CD HORIZON® SOLERA® VOYAGER™ Spinal System 55850017540Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 55850017540 5.5 VOYAGER MAS 7.5X40 · Class 1 · Implant
All models of CD HORIZON SOLERA VOYAGER Spinal System (60)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169768345), received between Apr 3, 2025 and Apr 24, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Mechanics Altered” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanics Altered1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 24, 2025Mechanics Altered; Mechanics Altered
INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (HEALTHCARE PROVIDER, OTHER) REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT DURING A LUMBAR FUSION PROCEDURE, A REDUCTION SCREW WAS INSERTED INTO THE SPINE'S VERTEBRAL BODY. THE SCREW IS EQUIPPED WITH A CLIP THAT IS INTENDED TO BE REMOVED ONCE THE SCREW IS SECURELY POSITIONED. HOWEVER, ONE OUT OF THE FOUR SCREWS USED HAD A CLIP THAT FAILED TO DETACH, RESULTING IN IT BEING RETAINED IN THE PATIENT AND CAUSING CONTINUED PAIN. THIS ISSUE WAS IDENTIFIED DURING A ROUTINE OFFICE VISIT ONE MONTH POST-OPERATION THROUGH AN X-RAY ON (B)(6) 2025. THE PATIENT IS SCHEDULED TO RETURN TO THE OPERATING ROOM TO HAVE THE CLIP REMOVED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: Pain
MAUDE report 21898144 ↗InjuryApr 3, 2025Use of Device Problem; Use of Device Problem
ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE STATING, PATIENT REPORTED BACK PAIN, BUT SURGEON DID NOT ATTRIBUTE IT TO RETAINED TAB. TAB WAS IDENTIFIED AT POST OP FOLLOW UP ROUGHLY 4 WEEKS POST OP DURING ROUTINE X-RAY. REVISION SURGERY WAS PERFORMED WITHOUT INCIDENT. NO FURTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21760068 ↗
Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices
- NA 511.701Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT2,164 reports
- SmartLife 7126-120-000Stryker Corporation · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT1,065 reports
- NA 05.001.224Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT987 reports
- NA 05.001.206Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT632 reports
- NA 532.021Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT521 reports
- NA 2570-04-501Depuy (Ireland) · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT500 reports
- NA 532.110Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT433 reports
- NA 05.001.223Synthes GmbH · INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT397 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page