Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169768284), received between Mar 23, 2021 and Jul 9, 2024 — classified as 5 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth3
  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Pain2
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 9, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED POST L4/5 PERC SCREWS/MIDLINE INCISION FOR INTERBODY IN A PATIENT DIAGNOSED WITH BACK AND LEG PAIN . IT WAS REPORTED THAT PATIENT PRESENTED FOR FOLLOW-UP COMPLAINING OF LEG PAIN. RESCANNED AND CAGE SHOWN TO HAVE BACKED OUT INTO CANAL. ON CT THE SCREWS HAVE ALSO OBSERVED TO HAVE LUCENCY AROUND THEM. SCREWS LOOSENED DUE TO THE LACK OF ANTERIOR COLUMN SUPPORT. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Pain

    MAUDE report 19699709 ↗
  • InjuryJul 18, 2022Break; Break

    INFORMATION RECEIVED FROM HEALTHCARE PROVIDER VIA MANUFACTURER REPRESENTATIVE REGARDING AN EVENT HAPPENED DURING POST OP OF THE REPO RTED PRODUCTS. IT WAS REPORTED THAT, SCREW HEAD WITH FEW THREADS LENGTH OF THE SHAFT HAS BEEN EXPLANTED. THE REST OF THE SHAFT HAS BEEN LEFT IN THE PATIENT. THE PROCEDURE INVOLVED WAS REVISION SURGERY TO REPLACE THE SCREW, AND INITIAL PROCEDURE PERFORMED WAS MIS L5/S1 INSTRUMENTED FUSION WITH RIGHT SIDED TLIF. THE LEVELS IMPLANTED DURING INITIAL SURGERY WAS LEFT S1. IT WAS UNKNOWN ABOUT THE PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT. NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 15037700 ↗
  • InjuryJul 5, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT PREVIOUSLY IMPLANTED WITH SCREWS IN A PERCUTANEOUS PSF AT T10-L2. IT WAS REPORTED THAT THE SCREWS MALFUNCTIONED SOMETIME BETWEEN IMPLANT AND REMOVAL, LIKELY DURING ONE OF THE PATIENT'S FALLS. UPON EXPLANTATION, THE HCP NOTICED THAT THE SCREWS TULIPS WERE WELDED TO THE SHANK IN A WAY THAT THE POLY AXIAL RANGE OF MOTION WAS LOST AND THE TULIP WAS ¿FIXATED¿ AT A DIFFERENT DEGREE FOR EACH OF THE RESPECTIVE SCREWS. THE SURGEON WAS UNABLE TO PUT THE SHANK OF THE DRIVER THROUGH THE TULIP TO SEAT IT INSIDE OF THE SCREW SHANK BECAUSE THE FIXATED TULIP WAS BLOCKING A STRAIGHT PATH WAY FOR THE DRIVER TO SEAT. THE MALFUNCTIONING OF THE SCREWS WAS LIKELY DUE TO A FALL. THE PATIENT IS OSTEOPOROTIC AND THE SCREWS WERE HALOING SEVERELY AND LOOSE. THE PATIENT HAS HAD MULTIPLE FALLS. THE HCP DETERMINED THAT THE SCREWS DID NOT STABILIZE THE PATIENT AS HE HAD HOPED. THE SURGEON DECIDED TO REMOVE HARDWARE AS PART OF HER PREVENTATIVE TREATMENT PLAN AND AT THE TIME OF THE HARDWARE REMOVAL, HE NOTICED THE SCREW MALFUNCTION. THERE IS NO INJURY TO THE PATIENT DUE TO THE SCREW MALFUNCTION. THE DATE OF

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14913679 ↗
  • InjuryJul 5, 2022Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT PREVIOUSLY IMPLANTED WITH SCREWS IN A PERCUTANEOUS PSF AT T10-L2. IT WAS REPORTED THAT THE SCREWS MALFUNCTIONED SOMETIME BETWEEN IMPLANT AND REMOVAL, LIKELY DURING ONE OF THE PATIENT'S FALLS. UPON EXPLANTATION, THE HCP NOTICED THAT THE SCREWS TULIPS WERE WELDED TO THE SHANK IN A WAY THAT THE POLY AXIAL RANGE OF MOTION WAS LOST AND THE TULIP WAS ¿FIXATED¿ AT A DIFFERENT DEGREE FOR EACH OF THE RESPECTIVE SCREWS. THE SURGEON WAS UNABLE TO PUT THE SHANK OF THE DRIVER THROUGH THE TULIP TO SEAT IT INSIDE OF THE SCREW SHANK BECAUSE THE FIXATED TULIP WAS BLOCKING A STRAIGHT PATH WAY FOR THE DRIVER TO SEAT. THE MALFUNCTIONING OF THE SCREWS WAS LIKELY DUE TO A FALL. THE PATIENT IS OSTEOPOROTIC AND THE SCREWS WERE HALOING SEVERELY AND LOOSE. THE PATIENT HAS HAD MULTIPLE FALLS. THE HCP DETERMINED THAT THE SCREWS DID NOT STABILIZE THE PATIENT AS HE HAD HOPED. THE SURGEON DECIDED TO REMOVE HARDWARE AS PART OF HER PREVENTATIVE TREATMENT PLAN AND AT THE TIME OF THE HARDWARE REMOVAL, HE NOTICED THE SCREW MALFUNCTION. THERE IS NO INJURY TO THE PATIENT DUE TO THE SCREW MALFUNCTION. THE DATE OF

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14913674 ↗
  • InjuryMar 23, 2021Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA FIELD REPRESENTATIVE REGARDING PATIENT WITH PRE-OP DIAGNOSIS- PAINFUL ORTH OPEDIC HARDWARE AFTER PROCEDURE INVOLVED IN MINIMALLY INVASIVE TECHNIQUES FOR TRANSFORAMINAL LUMBAR INTERBODY FUSION (MIS TLIF). LEVELS IMPLANTED- L5-S1. IT WAS REPORTED THAT DURING USE, THE PATIENT HEARD A POP IN THEIR BACK. REPORTED THE BREAKAGE WAS DUE TO THE WEIGHT OF THE PATIENT. PRODUCT WAS UTILIZED CORRECTLY ACCORDING TO THE DIRECTIONS GIVEN IN THE IFU/LABELING. REVISION SURGERY WAS PERFORMED TO EXPLANT THE PRODUCT, AS PATIENT HAD INCREASE BACK PAIN BECAUSE OF THE BROKEN SCREW. THERE WERE NO BROKEN FRAGMENTS LEFT IN THE PATIENT BODY. THERE WAS NO DELAY IN OVERALL PROCEDURE TIME. AFTER REVISION SURGERY, PATIENT IS RECOVERING. PATIENT IS ALIVE AND NO INJURY.

    Patient problems: Pain

    MAUDE report 11553695 ↗