa MRI lead is an implanted wire that carries electrical pulses between a heart pacemaker and the heart. Its 3,351 reports mostly describe the lead moving out of place, the heart not responding to pacing pulses, and leads being removed and replaced after infection or erosion. The reports include 38 deaths and 2,283 injuries. No recalls name this device. Reports were lower in the most recent year (299) than in the year before (356).

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
3,377
Deaths reported
38
Injuries reported
2,299
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3,377 adverse event reports that list this exact device (DI 00643169601833), received between Dec 9, 2016 and Sep 30, 2026 — classified as 38 deaths, 2,299 injuries, 1,040 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (691 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 36 reportsDec 2023: 47 reportsJan 2024: 38 reportsFeb 2024: 32 reportsMar 2024: 36 reportsApr 2024: 36 reportsMay 2024: 24 reportsJun 2024: 31 reportsJul 2024: 32 reportsAug 2024: 25 reportsSep 2024: 26 reportsOct 2024: 30 reportsNov 2024: 37 reportsDec 2024: 24 reportsJan 2025: 47 reportsFeb 2025: 21 reportsMar 2025: 29 reportsApr 2025: 18 reportsMay 2025: 29 reportsJun 2025: 29 reportsJul 2025: 35 reportsAug 2025: 32 reportsSep 2025: 28 reportsOct 2025: 27 reportsNov 2025: 36 reportsDec 2025: 36 reportsJan 2026: 20 reportsFeb 2026: 42 reportsMar 2026: 40 reportsApr 2026: 33 reportsMay 2026: 31 reportsJun 2026: 23 reportsJul 2026: 13 reportsAug 2026: 23 reportsSep 2026: 28 reportsOct 2026: 0 reports20242025202647
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1,290
  • High Capture Threshold691
  • Device Dislodged or Dislocated535
  • Failure to Capture397
  • Capturing Problem249
  • High impedance167
  • Activation, Positioning or Separation Problem165
  • Appropriate Device Problem Term/Code Not Available154

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,167
  • Unspecified Infection732
  • No Known Impact Or Consequence To Patient501
  • Undesired Nerve Stimulation218
  • Muscle Stimulation184
  • Device Overstimulation of Tissue181
  • No Consequences Or Impact To Patient91
  • Erythema57

Recent adverse event reports

The 25 most recent of 3,377 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Mechanical Problem; Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952730 ↗
  • InjurySep 29, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50935486 ↗
  • InjurySep 29, 2026Failure to Capture

    Patient problems: Chest Pain; Syncope/Fainting

    MAUDE report 50932892 ↗
  • MalfunctionSep 25, 2026Failure to Capture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903701 ↗
  • InjurySep 24, 2026Extracardiac stimulation

    It was reported that the Left Ventricular (LV) lead was "turned off since it is giving too much power that the patient can feel it". No further patient complications have been reported as a result of this event.

    Patient problems: Inappropriate Device Stimulation of Tissue

    MAUDE report 50885621 ↗
  • InjurySep 23, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that the Cardiac Resynchronization Therapy Pacemaker (CRT-P) system was explanted for an unknown/unspecified reason. A new system was implanted a few days later. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50871925 ↗
  • InjurySep 23, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that shortly after the Cardiac Resynchronization Therapy Defibrillator (CRT-D) implant, the patient experienced a large hematoma on the left side for which incision and drainage was performed on the left side. It was suspected that subsequently the patient developed infection with weakness, bacteremia, erosion and organism identified as Methicillin sensitive staphylococcus aureus (MSSA). The CRT-D system was explanted. No further patient complications have been reported as a result of this event.

    Patient problems: Bacteremia; Bacterial Infection; Erosion; Muscle Weakness/Atrophy

    MAUDE report 50869978 ↗
  • MalfunctionSep 23, 2026High Capture Threshold

    It was reported that the Left Ventricular (LV) lead exhibited chronic high thresholds. The LV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50866040 ↗
  • MalfunctionSep 18, 2026Impedance Problem

    It was reported that the Left Ventricular (LV) lead exhibited Out of Range (OOR) impedance measurement. The lead is still in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50819273 ↗
  • InjurySep 17, 2026Device Dislodged or Dislocated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50810089 ↗
  • InjurySep 17, 2026Extracardiac stimulation

    Patient problems: Inappropriate Device Stimulation of Tissue

    MAUDE report 50809104 ↗
  • InjurySep 16, 2026Device Dislodged or Dislocated; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50795300 ↗
  • InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced infection. The Cardiac Resynchronization Therapy Pacemaker (CRT-P) system was explanted and replaced with a Leadless implantable pulse generator (IPG). No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50767789 ↗
  • InjurySep 14, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50763627 ↗
  • InjurySep 14, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted due to pocket infection with erosion, redness and discomfort. No further patient complications have been reported as a result of this event.

    Patient problems: Discomfort; Erosion; Erythema; Medical device site infection

    MAUDE report 50757385 ↗
  • MalfunctionSep 11, 2026Capturing Problem; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50741374 ↗
  • MalfunctionSep 11, 2026High Capture Threshold

    It was reported that the Left Ventricular (LV) lead exhibited high capture thresholds. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50734997 ↗
  • MalfunctionSep 10, 2026High Capture Threshold

    It was reported that the Left Ventricular (LV) lead exhibited higher than programmed capture thresholds. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50716189 ↗
  • MalfunctionSep 9, 2026Failure to Capture; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50698336 ↗
  • MalfunctionSep 4, 2026Impedance Problem

    It was reported that the device was unable to be programmed to Magnetic Resonance Imaging (MRI) mode due to Out of Range (OOR) impedance values from the leads. The Right Ventricular (RV) lead and Left Ventricular (LV) lead remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50616304 ↗
  • InjurySep 3, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Medical device site infection

    MAUDE report 50597595 ↗
  • MalfunctionSep 3, 2026Capturing Problem; Impedance Problem

    Patient problems: Dyspnea; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50595596 ↗
  • MalfunctionSep 3, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50595373 ↗
  • MalfunctionSep 3, 2026High impedance

    It was reported that the Left Ventricular (LV) lead exhibited lead impedance that was high and undefined. The LV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50585034 ↗
  • InjurySep 2, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: Thrombosis/Thrombus

    MAUDE report 50574332 ↗

Other permanent pacemaker Electrode devices