CD HORIZON® SOLERA® VOYAGER™ Spinal System 55750016540Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 55750016540 5.5VOYAGER STMAS6.5X40 · Class 2 · Implant
All models of CD HORIZON SOLERA VOYAGER Spinal System (60)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169783836), received between Aug 24, 2019 and Oct 26, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Malposition of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Malposition of Device1
- Migration1
Patient problems
- Paralysis1
- Paraplegia1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryOct 26, 2023Malposition of Device; Malposition of Device
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR FUSION AT TH12-L5 FOR PYOGENIC SPONDYLITIS. IT WAS REPORTED THAT AFTER INSTALLING THE ILIAC REFERENCE FRAME, OARM/NAVI WAS USED TO BEGIN INSERTING THE TH12 SCREW. AFTER INSERTING THE TH12 SCREW, A SENSE OF DISCOMFORT WAS FELT WHEN CREATING THE PILOT HOLE FOR THE TH11 SCREW, SO THE SCREWS WERE INSERTED INTO L4 AND L5 VERTEBRAL BODIES FIRST, AND THEN TRANSFERRED TO THE TH12 AND TH11 SCREWS, BUT IT WAS CHANGED TO A SPINOUS PROCESS REFERENCE, AND THE OARM WAS ROTATED AGAIN. AT THAT TIME, THE SCREW OF TH12 WAS FOUND TO HAVE DEVIATED INTO THE SPINAL CANAL AND WAS REINSERTED. AFTER THAT, THE SCREW WAS INSERTED INTO THE REMAINING VERTEBRAL BODY AND THE SURGERY WAS COMPLETED. PATIENT DEVELOPED LOWER LIMB PARALYSIS. THERE WAS A DELAY OF LESS THAN 60 MINUTES IN THE OVERALL PROCEDURE TIME. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: Paraplegia
MAUDE report 18009791 ↗InjuryAug 24, 2019Migration; Migration
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT PRESENTED WITH DEGENERATIVE SCOLIOSIS; AND UNDERWENT TRANSFORAMINAL LUMBAR INTERBODY FUSION AT L3-L5. O-ARM/NAVIGATION WAS USED IN COMBINATION IN THIS SURGERY. POST-OP, SCREW IMPLANTED AT L3 DEVIATED FROM THE PEDICLE, DUE TO WHICH THE PATIENT GOT PARALYZED. HENCE, THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2019, IN WHICH THE DEVIATED SCREW WAS REPLACED. PATIENT'S HOSPITAL STAY WAS PROLONGED BECAUSE OF THE RE-OPERATION. PATIENT'S CURRENT STATUS (AFTER REPLACEMENT SURGERY) IS NOT KNOWN.
Patient problems: Paralysis
MAUDE report 8928725 ↗
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.