CD HORIZON® SOLERA® Spinal System 651003070Recalls & reported problems
Medtronic Sofamor Danek, Inc. · ROD 651003070 5.5 CCM CAPPED ROD 70MM · Class 2 · Implant
All models of CD HORIZON SOLERA Spinal System (34)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169748040), received between Sep 28, 2021 and Sep 28, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth1
- Mechanical Problem1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Post Operative Wound Infection1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2021Mechanical Problem; Loosening of Implant Not Related to Bone-Ingrowth; Mechanical Problem; Loosening of Implant Not Related to Bone-Ingrowth
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING PATIENT FOR UNKNOWN SPINAL THERAPY. IT WAS REPORTED THAT THERE IS POSSIBLE INFECTION. THERE WAS PATIENT INVOLVED IN THE EVENT AND NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE HEAD OF SCREW WAS MOBILE AND SET SCREW LOOSE. THE PEDICLE SCREW WAS SOLID WITHIN THE PEDICLE AND THE ROD HAD AN INDENTATION SIGNIFYING PROPER FINAL TIGHTENING. FAILURE BETWEEN HEAD OF SCREW, ROD AND SET SCREW CAUSING LOOSENING AND A METALLOSIS EFFECT. THE LOT NUMBERS OF PRODUCT ID: 5540030 IS H5453997, PRODUCT ID: 55840007550 IS H59226 AND PRODUCT ID: 651003070 IS 0587500W. INFECTION CAUSE IS UNKNOWN. HOWEVER, METALLOSIS WAS A CAUSE FOR CONCERN. EXACT DATE OF IMPLANTATION IS UNKNOWN. STATED TO BE ROUGHLY 2 ¿ 2 ½ YEARS AGO.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Post Operative Wound Infection
MAUDE report 12538872 ↗
Other Thoracolumbosacral pedicle screw system devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page