CG FUTURE™ 638RL28Recalls & reported problems
Medtronic, Inc. · RING 638RL28 CG FUTURE ANNULO 29L · Class 2 · Implant
All models of CG FUTURE (46)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic CG Future 28 mm annuloplasty ring is used in mitral valve repair surgery to reshape the heart's mitral valve. The reports mostly describe the ring being found deformed or not seating properly during surgery, leaving the valve leaking, so it was explanted and replaced with another product. Across 33 reports there are 32 injuries, no deaths, and 1 malfunction, with no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 0
- Injuries reported
- 32
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00643169793521), received between May 24, 2018 and Jun 11, 2026 — classified as 32 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (25 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information25
- Adverse Event Without Identified Device or Use Problem9
- Intravalvular regurgitation2
- Break1
- Incomplete Coaptation1
- Packaging Problem1
- Structural Problem1
Patient problems
- Insufficient Information20
- No Information10
- Mitral Regurgitation6
- Mitral Valve Insufficiency/ Regurgitation3
- Endocarditis1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 11, 2026Incomplete Coaptation
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28 MM CG FUTURE ANNULOPLASTY HEART RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS REPORTED AS SUBOPTIMAL LEAFLET COAPTATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 25470761 ↗InjuryMar 30, 2026Intravalvular regurgitation
IT WAS REPORTED THAT A CG FUTURE 28MM RING WAS USED DURING A MITRAL VALVE REPAIR PROCEDURE FOR A PATIENT BEING TREATED FOR INFECTIVE ENDOCARDITIS WITH MITRAL INSUFFICIENCY. DURING POST-IMPLANTATION TESTING, PERSISTENT CENTRAL REGURGITATION WAS OBSERVED, AND A PRODUCT ISSUE WAS SUSPECTED. THE RING WAS EXPLANTED AND REPLACED WITH ANOTHER MANUFACTURER¿S MITRAL VALVE, AND SUBSEQUENT TESTING SHOWED NO FURTHER CENTRAL REGURGITATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Mitral Valve Insufficiency/ Regurgitation
MAUDE report 24728836 ↗InjuryFeb 6, 2026Break
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE ANNULOPLASTY RING WAS NOT FULLY SUTURED AND TESTED PRIOR TO REMOVAL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 24283611 ↗InjurySep 18, 2025Structural Problem; Packaging Problem; Structural Problem; Packaging Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM CG FUTURE HEART RING, IT WAS FOUND TO BE DEFORMED AND REPLACED WITH AN UNKNOWN PRODUCT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 23084718 ↗InjurySep 8, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
ADDITIONAL INFORMATION RECEIVED REPORTED THE REASON FOR REPLACEMENT AS MITRAL VALVE PATHOLOGY AND THAT THE DEVICE WAS REPLACED BY A COMPETITORS VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 22992144 ↗InjuryMar 11, 2025Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation
IT WAS REPORTED THAT DURING IMPLANT OF THIS 28MM CG FUTURE MITRAL RING, IT WAS EXPLANTED AND REPLACED WITH A 27MM MOSAIC MITRAL VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation
MAUDE report 21572394 ↗InjuryMar 5, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING A MITRAL VALVULOPLASTY, A DEFORMATION OF A 28 MM CG FUTURE MITRAL VALVE RING WAS OBSERVED, LEADING TO A POOR FORMING EFFECT. THE DEFORMED MITRAL VALVE RING WAS EXPLANTED AND REPLACED WITH ANOTHER PRODUCT OF THE SAME MODEL DURING THE PROCEDURE, AND THE OPERATION WAS SUCCESSFULLY COMPLETED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 21524796 ↗InjuryOct 1, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 20346810 ↗InjuryJul 5, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 19680126 ↗InjuryFeb 9, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 1 YEAR POST IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 18677820 ↗InjuryDec 12, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 7 MONTHS POST IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 18310248 ↗InjuryAug 28, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT AFTER AN UNKNOWN DURATION POST IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, ANOTHER MANUFACTURERS TRANSCATHETER VALVE WAS IMPLANTED. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 17629948 ↗InjurySep 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS REPORTED TO BE DUE TO THE PHYSICIAN NOT BEING SATISFIED WITH THE FINAL RESULT OF THE REPAIR. IT WAS STATED THAT THE 38MM MITRAL ANNULOPLASTY RING DID NOT MALFUNCTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 15355469 ↗InjuryAug 16, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 15232540 ↗InjuryJun 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS REPORTED AS SYSTOLIC ANTERIOR MOTION. IT WAS REPORTED THAT THE 28MM MITRAL ANNULOPLASTY RING DID NOT MALFUNCTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 14753687 ↗InjuryJun 14, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 31MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 14686789 ↗InjuryMay 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS REPORTED AS AN "UNSATISFACTORY REPAIR". IT WAS REPORTED THERE WAS NO MALFUNCTION OF THE ANNULOPLASTY RING. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 14495897 ↗InjuryApr 5, 2022Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 27MM BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information; Mitral Valve Insufficiency/ Regurgitation
MAUDE report 14015497 ↗InjuryJan 10, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE EXPLANT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 13212208 ↗InjuryFeb 1, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY BAND, IT WAS EXPLANTED AND REPLACED WITH A 26MM BAND OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 11261188 ↗InjuryNov 5, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT THIS PATIENT'S NATIVE MITRAL VALVE WAS DILATED AND HAD SEVERE REGURGITATION. THE SURGEON IMPLANTED THIS 28 MM MITRAL ANNULOPLASTY RING IN ATTEMPT TO REPAIR THE PATIENT'S VALVE. TESTING WITH SALINE SHOWED A SUCCESSFUL REPAIR. UPON WEANING OFF OF CARDIOPULMONARY BYPASS (CPB) HOWEVER, RESIDUAL REGURGITATION WAS PRESENT. THE PATIENT WAS PLACED BACK ON CPB, AND THE VALVE WAS REPLACED WITH A BIOPROSTHETIC VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Regurgitation
MAUDE report 10792606 ↗InjuryNov 2, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A N ON-MEDTRONIC BIOPROSTHETIC VALVE. IT WAS REPORTED THAT THE PHYSICIAN "WAS NOT SATISFIED WITH THE RESULTS" OF THE REPAIR AND DECIDED TO PERFORM A FULL VALVE REPLACEMENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information; No Information
MAUDE report 10771458 ↗InjuryOct 28, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28 MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Information
MAUDE report 10749178 ↗InjurySep 2, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A NON-MEDTRONIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Mitral Regurgitation; No Information
MAUDE report 10485087 ↗InjuryAug 24, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT APPROXIMATELY 2 MONTHS POST IMPLANT OF THIS 28MM MITRAL ANNULOPLASTY RING, IT WAS EXPLANTED AND REPLACED WITH A 29MM MITRAL BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Information
MAUDE report 10442117 ↗
Other RING, ANNULOPLASTY devices
- CG FUTURE™ 638BL32Medtronic, Inc. · RING, ANNULOPLASTY1 death71 reports
- PROFILE 3D® 680R28Medtronic, Inc. · RING, ANNULOPLASTY1 death67 reports
- CG FUTURE™ 638BL30Medtronic, Inc. · RING, ANNULOPLASTY1 death65 reports
- CG FUTURE™ 638RL32Medtronic, Inc. · RING, ANNULOPLASTY59 reports
- Tri-Ad™ 900SFC228Medtronic, Inc. · RING, ANNULOPLASTY4 deaths56 reports
- CG FUTURE™ 638BL34Medtronic, Inc. · RING, ANNULOPLASTY51 reports
- Contour 3D® 690R30Medtronic, Inc. · RING, ANNULOPLASTY1 death50 reports
- CG FUTURE™ 638BL28Medtronic, Inc. · RING, ANNULOPLASTY1 death47 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page