CD HORIZON® SOLERA® Spinal System 651000040Recalls & reported problems
Medtronic Sofamor Danek, Inc. · ROD 651000040 5.5 CCM PERC ROD 40MM · Class 1 · Implant
All models of CD HORIZON SOLERA Spinal System (34)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169748521), received between Sep 22, 2026 and Sep 22, 2026 — classified as 2 injuries.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 22, 2026Use of Device Problem
Information was received from healthcare provider (HCP) via a manufacturer representative regarding a Patient having Open surgery Spinal therapy for degenerative spine condition. It was reported that during the procedure, difficult rod placement was encountered. The loaner set intended for the Voyager 4.75 mm system had been replenished with Voyager 5.5 mm implants, and wrong-diameter rods were placed. The surgeon was unable to lock the system with the locking nuts. The planned minimally invasive surgery was converted to open surgery, and the wound was opened on one side to remove the incorrect rods. The incorrect rods were explanted and replaced. Surgical time was extended by approximately 1 hour. There were unknown patient symptoms or complications reported. Additional information received from the manufacturer representative that there was no patient symptoms reported. Additional information received from the manufacturer representative that the levels involved was L5-S1. The quantity of the reported products were 2 minimal invasive rods 5.5.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50850147 ↗InjurySep 22, 2026Use of Device Problem
Information was received from healthcare provider (HCP) via a manufacturer representative regarding a Patient having Open surgery Sp inal therapy for degenerative spine condition. It was reported that during the procedure,difficult rod placement was encountered. The loaner set intended for the Voyager 4.75 mm system had been replenished with Voyager 5.5 mm implants, and wrong-diameter rods were placed. The surgeon was unable to lock the system with the locking nuts. The planned minimally invasive surgery was converted to open surgery, and the wound was opened on one side to remove the incorrect rods. The incorrect rods were explanted and replaced. Surgical time was extended by approximately 1 hour. There were unknown patient symptoms or complications reported. Additional information received from the manufacturer representative that there was no patient symptoms reported. Additional information received from the manufacturer representative that the levels involved was L5-S1. The quantity of the reported products were 2 minimal invasive rods 5.5.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50850145 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page