Skip to content

Contegra® 200SH20 (UDI 00643169616875)Recalls & reported problems

Medtronic, Inc. · CNDT 200SH20 CONTEGRA SUPP HDE · Class f · Implant

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00643169616875), received between Mar 24, 2017 and Jun 2, 2026 — classified as 7 injuries.

The most frequently reported specific device problem is “Gradient Increase” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Gradient Increase2
  • Insufficient Device Problem Information2
  • Intravalvular regurgitation2
  • Device Stenosis1

Patient problems

  • Non specific EKG/ECG Changes2
  • Pulmonary Valve Insufficiency/ Regurgitation2
  • Pulmonary Valve Stenosis2
  • Adhesion(s)1
  • Insufficient Information1
  • No Information1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 2, 2026Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 11 MONTHS POST IMPLANT OF THIS 20MM PULMONARY VALVED CONDUIT IMPLANTED IN A THEN 8 YEAR OLD PEDIATRIC PATIENT, A TRANSCATHETER PULMONARY BIOPROSTHETIC VALVE (TPBV) WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR REPLACEMENT WAS REPORTED AS SEVERE PULMONARY REGURGITATION (PR). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Pulmonary Valve Insufficiency/ Regurgitation

    MAUDE report 25360952 ↗
  • InjuryOct 20, 2025Gradient Increase; Device Stenosis; Intravalvular regurgitation; Gradient Increase; Device Stenosis; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND POST IMPLANT OF THIS 20MM CONTEGRA PULMONARY VALVED CONDUIT IMPLANTED IN A THEN 4 YEAR OLD PEDIATRIC PATIENT, IT WAS REPLACED WITH A TRANSCATHETER VALVE. THE REASON FOR REPLACEMENT WAS REPORTED AS MODERATE PULMONARY STENOSIS, MODERATE PULMONARY REGURGITATION AND A MEAN GRADIENT OF 24MMHG. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Pulmonary Valve Insufficiency/ Regurgitation; Pulmonary Valve Stenosis

    MAUDE report 23342832 ↗
  • InjuryJun 3, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS AND 11 MONTHS POST IMPLANT OF THIS 20 MM PULMONARY VALVED CONDUIT IN A 11YR OLD PEDIATRIC PATIENT, A TRANSCATHETER PULMONARY BIOPROSTHETIC VALVE (TPBV) WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 14598550 ↗
  • InjuryOct 23, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT 11 DAYS POST IMPLANT OF THIS 20MM PULMONARY VALVED CONDUIT THIS PATIENT RECEIVED A PERMANENT PACEMAKER DUE TO ATRIO-VENTRICULAR (AV) BLOCK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 9224918 ↗
  • InjuryJul 1, 2019Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 2 YEARS AND 3 MONTHS POST IMPLANT OF THIS 20MM BIOPROSTHETIC PULMONARY VALVED CONDUIT IN A PEDIATRIC PATIENT, IT WAS EXPLANTED AND REPLACED WITH A BIOPROSTHETIC PULMONARY VALVED CONDUIT OF THE SAME SIZE AND MODEL FOR UNKNOWN REASONS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Adhesion(s); No Information

    MAUDE report 8749380 ↗
  • InjuryApr 25, 2018Gradient Increase; Gradient Increase

    MEDTRONIC RECEIVED INFORMATION THAT 1 YEAR AND 1 MONTH POST IMPLANT OF THIS PULMONARY VALVED CONDUIT IN A PEDIATRIC PATIENT, A TRANSCATHETER PULMONARY BIOPROSTHETIC VALVE (TPBV) WAS IMPLANTED VALVE-IN-VALVE DUE TO STENOSIS AND HIGH GRADIENTS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Pulmonary Valve Stenosis

    MAUDE report 7462127 ↗
  • InjuryMar 24, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    MEDTRONIC RECEIVED INFORMATION THAT 8 DAYS POST IMPLANT OF THIS PULMONARY BIOPROSTHETIC VALVED CONDUIT A PERMANENT PACEMAKER IMPLANT WAS REQUIRED. NO FAILURE MECHANISM WAS PROVIDED. NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 6432210 ↗

Other pulmonic valved conduit devices

All pulmonic valved conduit devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.