Visualase 9735559 (UDI 00643169611986)Recalls & reported problems
Medtronic Navigation, Inc. · VCLAS 9735559 .4MM CORE FIBER 3MM TIP · Class 2
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Mar 23, 2018Terminated
Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.
Z-2501-2018 · Root cause: Equipment maintenance · FDA record ↗
Other Neurosurgical laser with MR thermography devices
- Visualase 9735560Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography1 recall59 reports
- Visualase 9735560Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography1 recall4 reports
- Visualase 9735559Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography1 recall3 reports
- Visualase 9735560Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography1 recall3 reports
- Visualase 9735560Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography44 reports
- Visualase 9735559Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography15 reports
- Visualase 9735542Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography15 reports
- Visualase 9735559Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography10 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.