Gallant™ CDDRA500QRecalls & reported problems

St. Jude Medical, Inc. · DR Implantable Cardioverter Defibrillator VVED DDDR · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Gallant implantable cardioverter defibrillator is a device implanted to treat heart rhythm problems. Its 2,997 adverse event reports often describe the device sensing electrical noise as heartbeats, losing Bluetooth connection, and shocks that were inappropriate or inadequate. The reports include 33 deaths and 835 injuries. Two open recalls name this device, both over Bluetooth problems that can prevent remote monitoring.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
3,028
Deaths reported
33
Injuries reported
840
Recalls naming it
2

Summary

The FDA's MAUDE database contains 3,028 adverse event reports that list this exact device (DI 05415067032003), received between Nov 10, 2020 and Sep 30, 2026 — classified as 33 deaths, 840 injuries, 2,153 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (864 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 58 reportsDec 2023: 58 reportsJan 2024: 52 reportsFeb 2024: 50 reportsMar 2024: 52 reportsApr 2024: 67 reportsMay 2024: 66 reportsJun 2024: 55 reportsJul 2024: 49 reportsAug 2024: 68 reportsSep 2024: 57 reportsOct 2024: 71 reportsNov 2024: 67 reportsDec 2024: 40 reportsJan 2025: 63 reportsFeb 2025: 62 reportsMar 2025: 55 reportsApr 2025: 45 reportsMay 2025: 67 reportsJun 2025: 54 reportsJul 2025: 43 reportsAug 2025: 52 reportsSep 2025: 42 reportsOct 2025: 40 reportsNov 2025: 50 reportsDec 2025: 55 reportsJan 2026: 59 reportsFeb 2026: 54 reportsMar 2026: 63 reportsApr 2026: 74 reportsMay 2026: 44 reportsJun 2026: 40 reportsJul 2026: 12 reportsAug 2026: 47 reportsSep 2026: 64 reportsOct 2026: 0 reports20242025202674
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 16, 2023Open, Classified

    Due to Bluetooth malfunction on a subset of Implantable Cardioverter Defibrillator (ICD) and Cardia Resynchronization Therapy Defibrillator *CRT-D) devices that were manufactured prior to April 2022 as a result of crystal component issue within the Bluetooth Low Energy (BLE) circuit. This results in the device entering an inductive telemetry only mode with loss of Bluetooth communication. In some instances, thought BLE communication is unavailable, the circuit may remain in a state which draws excessive power affecting battery longevity.

    Z-2614-2023 · Root cause: Component design/selection · FDA record ↗

  2. Mar 10, 2022Open, Classified

    Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

    Z-0988-2022 · Root cause: Component design/selection · FDA record ↗

Most reported problems

Device problems

  • Over-Sensing864
  • Failure to Interrogate640
  • Adverse Event Without Identified Device or Use Problem379
  • Inappropriate/Inadequate Shock/Stimulation326
  • Incorrect Interpretation of Signal229
  • Inappropriate or Unexpected Reset216
  • Under-Sensing145
  • Pacing Problem128

Patient problems

  • No Clinical Signs, Symptoms or Conditions2,403
  • Unspecified Infection321
  • Arrhythmia135
  • Insufficient Information111
  • Discomfort69
  • Pocket Erosion52
  • Wound Dehiscence32
  • Hematoma17

Recent adverse event reports

The 25 most recent of 3,028 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Failure to Interrogate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949907 ↗
  • MalfunctionSep 30, 2026Failure to Interrogate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949097 ↗
  • InjurySep 30, 2026Inappropriate/Inadequate Shock/Stimulation

    Patient problems: Insufficient Information

    MAUDE report 50948520 ↗
  • MalfunctionSep 30, 2026Failure to Interrogate

    It was reported that the implantable cardioverter defibrillator (ICD) exhibited Bluetooth (BLE) telemetry loss. Additional information received indicated the patient presented in clinic and the BLE was succesfully restored, however, no information was provided as to how this was achieved. There were no patient consequences reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947795 ↗
  • InjurySep 30, 2026Inappropriate/Inadequate Shock/Stimulation

    It was reported that the patient received an inappropriate shock from the Implantable Cardioverter Defibrillator (ICD). Further evaluation revealed that the right ventricular (RV) lead exhibited loss of capture, and a Computed Tomography (CT) scan confirmed cardiac perforation. At the patient's request, the entire system, including the ICD, right ventricular (RV) lead, and right atrial (RA) lead, was explanted on (b)(6)2026. The patient remained stable throughout the procedure.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941663 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient presented for a follow-up in the clinic with infection at the implanted device pocket site and pocket erosion. The implantable cardioverter defibrillator was visible from the opened incision site of the implanted device pocket. The physician explanted the entire system ¿ implantable cardioverter-defibrillator, right ventricular lead, and atrial lead due to resolve the issue. The patient was in stable condition.

    Patient problems: Pocket Erosion; Unspecified Infection

    MAUDE report 50933682 ↗
  • MalfunctionSep 29, 2026Under-Sensing

    It was reported that the patient presented remotely via merlin.net. Review of transmission revealed that the Implantable Cardioverter Defibrillator (ICD) exhibited under-sensing. No intervention was performed. The patient was in stable condition.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50928255 ↗
  • MalfunctionSep 29, 2026Environmental Compatibility Problem; Inappropriate or Unexpected Reset

    It was reported that the patient underwent to a Magnetic Resonance Imaging (MRI) procedure without enabling MRI mode. Later it was noted that the Implantable Cardioverter Defibrillator (ICD) was in backup mode with high voltage therapy disabled. Programming changes were made, and the ICD was restored. The patient was in stable condition.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927996 ↗
  • MalfunctionSep 28, 2026Failure to Interrogate

    It was reported that the implantable cardioverter defibrillator (ICD) exhibited Bluetooth (BLE) telemetry loss. No intervention was performed. There were no patient consequences reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915931 ↗
  • MalfunctionSep 26, 2026Over-Sensing

    Patient problems: Insufficient Information

    MAUDE report 50906530 ↗
  • MalfunctionSep 25, 2026Failure to Interrogate

    It was reported that the patient's Implantable Cardioverter Defibrillator (ICD) exhibited loss of Bluetooth telemetry. No intervention was performed. The patient had no consequences.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50900951 ↗
  • MalfunctionSep 25, 2026Failure to Interrogate

    It was reported that the patient's Implantable Cardioverter Defibrillator exhibited loss of Bluetooth telemetry. No intervention was performed. There were no patient consequences.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50897374 ↗
  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient presented to the hospital with drainage at the device implant site due to infection. The patient's implantable cardioverter-defibrillator, the right atrial lead and the right ventricular lead were explanted due to the infection. The patient was stable throughout.

    Patient problems: Unspecified Infection

    MAUDE report 50893288 ↗
  • MalfunctionSep 24, 2026Failure to Capture; Under-Sensing

    It was reported that the patient experienced an arrythmia. Subsequently, decreased sensing resulting in under-sensing and loss of capture were noted on the device. No intervention was performed. The patient was stable.

    Patient problems: Arrhythmia

    MAUDE report 50884663 ↗
  • MalfunctionSep 24, 2026Inappropriate or Unexpected Reset

    It was reported that the Implantable Cardioverter Defibrillator (ICD) went into backup mode. No intervention was reported and patient condition was not provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884533 ↗
  • InjurySep 24, 2026Incorrect Interpretation of Signal; Inappropriate/Inadequate Shock/Stimulation

    It was reported via a review of remote transmission that the device delivered an inappropriate shock due to an incorrect interpretation of signal. No patient symptoms were reported. It is unknown whether any changes were made.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50883907 ↗
  • InjurySep 23, 2026Insufficient Device Problem Information

    It was reported that the device was explanted and replaced due to an unspecified device problem. No further information was reported.

    Patient problems: Insufficient Information

    MAUDE report 50872954 ↗
  • MalfunctionSep 23, 2026Failure to Interrogate

    It was reported that a patient presented to the emergency room for unknown reasons. The Merlin on Demand transmitter was unable to perform reading on the patient's implantable cardioverter defibrillator (ICD). No changes or interventions were reported. The patient was discharged.

    Patient problems: Insufficient Information

    MAUDE report 50868808 ↗
  • MalfunctionSep 23, 2026Failure to Interrogate

    It was reported that the implantable cardioverter defibrillator (ICD) exhibited Bluetooth (BLE) telemetry loss. No intervention was performed. There were no patient consequences reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50868315 ↗
  • MalfunctionSep 23, 2026Over-Sensing; Pacing Problem

    During an in clinic follow up, far field R-wave oversensing and inappropriate automatic mode switch were noted on the device. No intervention has been performed. The patient was stable.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50867297 ↗
  • MalfunctionSep 19, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50831632 ↗
  • MalfunctionSep 18, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829944 ↗
  • InjurySep 18, 2026Capturing Problem; Device Sensing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829708 ↗
  • MalfunctionSep 18, 2026Failure to Interrogate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829490 ↗
  • InjurySep 18, 2026Inappropriate/Inadequate Shock/Stimulation

    It was reported that the patient presented remotely. A remote transmission showed that the patient was inappropriately shocked by their Implantable Cardioverter Defibrillator (ICD). No intervention was performed. The patient had no adverse consequences due to the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50828534 ↗

Other Pacemaker/icd/crt non-implanted components devices