Skip to content

CareLink Encore™ 29901ARecalls & reported problems

Medtronic, Inc. · PRGM 29901A ENCORE US EIFU · Class 3

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic CareLink Encore programmer is used to check implanted heart devices. Its 475 adverse event reports often describe the programmer freezing, crashing, or showing blank screens. There are no deaths and 3 injuries. Reports rose from 4 in the prior 12 months to 18 in the last 12. There are 2 recalls, 1 open, over security weaknesses in its telemetry, including improper access control and unencrypted data transmission.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
474
Deaths reported
0
Injuries reported
3
Recalls naming it
2

Summary

The FDA's MAUDE database contains 474 adverse event reports that list this exact device (DI 00643169763548), received between Apr 6, 2017 and May 11, 2026 — classified as 3 injuries, 471 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (223 reports).

2 FDA recall records name this device's identifier, 1 still open.

10 other recalls cover devices cleared under the same FDA submission (P890003); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 4 reportsJan 2024: 3 reportsFeb 2024: 7 reportsMar 2024: 8 reportsApr 2024: 3 reportsMay 2024: 8 reportsJun 2024: 1 reportsJul 2024: 4 reportsAug 2024: 5 reportsSep 2024: 5 reportsOct 2024: 5 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 7 reportsMay 2026: 6 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 30, 2020Open, Classified

    Medtronic Conexus Telemetry has been determined to contain two primary cyber vulnerabilities: improper access control and the cleartext transmission of sensitive information.

    Z-1525-2020 · Root cause: Software design · FDA record ↗

  2. Jan 17, 2019Terminated

    The Medtronic CareLink Encore 29901 programmer system is comprised of prescription devices indicated for use in the interrogation and programming of implantable medical devices. Prior to use, refer to the Programmer Reference Guide as well as the appropriate programmer software and implantable device technical manuals for more information related to specific implantable device models. Programming should be attempted only by appropriately trained personnel after careful study of the technical manual for the implantable device and after careful determination of appropriate parameter values based on the patient's condition and pacing system used. The Medtronic CareLink Encore 29901 programmer must be used only for programming implantable devices manufactured by Medtronic or Vitatron.

    Z-0796-2019 · Root cause: Software design · FDA record ↗

Most reported problems

Device problems

  • No Display/Image223
  • Device Difficult to Program or Calibrate137
  • Computer Operating System Problem108
  • Failure to Calibrate70
  • Computer Software Problem54
  • Unexpected Shutdown48
  • Display or Visual Feedback Problem32
  • Battery Problem28

Patient problems

  • No Clinical Signs, Symptoms or Conditions328
  • No Patient Involvement75
  • No Known Impact Or Consequence To Patient61
  • No Consequences Or Impact To Patient9
  • Cardiac Arrest1
  • Dizziness1
  • Syncope1

Recent adverse event reports

The 25 most recent of 474 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 11, 2026Computer Software Problem; Operating System Becomes Nonfunctional

    IT WAS REPORTED THAT THE PROGRAMMER CRASHED DURING THE INTERROGATION OF A PATIENT¿S IMPLANTABLE DEVICE AND DISPLAYED AN ERROR MESSAGE. IT WAS NOTED THAT THE SCREEN WENT BLACK AND THEN DISPLAYED AN ERROR. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25137402 ↗
  • MalfunctionMay 7, 2026Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE PROGRAMMER BECAME UNRESPONSIVE AND FROZE WHEN AN ERROR MESSAGE APPEARED ON THE SCREEN WHILE ATTEMPTING TO SAVE DATA TO A FLASH DRIVE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25108393 ↗
  • MalfunctionMay 6, 2026Computer Software Problem; Operating System Becomes Nonfunctional

    IT WAS REPORTED THAT THE PROGRAMMER CRASHED DURING A PATIENT SESSION. IT WAS NOTED THAT AN ERROR MESSAGE INDICATING SERIOUS SOFTWARE PROBLEM WAS DISPLAYED AND KICKED OUT OF PATIENT SESSION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25091705 ↗
  • MalfunctionMay 5, 2026Computer Software Problem; No Display/Image; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE PROGRAMMER GENERATED AN ERROR MESSAGE INDICATING A SERIOUS SOFTWARE PROBLEM. IT WAS NOTED THAT WHEN THE PRO GRAMMER WAS TURNED ON TO DISABLE THE AUTO-PRINTING, IT FROZE AT THE FIND MODEL AND FIND PATIENT SCREEN. IT WAS UNABLE TO NAVIGATE TO ANYTHING ON THE SCREEN, AND THE PROGRAMMER BECAME UNRESPONSIVE TO FINGER TOUCH OR THE STYLUS PEN. TROUBLESHOOTING STEPS WERE TAKEN, INCLUDING REBOOTING THE PROGRAMMER AND TAKING THE BATTERY OUT, WHICH DID NOT RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25083316 ↗
  • MalfunctionMay 5, 2026Computer Software Problem; Failure to Interrogate; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE PROGRAMMER FROZE AT FORMATTING REPORTS AND UNABLE TO PROCEED FURTHER WITH INTERROGATION. IT WAS ALSO NOTED THAT THE PROGRAMMER GENERATED A SOFTWARE ERROR AFTER BOOT UP WHEN ATTEMPTED INTERROGATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25076999 ↗
  • MalfunctionMay 4, 2026Computer Software Problem; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE USER REPORTED DIFFICULTY USING THE PROGRAMMER DURING A DEVICE FOLLOW-UP SESSION. THE USER SELECTED THE APPROPRIATE DEVICE MODEL AND INITIATED INTERROGATION, WHICH WAS SUCCESSFULLY COMPLETED. HOWEVER, THE PROGRAMMER DISPLAYED A PERSISTENT MESSAGE AT THE TOP OF THE SCREEN STATING, ¿SERIOUS SOFTWARE PROBLEM.¿ AN ADDITIONAL POP-UP MESSAGE APPEARED STATING, ¿QUICK LOOK FOLLOW UP GENERATING COMPLETE REPORT,¿ AT WHICH POINT THE SYSTEM FROZE. THE PROGRAMMER SUBSEQUENTLY RESUMED NORMAL FUNCTION AFTER A PERIOD OF TIME. IT WAS NOTED A SEPARATE MESSAGE WAS DISPLAYED INDICATING, ¿INCONSISTENT DATE AND TIME SETTINGS HAVE BEEN DETECTED ON THE PROGRAMMER. IT WAS CONFIRMED PROGRAMMER DATE AND TIME WERE ACCURATE. TROUBLESHOOTING STEPS WERE TAKEN TO CHECK THE SOFTWARE VERSION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25066106 ↗
  • MalfunctionApr 21, 2026Computer Software Problem; Operating System Becomes Nonfunctional

    IT WAS REPORTED THAT THE PROGRAMMER CRASHED WITH THE ERROR "SERIOUS SOFTWARE PROBLEM." TROUBLESHOOTING STEPS WERE PERFORMED, INCLUDING CHANGES TO SETTINGS, AND IT WAS CONFIRMED THAT IT WORKED AND WAS ABLE TO SAVE THE PORTABLE DIGITAL DOCUMENT (PDD) FILE TO SEND TO A DATA MANAGEMENT SYSTEM. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24941936 ↗
  • MalfunctionApr 16, 2026Computer Software Problem; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE PROGRAMMER DISPLAYED A ¿SERIOUS SOFTWARE PROBLEM¿ ERROR MESSAGE. IT WAS NOTED THAT THE PROGRAMMER PROMPTED THAT INCONSISTENT DATE AND TIME SETTINGS WERE DETECTED; HOWEVER, THE DATE AND TIME WERE CONFIRMED TO BE CORRECT. IT WAS FURTHER NOTED THAT DURING INTERROGATION, THE PROGRESS BAR COMPLETED, BUT THE PROGRAMMER REMAINED ON THE FORMATTING REPORT SCREEN FOR SEVERAL MINUTES AND SUBSEQUENTLY DISPLAYED A BLACK SCREEN WITH A SYSTEM ERROR MESSAGE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24894353 ↗
  • MalfunctionApr 15, 2026Computer Software Problem; Communication or Transmission Problem; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE PROGRAMMER SCREEN WAS STUCK, THEN WENT TO A BLACK SCREEN DISPLAYING A SYSTEM ERROR WHEN IT WAS TRYING TO FORMAT REPORTS WHILE INTERROGATING, AN IMPLANTABLE DEVICE. IT WAS NOTED THAT WHEN THE SCREEN CAME BACK ON, IT DISPLAYED AN ERROR MESSAGE INDICATING "SERIOUS SOFTWARE PROBLEM." NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24883941 ↗
  • MalfunctionApr 10, 2026Computer Software Problem; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE PROGRAMMER¿S SCREEN WAS FROZEN WITH A POP UP MESSAGE DISPLAYED INDICATING "FORMATTING REPORT, PLEASE WAIT." THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24842685 ↗
  • MalfunctionApr 10, 2026Data Problem; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE PROGRAMMER FROZE WHILE GENERATING A REPORT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24842645 ↗
  • MalfunctionApr 9, 2026Unstable Capture Threshold; Application Program Freezes or Fails to Launch

    IT WAS REPORTED THAT THE PROGRAMMER WAS UNABLE TO RUN A THRESHOLD TEST AND SCREEN WAS FROZEN ON QUICK LOOK FOLLOW-UP REPORT. TROUBLESHOOTING STEPS WERE PERFORMED BY POWER CYCLING, REMOVED THE BACK UP BATTERY AND REMOVED THE POWER SUPPLY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24827451 ↗
  • MalfunctionApr 9, 2026Computer Software Problem; Operating System Becomes Nonfunctional

    IT WAS REPORTED THAT THE PROGRAMMER CRASHED DURING AN ACTIVE PATIENT SESSION AND DISPLAYED ERROR MESSAGE INDICATING SERIOUS PROGRAMMER PROBLEM. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24824739 ↗
  • MalfunctionMar 26, 2026Electrical Shorting

    IT WAS REPORTED THAT THE PROGRAMMER'S POWER PLUG WAS BROKEN, AND THE PROGRAMMER HAD A SHORT CIRCUIT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698007 ↗
  • MalfunctionMar 17, 2026Overheating of Device; Date/Time-Related Software Problem

    IT WAS REPORTED THAT THE PROGRAMMER DISPLAYED AN OVERHEATING MESSAGE AND THAT THE DATE WAS INCORRECT. TROUBLESHOOTING STEPS WERE PROVIDED FOR RESOLVING THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24615018 ↗
  • MalfunctionJan 5, 2026Unexpected Shutdown

    IT WAS REPORTED THAT THE PROGRAMMER REBOOTED WHILE CHECKING WITH PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23955900 ↗
  • MalfunctionJan 5, 2026Unexpected Shutdown

    IT WAS REPORTED THAT THE PROGRAMMER REBOOTED WHILE CHECKING WITH PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23955886 ↗
  • MalfunctionAug 29, 2025Computer Operating System Problem; Computer Operating System Problem

    IT WAS REPORTED THAT THE PROGRAMMER'S SCREEN FROZE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22920822 ↗
  • MalfunctionJul 24, 2025Computer Software Problem; Failure to Interrogate; Computer Operating System Problem; Computer Software Problem; Failure to Interrogate; Computer Operating System Problem

    IT WAS REPORTED THAT THE PROGRAMMER SCREEN FROZE AND BECAME UNRESPONSIVE WHILE TRYING TO INTERROGATE. IT WAS NOTED THAT AN ERROR MESSAGE WAS GENERATED ALONG WITH A BLACK SCREEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22585236 ↗
  • MalfunctionDec 4, 2024Computer Operating System Problem; Computer Operating System Problem

    IT WAS REPORTED THAT THE PROGRAMMER'S SCREEN WAS FROZEN AND RADIOFREQUENCY (RF) HEAD LIGHTS WERE RED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20841064 ↗
  • MalfunctionNov 5, 2024Computer Operating System Problem; Problem with Software Installation; Computer Operating System Problem; Problem with Software Installation

    IT WAS REPORTED THAT THE PROGRAMMER'S SCREEN FROZE WHEN TRYING TO UPDATE THE SOFTWARE USING THE UNIVERSAL SERIAL BUS (USB). TROUBLESHOOTING STEPS WERE TAKEN TO REBOOT THE PROGRAMMER AND THE USB WAS PLUGGED IN, AFTER WHICH THE UPDATES CONTINUED INSTALLING. HOWEVER, THE SOFTWARE UPDATE WAS NOT SUCCESSFUL. THE PROGRAMMER HAS BEEN RETURNED FOR EVALUATION AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20608599 ↗
  • MalfunctionOct 29, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE PROGRAMMER HAD TOUCHSCREEN ISSUES. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20556347 ↗
  • MalfunctionOct 29, 2024No Display/Image; No Display/Image

    IT WAS REPORTED THAT THE PROGRAMMER HAD TOUCHSCREEN ISSUES. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20556275 ↗
  • MalfunctionOct 10, 2024No Display/Image; Computer Operating System Problem; No Display/Image; Computer Operating System Problem

    IT WAS REPORTED THAT THE PROGRAMMER FROZE ON START UP. THE PROGRAMMER HAS BEEN RETURNED FOR EVALUATION AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20414039 ↗
  • MalfunctionOct 4, 2024No Display/Image; Display or Visual Feedback Problem; No Display/Image; Display or Visual Feedback Problem

    IT WAS REPORTED THAT THE PROGRAMMER DISPLAYED A BLACK SCREEN UPON START UP, AFTER THE PROGRAMMER WAS LEFT IDLE AND DURING INTERROGATION CHECKS. IT WAS NOTED THAT THE PROGRAMMER SOFTWARE HAD BEEN UPDATED AND THE ISSUE HAD PERSISTED. THE PROGRAMMER HAS BEEN RETURNED FOR EVALUATION AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20366949 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P890003. One submission often covers many models, so check the product and lot details in each record.

  • Z-2168-2025 · Jun 18, 2025 · Open, Classified
    Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria
  • Z-2169-2025 · Jun 18, 2025 · Open, Classified
    Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria A
  • Z-2128-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
  • Z-2127-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
  • Z-2506-2020 · Jun 4, 2020 · Open, Classified
    Patient Connector, Model Number 24967
  • Z-2505-2020 · Jun 4, 2020 · Open, Classified
    CareLink SmartSync Device Manager, Model Number 24970A
  • Z-0845-2019 · Jan 17, 2019 · Terminated
    Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
  • Z-0400-2018 · Jul 14, 2017 · Terminated
    MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates
  • Z-0399-2018 · Jul 14, 2017 · Terminated
    MyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with
  • Z-2209-2010 · Jun 11, 2010 · Terminated
    Medtronic Carelink Monitor Model 2490C. The Medtronic CareLink Monitor Model 2490C is designed to automatically gather information from the implanted heart device. This automatic w

Other Pacemaker/icd/crt non-implanted components devices

All Pacemaker/icd/crt non-implanted components devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page