CD HORIZON® SOLERA® VOYAGER™ Spinal System 55850014545Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SCREW 55850014545 5.5 VOYAGER MAS 4.5X45 · Class 1 · Implant

All models of CD HORIZON SOLERA VOYAGER Spinal System (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169768123), received between Jul 5, 2022 and Jul 5, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 5, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT PREVIOUSLY IMPLANTED WITH SCREWS IN A PERCUTANEOUS PSF AT T10-L2. IT WAS REPORTED THAT THE SCREWS MALFUNCTIONED SOMETIME BETWEEN IMPLANT AND REMOVAL, LIKELY DURING ONE OF THE PATIENT'S FALLS. UPON EXPLANTATION, THE HCP NOTICED THAT THE SCREWS TULIPS WERE WELDED TO THE SHANK IN A WAY THAT THE POLY AXIAL RANGE OF MOTION WAS LOST AND THE TULIP WAS ¿FIXATED¿ AT A DIFFERENT DEGREE FOR EACH OF THE RESPECTIVE SCREWS. THE SURGEON WAS UNABLE TO PUT THE SHANK OF THE DRIVER THROUGH THE TULIP TO SEAT IT INSIDE OF THE SCREW SHANK BECAUSE THE FIXATED TULIP WAS BLOCKING A STRAIGHT PATH WAY FOR THE DRIVER TO SEAT. THE MALFUNCTIONING OF THE SCREWS WAS LIKELY DUE TO A FALL. THE PATIENT IS OSTEOPOROTIC AND THE SCREWS WERE HALOING SEVERELY AND LOOSE. THE PATIENT HAS HAD MULTIPLE FALLS. THE HCP DETERMINED THAT THE SCREWS DID NOT STABILIZE THE PATIENT AS HE HAD HOPED. THE SURGEON DECIDED TO REMOVE HARDWARE AS PART OF HER PREVENTATIVE TREATMENT PLAN AND AT THE TIME OF THE HARDWARE REMOVAL, HE NOTICED THE SCREW MALFUNCTION. THERE IS NO INJURY TO THE PATIENT DUE TO THE SCREW MALFUNCTION. THE DATE OF

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14913664 ↗