Skip to content

CD HORIZON® SOLERA® VOYAGER™ Spinal System 55750015540Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SCREW 55750015540 5.5VOYAGER STMAS5.5X40 · Class 2 · Implant

All models of CD HORIZON SOLERA VOYAGER Spinal System (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00643169783768), received between Oct 14, 2019 and Apr 22, 2025 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Malposition of Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Malposition of Device4
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Unspecified Nervous System Problem4
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 22, 2025Malposition of Device; Malposition of Device

    ADDITIONAL INFORMATION WAS RECEIVED REGARDING THE EVENT. SURGERY WAS PERFORMED FOR LCS WITHOUT L5 RADICULAR SYMPTOMS. SCREW WAS INSERTED USING MAZOR AND THE SCREW DEVIATED INTO THE SPINAL CANAL AND RADICULAR SYMPTOMS OCCURRED. RE-OPERATION OCCURRED AND IT WAS RE-INSERTED BY FREEHAND. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Unspecified Nervous System Problem

    MAUDE report 21880710 ↗
  • InjuryApr 22, 2025Malposition of Device; Malposition of Device

    ADDITIONAL INFORMATION WAS RECEIVED REGARDING THE EVENT. SURGERY WAS PERFORMED FOR LCS WITHOUT L5 RADICULAR SYMPTOMS. SCREW WAS INSERTED USING MAZOR AND THE SCREW DEVIATED INTO THE SPINAL CANAL AND RADICULAR SYMPTOMS OCCURRED. RE-OPERATION OCCURRED AND IT WAS RE-INSERTED BY FREEHAND. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Unspecified Nervous System Problem

    MAUDE report 21880700 ↗
  • InjuryApr 22, 2025Malposition of Device; Malposition of Device

    ADDITIONAL INFORMATION WAS RECEIVED REGARDING THE EVENT. SURGERY WAS PERFORMED FOR LCS WITHOUT L5 RADICULAR SYMPTOMS. SCREW WAS INSERTED USING MAZOR AND THE SCREW DEVIATED INTO THE SPINAL CANAL AND RADICULAR SYMPTOMS OCCURRED. RE-OPERATION OCCURRED AND IT WAS RE-INSERTED BY FREEHAND. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Unspecified Nervous System Problem

    MAUDE report 21880673 ↗
  • InjuryMar 28, 2025Malposition of Device; Malposition of Device

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING L3/S2 SPINAL TH ERAPY FOR AN INDICATION OF LUMBAR CANAL STENOSIS (LCS). IT WAS REPORTED THAT THERE WAS S2SCREW MISALIGNMENT. THE SCREW WAS INSERTED ACCORDING TO THE PLAN USING THE MAZOR SYSTEM AND POSTOPERATIVE CT CONFIRMED THAT THE S2 SCREW WAS INSERTED ALONG THE SACROILIAC JOINT, WITH THE SCREW THREADS PROTRUDING INTO THE PELVIS. THE PATIENT PRESENTED NEUROLOGICAL SYMPTOMS AND THE ATTENDING PHYSICIAN'S OPINION IS THAT THE PATIENT'S POSTOPERATIVE NEUROLOGICAL SYMPTOMS ARE LIKELY DUE TO THE DEVIATION OF THE L5 SCREW WHICH WAS INSERTED FREEHAND BECAUSE THE MAZOR SYSTEM COULD NOT BE USED. THERE WAS A DELAY OF LESS THAN 60 MINS REPORTED IN OVERALL PROCEDURE TIME. THERE WAS EXTENSION OF EXISTING HOSPITALIZATION AS A RESULT OF THE EVENT AND THE PATIENT WAS SCHEDULED FOR REPEAT SURGERY ON (B)(6) 2025 FOR SCREW REPLACEMENT. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: Unspecified Nervous System Problem

    MAUDE report 21716558 ↗
  • InjuryJun 1, 2023Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A FUNNEL USED IN AN L4/5 CAGE REPLACEMENT REOPERATION (L4-S PPS, L4/5 TLIF, 5/S TLIF) FOR LCS, LEG PARALYSIS, AND NUMBNESS. IT WAS REPORTED THAT THE BONE GRAFT FUNNEL WAS FILLED WITH BONE IN AN ATTEMPT TO FILL THE INTERVERTEBRAL DISC WITH BONE GRAFT, BUT GOT STUCK AND WOULD NOT COME OUT. AFTER ATTEMPTS TO REMOVE THE BLOCKAGE WITH STEEL OBJECTS FROM THE HOSPITAL, THE PROBLEM PERSISTED AND THEREFORE WAS IDENTIFIED AS A FAILURE EVENT. THE FUNNEL WAS NOT USED ON THE PATIENT, EVEN THOUGH THE PATIENT WAS ON THE OPERATING TABLE WHEN THE MALFUNCTION OCCURRED. THERE WAS NO PATIENT IMPACT. THE L4/5 CAGE BACKED OUT DUE TO A SCREW LOOSENING ON THE RIGHT SIDE OF L4. THIS WAS A REPEAT SURGERY DUE TO SCREW LOOSENING AND CAGE BACK OUT. THE DATE OF INITIAL SURGERY WAS ON (B)(6) 2023. ON (B)(6) 2023, REPEAT SURGERY WAS PERFORMED. THE RIGHT L4 SCREW AND THE L4/5 CAGE WERE REMOVED AND REPLACED WITH A NEW ONE. THE CAUSE OF SCREW LOOSENING WAS THAT THE L4 RIGHT SCREW WAS PLACED OUTWARD FROM THE PEDICLE AND DEVIATED. NO FURTHER COMPLICATIONS WERE REPORTED/ ANTICIPATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17036362 ↗
  • MalfunctionOct 21, 2019Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SCREW REMOVAL SURGERY AT L2 DUE TO BONE UNION HAD ACHIEVED. INTRA-OP, SCREW ON THE LEFT SIDE OF L2 HAD BROKEN DURING REMOVING THE SCREW AS THE BONE UNION ACHIEVED. FRAGMENTS OF THE IMPLANT WERE REMAINING IN THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED DUE TO THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9213161 ↗
  • InjuryOct 14, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT L4 VERTEBRAL REPLACEMENT AND POSTERIOR FIXATION AT L2-S1. ON AN UNKNOWN DATE, POST-OP, THERE WAS A CLEAR ZONE AT L2 RIGHT SCREW PART AND INFECTION WAS REPORTED AT THAT SITE. HENCE, THE PATIENT UNDERWENT A REVISION SURGERY, IN WHICH THE SCREW WAS COMPLETELY EXPLANTED. THE INFECTED SITE WAS WASHED; AND A HOOK WAS PLACED. THEN, WOUND CLOSURE WAS PERFORMED.

    Patient problems: Unspecified Infection

    MAUDE report 9187963 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page