SelectSecure™ MRI SureScan™ 383069 (UDI 00763000130800)Recalls & reported problems

Medtronic, Inc. · LEAD 383069 MRI US · Class 3 · Implant

All models of SelectSecure MRI SureScan (9)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic SelectSecure MRI is a pacemaker electrode lead placed in the heart. Reports most often describe the lead needing too much power to pace the heart, and some describe leads moving out of place or infections. The device appears in 4,829 reports, including 56 deaths and 3,070 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
4,836
Deaths reported
56
Injuries reported
3,060
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4,836 adverse event reports that list this exact device (DI 00763000130800), received between Dec 26, 2018 and Sep 30, 2026 — classified as 56 deaths, 3,060 injuries, 1,720 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (1,517 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 47 reportsJan 2024: 53 reportsFeb 2024: 51 reportsMar 2024: 61 reportsApr 2024: 51 reportsMay 2024: 32 reportsJun 2024: 34 reportsJul 2024: 43 reportsAug 2024: 34 reportsSep 2024: 45 reportsOct 2024: 50 reportsNov 2024: 31 reportsDec 2024: 18 reportsJan 2025: 35 reportsFeb 2025: 30 reportsMar 2025: 38 reportsApr 2025: 22 reportsMay 2025: 44 reportsJun 2025: 29 reportsJul 2025: 45 reportsAug 2025: 29 reportsSep 2025: 24 reportsOct 2025: 32 reportsNov 2025: 32 reportsDec 2025: 33 reportsJan 2026: 26 reportsFeb 2026: 35 reportsMar 2026: 29 reportsApr 2026: 29 reportsMay 2026: 16 reportsJun 2026: 27 reportsJul 2026: 11 reportsAug 2026: 22 reportsSep 2026: 25 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202661
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Capture Threshold1,517
  • Failure to Capture729
  • Device Dislodged or Dislocated725
  • Adverse Event Without Identified Device or Use Problem708
  • Over-Sensing585
  • Under-Sensing574
  • Capturing Problem503
  • Decreased Sensitivity242

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,757
  • No Known Impact Or Consequence To Patient469
  • Unspecified Infection466
  • No Consequences Or Impact To Patient207
  • Bradycardia151
  • Dizziness132
  • Dyspnea107
  • Syncope/Fainting90

Recent adverse event reports

The 25 most recent of 4,836 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Over-Sensing

    It was reported that the Right Ventricular (RV) Left Bundle Branch (LBB) lead exhibited T-Wave Oversensing (TWOS). The RV LBB lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50943191 ↗
  • MalfunctionSep 29, 2026Failure to Capture; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937171 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced discomfort and an infection. The Implantable Pulse Generator (IPG) system was explanted and replaced with a leadless IPG. No further patient complications have been reported as a result of this event.

    Patient problems: Discomfort; Unspecified Infection

    MAUDE report 50933800 ↗
  • MalfunctionSep 28, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926349 ↗
  • MalfunctionSep 24, 2026High Capture Threshold

    It was reported that there was a failed Atrial Lead Position Check on the Right Atrial (RA) lead and high thresholds on the Right Ventricular (RV) lead. The RV and RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884052 ↗
  • MalfunctionSep 24, 2026Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited undersensing. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50880140 ↗
  • InjurySep 23, 2026High Capture Threshold

    It was reported that the Left Bundle Branch (LBB) Right Ventricular (RV) lead exhibited high thresholds. The lead was removed and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50869739 ↗
  • InjurySep 18, 2026High impedance; High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high thresholds and high impedance. The was explanted and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50815933 ↗
  • MalfunctionSep 17, 2026Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited under sensing. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50800795 ↗
  • MalfunctionSep 15, 2026Over-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited T-Wave Oversensing (TWOS) on the current electrograms (EGM). The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50770591 ↗
  • MalfunctionSep 15, 2026Over-Sensing

    It was reported that the patient experienced bradycardia due to T-Wave Oversensing (TWOS) that is occurring on the current electrograms (EGM). A polarity switch and high and undefined impedance warnings were also noted on the Right Atrial (RA) lead. The Right Ventricular Left Bundle Branch (RV-LBB) lead and RA lead remain in use. No further patient complications have been reported as a result of this event.

    Patient problems: Bradycardia

    MAUDE report 50770542 ↗
  • MalfunctionSep 15, 2026Break

    It was reported that the Right Ventricular (RV) lead returned for analysis and tested out-of-specification. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50769653 ↗
  • MalfunctionSep 14, 2026High impedance; Capturing Problem; Impedance Problem; High Capture Threshold

    It was reported that the left bundle branch (LBB) lead exhibited rising thresholds, increasing impedance and high impedance and high thresholds. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50752346 ↗
  • MalfunctionSep 10, 2026Failure to Capture; Pacing Problem; Under-Sensing

    It was reported that the patient attended the emergency room as they experienced shortness of breath. Upon device assessment it was noted that the HIS BUNDLE Right Ventricular (RV) lead exhibited under sensing, resulting in over pacing. Loss of capture was noted. The HIS BUNDLE RV lead remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Dyspnea; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50709652 ↗
  • MalfunctionSep 9, 2026Device Sensing Problem

    It was reported that the Right Ventricular HIS Bundle (RV-HIS) lead exhibited abnormal sensing. The RV-HIS remains in use. No patient complications have been reported as a result of this event. The patient is a participant in the Post Approval Clinical Surveillance Product Surveillance Registry.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50688444 ↗
  • MalfunctionSep 8, 2026Intermittent Capture

    Patient problems: Palpitations

    MAUDE report 50677354 ↗
  • MalfunctionSep 8, 2026Over-Sensing; Capturing Problem

    It was reported that the Right Ventricular (RV) ¿ HIS bundle lead exhibited an increase in short V-V intervals on the Sensing Integrity Counter (SIC) and higher thresholds were noted. The RV-HIS lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50663943 ↗
  • MalfunctionSep 8, 2026High Capture Threshold

    It was reported that the HIS BUNDLE Right Ventricular (RV) lead exhibited  intermittent high thresholds. The Right Atrial (RA) lead also exhibited Far Field R-Wave (FFRW) oversensing. Both leads remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50663784 ↗
  • MalfunctionSep 7, 2026Pacing Problem; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50645966 ↗
  • InjurySep 2, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that the patient experienced deep vein thrombus (DVT). The patient was administered anticoagulant medication that resolved the DVT. The Right Atrial (RA) lead, Right Ventricular (RV) left bundle branch lead and Left Ventricular (LV) lead remained in use. No further patient complications have been reported as a result of this event.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 50574250 ↗
  • InjurySep 2, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that the patient experienced left shoulder pain, severe pain in left neck, upper chest pain, facial swelling, redness, vein distension and certain movements of head and neck cause pain, that improved with muscle relaxant. The patient also indicated when bent over there is head pressure, becomes very red which takes a while to resolve. The patient also reports prominent veins in left upper chest since pacemaker was implanted. A venogram was performed that revealed occluded subclavian vein thus the patient experienced Superior Vena Cava (SVC) Syndrome. A SVC venoplasty was performed and the Right Atrial (RA) lead, Right Ventricular (RV) left bundle branch lead were explanted and replaced. The Left Ventricular (LV) lead was explanted. No further patient complications have been reported as a result of this event.

    Patient problems: Chest Pain; Erythema; Neck Pain; Obstruction/Occlusion; Pain; Swelling/ Edema

    MAUDE report 50573535 ↗
  • MalfunctionSep 2, 2026Failure to Capture; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50562564 ↗
  • MalfunctionSep 2, 2026Over-Sensing

    Patient problems: Bradycardia; No Clinical Signs, Symptoms or Conditions

    MAUDE report 50559063 ↗
  • InjurySep 1, 2026Failure to Capture; Over-Sensing; Under-Sensing; Capturing Problem

    It was reported that the patient experienced asystole and the physician thought the patient was in a torsade's ventricular event that they are thinking was ischemic event. The Right Ventricular (RV) Left Bundle Branch (LBB) lead exhibited non capture, a rise in thresholds which has been relatively consistent last two weeks, short v-v intervals with some double counting of fibrillation waves and some undersensing. The patient was defibrillated by the emergency team and had to move to temporary pacing  The RV LBB lead was reprogrammed and remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Arrhythmia; Asystole; Ischemia

    MAUDE report 50542698 ↗
  • MalfunctionSep 1, 2026Pacing Problem; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50541182 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P030036 on Aug 3, 2005. Submitted by Medtronic Cardiac Rhythm Disease Management as “Medtronic Selectsecure Lead Model 3830”. The FDA review took 699 days. The FDA has since approved 159 supplements (changes to the device or its labeling), most recently on Oct 16, 2026. FDA record ↗

The FDA regulates product code DTB (permanent pacemaker electrode) under 21 CFR 870.3680 as a Class III device. 21 CFR 870.3680 ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Common questions about the SelectSecure™ MRI SureScan™ 383069

Has the SelectSecure™ MRI SureScan™ 383069 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00763000130800), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SelectSecure™ MRI SureScan™ 383069?

The FDA's MAUDE database lists 4,836 adverse event reports for this device, received between Dec 26, 2018 and Sep 30, 2026. In those reports, the most common specific device problems are “High Capture Threshold” (1,517), “Failure to Capture” (729) and “Device Dislodged or Dislocated” (725); the most common patient problems are “Unspecified Infection” (466), “Bradycardia” (151) and “Dizziness” (132). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SelectSecure™ MRI SureScan™ 383069?

56 reports are classified as deaths, 3,060 as injuries and 1,720 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SelectSecure™ MRI SureScan™ 383069 FDA approved?

It was approved through premarket approval (PMA) P030036 on Aug 3, 2005.

Is the SelectSecure™ MRI SureScan™ 383069 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the SelectSecure™ MRI SureScan™ 383069, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand SelectSecure™ MRI SureScan™, model 383069). Its record shows it as in commercial distribution.