Skip to content

CD HORIZON® SOLERA® VOYAGER™ Spinal System 55750018540Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SCREW 55750018540 5.5VOYAGER STMAS8.5X40 · Class 1 · Implant

All models of CD HORIZON SOLERA VOYAGER Spinal System (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169783966), received between Jul 19, 2021 and Jan 29, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Device Dislodged or Dislocated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Unspecified Infection1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2025Break; Break

    (B)(6) 2023 MPXR (B)(6) (REP, HCP, FOR): INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR LUMBAR VERTEBRAL FUSION FOR ADJACENT INTERVERTEBRAL DISORDER. LEVEL AT WHICH THE IMPLANT WAS PERFORMED: L3-S2 IT WAS REPORTED THAT WHEN INSERTING THE LEFT L4 SCREW, A LOAD WAS APPLIED TO THE TIP OF THE SCREWDRIVER, RESULTING IN BREAKAGE. THE BROKEN PART OF THE SCREWDRIVER TIP WAS CONFIRMED INSIDE THE SCREW HEAD. THE SCREW WAS REMOVED ONCE AND REPLACED WITH A NEW SCREW OF THE SAME SIZE. DATE OF INITIAL SURGERY: (B)(6) 2010, LEVEL AT WHICH IMPLANT WAS PERFORMED: L4/5, NO PRODUCT MALFUNCTION WAS REPORTED AGAINST THE PRODUCT USED IN THE INITIAL SURGERY. THE REPORTED SCREW WAS FRACTURED. THERE WAS A DELAY IN THE PROCEDURE BY LESS THAN 60 MINS. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21263585 ↗
  • InjuryJul 19, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE INFORMATION WAS RECEIVED FROM HEALTH CARE PROVIDER VIA A MANUFACTURING REPRESENTATIVE REGARDING A PATIENT WITH L4 / 5 / S SPINAL CANAL STENOSIS FOR OLIF OF L4 / 5 / S AND POSTERIOR L4 / 5 / S FIXATION SPINAL THERAPY. IT WAS REPORTED THAT THE L5 / S SCREWS AND THE POSTERIOR SCREWS WERE LOOSENED DUE TO INFECTION, THE ANTERIOR CAGE, SCREWS AND POSTERIOR SCREWS WERE REMOVED, AND THE ILIUM WAS COLLECTED, AND ANTERIOR FIXATION WAS PERFORMED. AFTER THAT, POSTERIOR L4 / 5 / S / S2 FIXATION WAS PERFORMED. THERE WAS NO FURTHER INFORMATION REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Infection

    MAUDE report 12187109 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →

Sources and updates