Skip to content

Angiographic Guide Wire 010388Recalls & reported problems

Medtronic, Inc. · WIRE ANG 010388 10PK .025X145CM 3MMJPCS · Class 2

All models of Angiographic Guide Wire (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169857278), received between Apr 5, 2019 and Apr 5, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Deformation1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 5, 2019Material Deformation; Material Deformation

    AN ANGIOGRAPHIC WIRE WAS USED DURING A PROCEDURE. THE LESION WAS LOCATED IN THE CORONARY SINUS OSTIUM (DURING A CRT IMPLANT). NO DAMAGE NOTED TO PACKAGING. NO ISSUES REMOVING THE DEVICE FROM THE PACKAGING AS PER IFU. THE DEVICE WAS INSPECTED AND PREPPED PER IFU WITH NO ISSUES. THE WIRE TIP WAS NOT FORMED BY THE PHYSICIAN. IT WAS REPORTED THAT DAMAGE WAS NOTED TO THE DEVICE AFTER REMOVAL FROM THE PATIENT. THE PHYSICIAN USED THE ANGIOGRAPHIC WIRE INSIDE AN ATTAIN COMMAND CATHETER TO ACCESS THE CORONARY SINUS . AFTER TRYING TO PASS THE SUPERIOR VENA CAVA (ATRIAL AND RV LEAD WERE PREVIOUSLY IMPLANTED - DURING THE SAME PROCEDURE) THE PHYSICIAN NOTED THAT THE GUIDE WIRE WAS KINKED. THE PROCEDURE WAS COMPLETED USING ANOTHER 0.025" WIRE. NO PATIENT INJURY REPORTED. ANALYSIS REVEALED COILS SEPARATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8486647 ↗