SelectSecure™ MRI SureScan™ 383069 (UDI 00763000130800)Recalls & reported problems

Medtronic, Inc. · LEAD 383069 MRI US · Class 3 · Implant

Quick summary

This Medtronic SelectSecure MRI is a pacemaker electrode lead placed in the heart. Reports most often describe the lead needing too much power to pace the heart, and some describe leads moving out of place or infections. The device appears in 4,829 reports, including 56 deaths and 3,070 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
4,850
Deaths reported
56
Injuries reported
3,069
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4,850 adverse event reports that list this exact device (DI 00763000130800), received between Dec 26, 2018 and Sep 30, 2026 — classified as 56 deaths, 3,069 injuries, 1,725 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (1,522 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 61 reportsDec 2023: 47 reportsJan 2024: 53 reportsFeb 2024: 51 reportsMar 2024: 61 reportsApr 2024: 51 reportsMay 2024: 32 reportsJun 2024: 34 reportsJul 2024: 43 reportsAug 2024: 34 reportsSep 2024: 45 reportsOct 2024: 50 reportsNov 2024: 31 reportsDec 2024: 18 reportsJan 2025: 35 reportsFeb 2025: 30 reportsMar 2025: 38 reportsApr 2025: 22 reportsMay 2025: 44 reportsJun 2025: 29 reportsJul 2025: 45 reportsAug 2025: 29 reportsSep 2025: 24 reportsOct 2025: 32 reportsNov 2025: 32 reportsDec 2025: 33 reportsJan 2026: 26 reportsFeb 2026: 35 reportsMar 2026: 29 reportsApr 2026: 30 reportsMay 2026: 16 reportsJun 2026: 31 reportsJul 2026: 11 reportsAug 2026: 24 reportsSep 2026: 32 reportsOct 2026: 0 reports20242025202661
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Capture Threshold1,522
  • Failure to Capture729
  • Device Dislodged or Dislocated725
  • Adverse Event Without Identified Device or Use Problem709
  • Over-Sensing584
  • Under-Sensing575
  • Capturing Problem504
  • Decreased Sensitivity243

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,770
  • No Known Impact Or Consequence To Patient469
  • Unspecified Infection467
  • No Consequences Or Impact To Patient207
  • Bradycardia151
  • Dizziness132
  • Dyspnea107
  • Syncope/Fainting90

Recent adverse event reports

The 25 most recent of 4,850 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Decreased Sensitivity; Capturing Problem; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high and rising thresholds as well as diminished sensimg. It was further reported that the Right Atrial (RA) lead exhibited decreased pacing. Both leads remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50944794 ↗
  • MalfunctionSep 30, 2026Over-Sensing

    It was reported that the Right Ventricular (RV) Left Bundle Branch (LBB) lead exhibited T-Wave Oversensing (TWOS). The RV LBB lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943191 ↗
  • MalfunctionSep 29, 2026Failure to Capture; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937171 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Discomfort; Unspecified Infection

    MAUDE report 50933800 ↗
  • MalfunctionSep 29, 2026Low impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932374 ↗
  • MalfunctionSep 28, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926349 ↗
  • InjurySep 28, 2026Under-Sensing; Low impedance; Impedance Problem

    It was reported that the Right Atrial (RA) lead exhibited low impedance. The lead was explanted and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50912842 ↗
  • MalfunctionSep 24, 2026High Capture Threshold

    It was reported that there was a failed Atrial Lead Position Check on the Right Atrial (RA) lead and high thresholds on the Right Ventricular (RV) lead. The RV and RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884052 ↗
  • InjurySep 24, 2026High impedance

    It was reported that the His bundle branch Right Ventricular (RV-His) lead exhibited high out of range impedance warnings. The lead was explanted and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883576 ↗
  • MalfunctionSep 24, 2026Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited undersensing. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50880140 ↗
  • InjurySep 23, 2026High Capture Threshold

    It was reported that the Left Bundle Branch (LBB) Right Ventricular (RV) lead exhibited high thresholds. The lead was removed and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869739 ↗
  • InjurySep 18, 2026High impedance; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high thresholds and high impedance. The was explanted and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50815933 ↗
  • MalfunctionSep 17, 2026Break

    It was reported that the Right Ventricular (RV) lead returned for analysis and tested out-of-specification. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50802761 ↗
  • MalfunctionSep 17, 2026Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited under sensing. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50800795 ↗
  • MalfunctionSep 16, 2026Low impedance

    It was reported that the Left Ventricular (LV) lead exhibited low impedance. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50788138 ↗
  • MalfunctionSep 15, 2026Over-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited T-Wave Oversensing (TWOS) on the current electrograms (EGM). The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50770591 ↗
  • MalfunctionSep 15, 2026Over-Sensing

    It was reported that the patient experienced bradycardia due to T-Wave Oversensing (TWOS) that is occurring on the current electrograms (EGM). A polarity switch and high and undefined impedance warnings were also noted on the Right Atrial (RA) lead. The Right Ventricular Left Bundle Branch (RV-LBB) lead and RA lead remain in use. No further patient complications have been reported as a result of this event.

    Patient problems: Bradycardia

    MAUDE report 50770542 ↗
  • MalfunctionSep 15, 2026Break

    It was reported that the Right Ventricular (RV) lead returned for analysis and tested out-of-specification. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50769653 ↗
  • InjurySep 15, 2026High Capture Threshold

    It was reported that the Right Ventricular (RV) left bundle branch lead exhibited high thresholds leading to premature battery depletion of the Implantable Pulse Generator (IPG). The lead was capped and replaced and the device was explanted and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50769169 ↗
  • MalfunctionSep 14, 2026High impedance; Capturing Problem; Impedance Problem; High Capture Threshold

    It was reported that the left bundle branch (LBB) lead exhibited rising thresholds, increasing impedance and high impedance and high thresholds. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50752346 ↗
  • InjurySep 11, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that the pacing lead exhibited an unknown performance issue and was not functioning within normal tolerances. The pacing lead was explanted and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50730731 ↗
  • MalfunctionSep 10, 2026Failure to Capture; Pacing Problem; Under-Sensing

    It was reported that the patient attended the emergency room as they experienced shortness of breath. Upon device assessment it was noted that the HIS BUNDLE Right Ventricular (RV) lead exhibited under sensing, resulting in over pacing. Loss of capture was noted. The HIS BUNDLE RV lead remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Dyspnea; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50709652 ↗
  • MalfunctionSep 9, 2026Device Sensing Problem

    It was reported that the Implantable Pulse Generator (IPG) exhibited abnormal sensing. The IPG remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50688444 ↗
  • MalfunctionSep 8, 2026Intermittent Capture

    Patient problems: Palpitations

    MAUDE report 50677354 ↗
  • MalfunctionSep 8, 2026Over-Sensing; Capturing Problem

    It was reported that the Right Ventricular (RV) ¿ HIS bundle lead exhibited an increase in short V-V intervals on the Sensing Integrity Counter (SIC) and higher thresholds were noted. The RV-HIS lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50663943 ↗

Other permanent pacemaker Electrode devices