Skip to content

CD HORIZON® SOLERA® VOYAGER™ Spinal System 55750017555Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SCREW 55750017555 5.5VOYAGER STMAS7.5X55 · Class 2 · Implant

All models of CD HORIZON SOLERA VOYAGER Spinal System (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169783935), received between Jun 30, 2023 and Jun 30, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 30, 2023Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING L4 SET SCREW RE PLACEMENT FOR AN INDICATION OF INADEQUATE BONE FUSION AFTER POSTERIOR FUSION. IT WAS REPORTED THAT THE PATIENT HAD INITIAL SURGERY OF POSTERIOR FIXATION BY L3/4DEPS TECHNIQUE AND POSTERIOR FIXATION BY S1/S2AI A ND REVISION SURGERY WAS PERFORMED DUE TO BONE FUSION FAILURE AFTER POSTERIOR FIXATION. THERE WAS NO MALFUNCTION WITH MDT PRODUCTS THAT RESULTED IN BONE FUSION FAILURE. DURING REVISION SURGERY, WHILE THE SET SCREW WAS ABOUT TO BE INSTALLED, IT WAS HARD LIKE A CROSS-THREAD, AND THE SET SCREW WAS REPLACED WITH A NEW ONE, BUT NO IMPROVEMENT. IT APPEARED THAT THE SCREW HEAD THREAD ITSELF MAY HAVE BEEN DEFORMED, AND WHEN THE SCREW WAS REPLACED AND THE SET SCREW WAS REPLACED AGAIN WITH ANOTHER ONE, THE PROBLEM WAS RESOLVED. WHILE THE SET SCREW IS A TITANIUM ALLOY, IT IS UNLIKELY THAT THE THREAD OF THE COBALT-CHROME ALLOY SCREW HEAD WAS PUSHED DOWN AND CRUSHED, BUT IT IS UNCLEAR WHETHER THE PROBLEM WAS ON THE SET SCREW OR THE SCREW. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17237175 ↗