Visualase recallZ-2501-2018 · Medtronic Navigation, Inc. · initiated Mar 23, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Navigation, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.
FDA root cause category: Equipment maintenance
What the company is doing
Customers were notified on about 03/23/2018 via URGENT: MEDICAL DEVICE RECALL letter. Instructions included to examine inventory for affected devices, contact Medtronic Technical Services at 1-888-826-5603 to arrange for the return of affected devices, and complete and return the Consignee Response Form. The recall was later expanded and customers were notified on about 05/16/2018 URGENT: MEDICAL DEVICE RECALL letter. Instructions did not change from the initial letter, but informed customers that the recall had been expanded to include two (2) additional lots.
Recall details
- Recall number
- Z-2501-2018
- Status
- Terminated
- Initiated by the company
- Mar 23, 2018
- Recalling firm
- Medtronic Navigation, Inc.
- Quantity
- 1,283 devices total
- Product code
- GEX
- FDA submissions
- K053087
- Distribution
- Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI. International distribution to Israel.
Product description
Visualase Cooled Laser Applicator System (VCLAS) labeled as: a. Part Number 9735559, Visualase Cooled Laser Applicator System 3mm Tip; b. Part Number 9735560 Visualase Cooled Laser Applicator System 10mm Tip; c. Part Number 9735561 Visualase Cooled Laser Applicator System 15mm Tip Product Usage: The Visualase Thermal Therapy System is indicated for use to necrotize or coagulate soft tissue through interstitial irradiation or thermal therapy under magnetic resonance imaging (MRI) guidance in medicine and surgery in cardiovascular thoracic surgery (excluding the heart and the vessels in the pericardial sac), dermatology, ear-nose-throat surgery, gastroenterology, general surgery, gynecology, head and neck surgery, neurosurgery, plastic surgery, orthopedics, pulmonology, radiology, and urology, for wavelengths 800 nm - 1060 nm. When therapy is performed under MRI guidance, and when data from compatible MRI sequences is available, the Visualase system can process images to determine relative changes in tissue temperature during therapy. The image data may be manipulated and viewed in a number of different ways, and the values of data at certain selected points may be monitored and/or displayed over time.
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
a. Part Number 9735559, UDI (00643169611986, 00643169933361), Lot Codes (0212782059, 0212894256, 0213251294, 0213331774, 0213383266, 0212783784, 0212926008, 0213259125, 0213335148, 0213383267, 0212823883, 0212956163, 0213299882, 0213375479, 0213387215, 0212823919, 0212978141, 0213299885, 0213378982, 0213537710, 0212854585, 0212996201, 0213327285, 0213383244, 0213966003, 0212882619, 0213519510); b. Part Number 9735560, UDI (00643169611993, 00643169933385), Lot Codes (0212086315, 0213158249, 0213299883, 0213427083, 0213728326, 0212698642, 0213158631, 0213304570, 0213455321, 0213735817, 0212715926, 0213158719, 0213304571, 0213458938, 0213737565, 0212745633, 0213165510, 0213304573, 0213507022, 0213760863, 0212753561, 0213165511, 0213308717, 0213507023, 0213760864, 0212756095, 0213205023, 0213309087, 0213512428, 0213760866, 0212760821, 0213205024, 0213334598, 0213512430, 0213769912, 0212827856, 0213208868, 0213368255, 0213513054, 0213917334, 0212838837, 0213212692, 0213368499, 0213513055, 0213922188, 0212843810, 0213223891, 0213368573, 0213513056, 0213924407, 0212846727, 0213223892, 0213373657, 0213540031, 0213924418, 0212856596, 0213242839, 0213374559, 0213576515, 0213928892, 0212873863, 0213245723, 0213375535, 0213576516, 0213945813, 0212873873, 0213258625, 0213375537, 0213576517, 0213945814, 0212879480, 0213259039, 0213375538, 0213594271, 0213966370, 0212913269, 0213259673, 0213375991, 0213594272, 0213972422, 0212922387, 0213259674, 0213376135, 0213594273, 0213973527, 0212935278, 0213267742, 0213376136, 0213596471, 0214032331, 0213000097, 0213289058, 0213376138, 0213699385, 0214043408, 0213001255, 0213289059, 0213376139, 0213699386, 0214045295, 0213009483, 0213292765, 0213376655, 0213699390, 0214049428, 0213041517, 0213292893, 0213376656, 0213710007, 0214062101, 0213044327, 0213296890, 0213376780, 0213710711, 0214066559, 0213056030, 0213296891, 0213377536, 0213710712, 0214068011, 0213107317, 0213298978, 0213378061, 0213720928, 0214068013, 0213116115, 0213299342, 0213387201, 0213720929, 0214068921, 0213134866, 0213299525, 0213388160, 0213728324, 0214068922, 0213136534, 0213518754, 0213138278); c. Part Number 9735561, UDI (00643169612006, 00763000015695, 00643169933378), Lot Codes (0212701679, 0212912501, 0212991152, 0213647807, 0213669275, 0212707035, 0212935757, 0213555835, 0213662428, 0213669402, 0212745660, 0212939348, 0213555836, 0213663333, 0213669408, 0212796559, 0212963182, 0213645747, 0213663500, 0213670155, 0212846726, 0212965542, 0213645760, 0213667299, 0213670722, 0212901580, 0212904407)
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- Visualase 9735560Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography1 recall4 reports
- Visualase 9735560Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography1 recall3 reports
- Visualase 9735559Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography1 recall3 reports
- Visualase 9735561Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall1 report
- Visualase 9735559Medtronic Navigation, Inc. · Neurosurgical laser with MR thermography1 recall0 reports
- Visualase 9735561Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall0 reports
- Visualase 9735561Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page