Skip to content

CD HORIZON® SOLERA® VOYAGER™ Spinal System 55750047535Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SCREW 55750047535 ST 5/6 SV ATS 7.5X35 · Class 2 · Implant

All models of CD HORIZON SOLERA VOYAGER Spinal System (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169749085), received between Nov 14, 2022 and Nov 14, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Migration1
  • Use of Device Problem1

Patient problems

  • Unspecified Nervous System Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 14, 2022Use of Device Problem; Migration; Use of Device Problem; Migration

    IT WAS REPORTED THAT AFTER THE PROCEDURE THE SCREWS WERE DEVIATED. THERE IS A POSSIBILITY OF L5 ROOT SYMPTOMS AFTER SURGERY AND PROLONGED HOSPITALIZATION TO FIX AGAIN THROUGH REVISION SURGERY. NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED. ADDITIONAL INFORMATION HAS BEEN RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT ALL THE REPORTED SCREWS WERE DEVIATED AFTER INITIAL SURGERY AND WERE EXPLANTED IN THE REVISION SURGERY. THE LOT NUMBERS OF THE REPORTED PRODUCTS ARE UNKNOWN. ADDITIONAL INFORMATION HAS BEEN RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT AS A RESULT OF THE VERIFICATION, IT BECAME CLEAR THAT IT WAS A HUMAN ERROR. IN A LATERAL DECUBITUS POSITION, A CAGE WAS INSERTED USING THE OLIF AT THIS HOSPITAL, AND THEN THE PERCUTANEOUS PEDICLE SCREW (PPS) WAS PERFORMED AS IS IN A LATERAL DECUBITUS POSITION. UPON NAVIGATION AFTER THE CAGE INSERTION, THE VERTEBRAL BODY MOVES DUE TO THE CAGE, SO THE O-ARM IMAGING WAS PERFORMED AGAIN NAVIGATION WAS USED; ACCIDENTALLY SELECTED THE O-ARM IMAGING DATA BEFORE INSERTING THE CAGE ON THE NAVIGATION EVEN DURING PPS, AND THE SCREW WAS INSERTED UNNOTICED AS IS. THEREFORE, IT WAS DECIDED THAT EVEN THOUGH THE SCREW INSERTION WAS ACCURATE ON THE N

    Patient problems: Unspecified Nervous System Problem

    MAUDE report 15786854 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page