Activa® RC 37612 (UDI 00643169864238)Recalls & reported problems

Medtronic, Inc. · INS 37612 ACTIVA RC NO PRYLN EMN DBS ALP · Class 3 · Implant

All models of Activa RC (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Activa RC deep brain stimulator has 1,849 adverse event reports on record, including 2 deaths and 277 injuries. Many reports describe connection and charging problems, material integrity problems and batteries running down, with some describing a broken connector pin in the recharger and patients feeling too much stimulation. Reports fell to 102 in the last 12 months, from 168 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,846
Deaths reported
2
Injuries reported
277
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,846 adverse event reports that list this exact device (DI 00643169864238), received between May 10, 2018 and Sep 17, 2026 — classified as 2 deaths, 277 injuries, 1,567 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (637 reports).

20 other recalls cover devices cleared under the same FDA submission (P960009); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 21 reportsJan 2024: 51 reportsFeb 2024: 75 reportsMar 2024: 85 reportsApr 2024: 20 reportsMay 2024: 9 reportsJun 2024: 11 reportsJul 2024: 9 reportsAug 2024: 18 reportsSep 2024: 15 reportsOct 2024: 9 reportsNov 2024: 12 reportsDec 2024: 8 reportsJan 2025: 17 reportsFeb 2025: 15 reportsMar 2025: 14 reportsApr 2025: 6 reportsMay 2025: 22 reportsJun 2025: 16 reportsJul 2025: 22 reportsAug 2025: 11 reportsSep 2025: 15 reportsOct 2025: 10 reportsNov 2025: 12 reportsDec 2025: 16 reportsJan 2026: 10 reportsFeb 2026: 16 reportsMar 2026: 21 reportsApr 2026: 6 reportsMay 2026: 4 reportsJun 2026: 3 reportsJul 2026: 5 reportsAug 2026: 3 reportsSep 2026: 3 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202685
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem637
  • Material Integrity Problem573
  • Charging Problem455
  • Battery Problem305
  • Failure to Deliver Energy240
  • Communication or Transmission Problem235
  • Adverse Event Without Identified Device or Use Problem157
  • Improper or Incorrect Procedure or Method118

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,060
  • No Known Impact Or Consequence To Patient215
  • Shaking/Tremors177
  • Insufficient Information155
  • Electric Shock88
  • Ambulation or Postural Difficulties66
  • Complaint, Ill-Defined57
  • Unspecified Infection53

Recent adverse event reports

The 25 most recent of 1,846 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Electromagnetic Compatibility Problem

    It was reported that patient said their DBS therapy feels less effective when they play video games and use their washing machine. There are no known contributing factors. All impedances were checked and were in range. This issue is considered resolved.

    Patient problems: Insufficient Information

    MAUDE report 50802948 ↗
  • MalfunctionSep 17, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that caller had a frayed recharger antenna.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50799106 ↗
  • MalfunctionSep 17, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the device was returned with no known allegation, analysis found non-conformance.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50797931 ↗
  • InjuryAug 28, 2026Failure to Deliver Energy

    It was reported that the patient said they would like to know how the stimulation starts to lower or reduced when the battery reaches a quarter. Patient said they can tell the stark difference once the battery hits 25%. Patient said the HCP programmed the device this way so the patient would remember to charge, Patient said it has been like this for as long as the patient can remember. Agent reviewed the device will deliver stim at the programmed settings till the implantable neurostimulator (INS) depletes. The patient was redirected to their healthcare provider to further address the issue.

    Patient problems: Insufficient Information

    MAUDE report 50507868 ↗
  • InjuryAug 23, 2026Overheating of Device; Charging Problem

    It was reported that  patient can't charge past 75% because it gets too hot and patient feels uncomfortable. Representative said they told patient to call technical services to lower the charging speed to see if patient doesn't get as hot. The patient was redirected to their healthcare provider to further address the issue.

    Patient problems: Discomfort

    MAUDE report 50397467 ↗
  • InjuryAug 17, 2026Inappropriate/Inadequate Shock/Stimulation; Insufficient Device Problem Information

    It was reported that patient said they have been having trouble with their head and the DBS system is acting weird. Patient said it is like there is too much current. Patient said they did discuss w/ healthcare provider (HCP) and they didn't know.

    Patient problems: Electric Shock; Insufficient Information

    MAUDE report 50287753 ↗
  • MalfunctionJul 16, 2026Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT PATIENT (PT) WAS HAVING ISSUES NOT BEING ABLE TO SPEAK. THE PATIENT REPRESENTATIVE SAID THAT THE PATIENT THOUGHT SOMETHING WAS WRONG WITH THEIR DBS IMPLANT BECAUSE THEY'VE HAD A RETURN OF SYMPTOMS THE PAST COUPLE OF DAYS. THE PATIENT REPRESENTATIVE SAID THAT THE PATIENT HAD BEEN FEELING SOME SHOCKS AND THE PATIENT'S LEFT LEG WAS SHAKING BUT THAT WAS EFFECTING THE PATIENT'S WHOLE BODY. THE PATIENT REPRESENTATIVE SAID THEY HAD TAKEN THE PATIENT INTO THE EMERGENCY ROOM (ER) AND THEY CHECKED THE IMPLANT AND TOLD THE PATIENT THAT EVERYTHING WAS OKAY. DURING THE CALL AGENT WALKED CALLER THROUGH HOW TO CONNECT WITH THE IMPLANT, THERAPY WAS SHOWING ON AND OK. PATIENT REPRESENTATIVE WAS REDIRECTED TO THE HEALTH CARE PROVIDER TO FURTHER ADDRESS THE ISSUE. ONCE SPEAKING WITH HEALTH CARE PROVIDER THE CALLER MAY CALL BACK FOR ASSISTANCE ON HOW TO MAKE A DOCTOR PROVIDED ADJUSTMENT.

    Patient problems: Dysphasia; Electric Shock; Shaking/Tremors

    MAUDE report 25847422 ↗
  • MalfunctionJul 14, 2026Battery Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT BELIEVES THE IMPLANT IS NOT FUNCTIONING RIGHT NOW. CALLER SAID THE PATIENT SAID SOMETHING ABOUT THE BATTERY NOT BEING CHARGED, BUT THEY WERE LOOKING IN THE PATIENT'S CHART AND IT LOOKED LIKE IT COULD HAVE BEEN AN INTERROGATION FROM JUNE 6TH AND WHEN THEY LOOKED AT THE REPORT, IT SAID THE BATTERY LEVEL WAS AT 0%. THE PATIENT WAS REDIRECTED TO THEIR HEALTHCARE PROVIDER TO FURTHER ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25813106 ↗
  • MalfunctionJul 14, 2026Battery Problem; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT BELIEVES THE IMPLANT IS NOT FUNCTIONING RIGHT NOW. CALLER SAID THE PATIENT SAID SOMETHING ABOUT THE BATTERY NOT BEING CHARGED, BUT THEY WERE LOOKING IN THE PATIENT'S CHART AND IT LOOKED LIKE IT COULD HAVE BEEN AN INTERROGATION FROM JUNE 6TH AND WHEN THEY LOOKED AT THE REPORT, IT SAID THE BATTERY LEVEL WAS AT 0%. THE PATIENT WAS REDIRECTED TO THEIR HEALTHCARE PROVIDER TO FURTHER ADDRESS THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25813093 ↗
  • MalfunctionJul 14, 2026Charging Problem; Connection Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE RECHARGER WON'T CHARGE. PATIENT SAID THE ISSUE STARTED RECENTLY, IN JULY. PATIENT CONFIRMED THE GREEN LIGHT WAS ON THE POWER SUPPLY AND THE CONNECTOR PIN WAS LOOSE. THE ISSUE WAS NOT RESOLVED THROUGH TROUBLESHOOTING. AN EMAIL WAS SENT TO THE REPAIR DEPARTMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25811128 ↗
  • MalfunctionJul 9, 2026Energy Output Problem

    IT WAS REPORTED THAT PATIENT REPORTED IT IS TAKING LONGER TO CHARGE THEIR IMPLANTED NEUROSTIMULATOR AT NIGHT. PATIENT STATED THEY USED TO CHARGE EVERY NIGHT FOR 7 MINUTES AND NOW IT TAKES ANYWHERE FROM 17 TO 20 MINUTES TO CHARGE. PATIENT REPORTED THIS STARTED ABOUT A YEAR AGO. PATIENT CONFIRMED THERE IS NO VISIBLE DAMAGE TO RECHARGER ANTENNA CORD. PATIENT'S NEUROLOGIST CHECKED PATIENT'S IMPLANT AND TOLD PATIENT THEIR IMPLANT BATTERY IS STILL IN GOOD SHAPE AND ADVISED PATIENT GET THEIR RECHARGER REPLACED. PATIENT CONFIRMED GETTING ALL COUPLING BOXES FILLED IN WHEN CHARGING. PATIENT CONFIRMED THEY HAVE HAD A LITTLE THERAPY INCREASE ON THEIR LEFT SIDE AND STATED THEIR NEUROLOGIST ADJUSTED IT UP A LITTLE FROM 0 LAST YEAR BUT DID NOT CHANGE IT THIS TIME. AGENT REVIEWED THERAPY ADJUSTMENTS RELATED TO CHARGING DURATION AND PATIENT CONFIRMED UNDERSTANDING. PATIENT MENTIONED THEY HAVE TRIED PLUGGING IN \"THE OTHER PART THAT CHARGES IT\" TO SEE IF THAT WOULD HELP AND SATED IT DOES HELP BUT STATED IT DOESN'T BRING THEM \"BACK TO STANDARD\" (PATIENT DID NOT CLARIFY THIS). THE ISSUE WAS NOT RESOLVED THROUGH TROUBLESHOOTING. REVIEWED WIRELESS RECHARGER OVERVIEW. A REQUEST WAS SENT TO THE REPAIR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25759192 ↗
  • MalfunctionJun 11, 2026Charging Problem

    IT WAS REPORTED THAT THE RECHARGING ANTENNA THE GRAY CORD IS DAMAGED AND WIRE IS EXPOSED. IT HAS BEEN DAMAGED FOR A LITTLE WHILE. JUST RECENTLY TO THE POINT IN THE PAST 6 MONTHS IT WON'T HOLD A CHARGE THAT LONG. THE RECHARGER WON'T CHARGE PAST A QUARTE OF THE WAY, SO THEY CAN'T FULLY CHARGE THEIR NEUROSTIMULATOR. SENT REQUEST TO REPAIR TO RELACE THE RECHARGE ANTENNA AND THE AC/DTC. 2026-MAY-12 (B)(4) (CON): ADDITIONAL INFORMATION HAS BEEN RECEIVED THAT THEY ATTEMPTED TO CHARGE THE RECHARGER BUT THE DEVICE WAS STILL NOT CHARGED AFTER HAVING IT PLUGGED INTO THE WALL. CALLER STATED THAT THE REPLACEMENT DESKTOP CHARGER (DTC) DID NOT RESOLVE THE ISSUE. AGENT SENT MESSAGE TO THE FIELD TO TRANSITION THE PATIENT TO THE WR. 2026-MAY-15 (B)(4) (CON): ADDITIONAL INFORMATION HAS BEEN RECEIVED THAT THE REPLACEMENT DTC AND ANTENNA DID NOT RESOLVE THE ISSUE. THE CALLER STATED THAT THE PATIENT HAS BEEN UNABLE TO CHARGE THE IMPLANTABLE NEUROSTIMULATOR (INS) FOR ALMOST A WEEK. CALLER HAS NOT HEARD FROM A MANUFACTURER REP TO TRANSITION THE PATIENT TO THE WR. CALLER MENTIONED THAT THEY SENT A MESSAGE TO THE PATIENT'S HCP A COUPLE OF WEEKS AGO BUT HAVEN'T HEARD BACK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25468644 ↗
  • MalfunctionJun 8, 2026Overheating of Device; Charging Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT EXPERIENCED DULL PAIN, BURNING SENSATION, AND A HEATING SENSATION DURING RECHARGING. THE ISSUE WAS R EPORTED TO HAVE STARTED A COUPLE OF MONTHS PRIOR. THERE WERE NO RECENT THERAPY SETTING CHANGES AND NO REPORTED FALL OR TRAUMA. IMPEDANCES WERE CHECKED AND REPORTED AS FINE AND THE PATIENT WAS REPORTED TO BE RECEIVING GOOD SYMPTOM RELIEF. THE PATIENT WAS EVALUATED BY A PHYSICIAN AND THE POCKET SITE WAS NOTED TO LOOK FINE WITH NO PROBLEMS. TROUBLESHOOTING GUIDANCE INCLUDED RECOMMENDING CHARGING WITH CLOTHING OVER THE IMPLANTABLE NEUROSTIMULATOR SITE, PERFORMING MORE FREQUENT BUT SHORTER CHARGING SESSIONS, AND A PLANT TO TRANSITION FROM A LEGACY RECHARGER TO A WIRELESS RECHARGER. ADDITIONAL INFORMATION HAS BEEN RECEIVED THAT THE PATIENT RECEIVED A WIRELESS RECHARGER BUT THEY COULD NOT GET IT TO PAIR WITH THE HANDSET. THE PATIENT REP HAD ALREADY SPOKEN WITH MANUFACTURER REP, WHO TOLD THE CALLER THAT THE PATIENT MIGHT NEED A NEWER VERSION OF THE HANDSET. AGENT HAD CALLER CHECK HANDSET (J3) AND THE RECHARGER APP WAS NOT INSTALLED. AGENT RE-OPENED CASE AND SENT REQUEST TO THE REPAIR DEPARTMENT TO REPLACE THE HANDSET. ADDITIONAL INFORMATION RECEIVED FROM REP I

    Patient problems: Burning Sensation; No Clinical Signs, Symptoms or Conditions; Pain

    MAUDE report 25431084 ↗
  • MalfunctionJun 4, 2026Migration or Expulsion of Device; Unintended Collision

    IT WAS REPORTED THAT PATIENT FELL AND BROKE THEIR NECK THE OTHER DAY AND IS IN A HOSPITAL AND NEEDS TO DO AN MRI BEFORE THEY CAN DO SURGERY. CALLER SAID THAT THE FOLKS AT THE HOSPITAL ARE CONFUSED ABOUT WHAT TO DO IN TERMS OF GETTING THE PHYSICAL EXAMINATION OF THE DEVICE AND WIRES AND THE BATTERIES BEFORE THEY GO IN THE MRI MACHINE. CALLER ASKED IF THERE IS A LOCAL MANUFACTURER REPRESENTATIVE (REP) THAT CAN DO THE MRI. REVIEWED MRI GUIDELINES AND REDIRECTED CALLER TO PATIENT'S MANAGING HEALTHCARE PROVIDER TO FURTHER ADDRESS THE ISSUE. WHEN ASKED, CALLER SAID THAT PATIENT'S NEUROLOGIST IS AT THE CLINIC, WHICH IS ABOUT 2 HOURS AWAY FROM THE HOSPITAL AND THE PRIMARY NEUROLOGIST ISN'T ABLE TO INSPECT THE DEVICE HERSELF. CALLER SAID THAT THEY WERE DIRECTED TO CONTACT SOMEONE LOCALLY TO ASSIST. CALLER ALSO SAID THAT THEY JUST GOT BACK FROM AN X-RAY AND THE MRI FACILITY SAID THEY WON'T DO THE SCAN UNLESS SOME VERIFICATION IS PROVIDED. CALLER WANTED TO KNOW ABOUT THE MRI ELIGIBILITY FORM. AFTER CONSULTING WITH TECHNICAL SERVICES AND REVIEWED INFORMATION. EMAIL WAS SENT TO FIELD STAFF TO NOTIFY THEM OF THE SITUATION AND REQUEST THAT THEY FOLLOW UP WITH THE PATIENT REP AND THE HOSPITAL.\" C

    Patient problems: Bone Fracture(s); Fall; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25396355 ↗
  • MalfunctionMay 29, 2026Migration or Expulsion of Device

    IT WAS REPORTED THAT THE PATIENT'S IMPLANTABLE NEUROSTIMULATOR (INS) IS LOCATED IN THEIR ABDOMEN AND THAT FOR THE LAST FIVE-SIX MONTHS THEIR INS HAS BEEN MOVING AROUND. THE PATIENT WAS REDIRECTED TO THE HCP TO FURTHER ASSIST WITH THE INS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25332893 ↗
  • MalfunctionMay 8, 2026High impedance; Low impedance

    IT WAS REPORTED THAT LOW IMPEDANCE/SHORT, 223 OHMS CONTACT 3 ON MONOPOLAR IMPEDANCE ON LEFT CONTACT. PATIENT (PT) REPORTING NO LOSS OF THERAPY OR CHANGE IN RECHARGE RESPONSIBILITY. HIGH IMPEDANCE, 2025 OHMS, CONTACT 8 ON MONOPOLAR IMPEDANCE RIGHT CONTACT. PT REPORTING NO LOSS OF THERAPY OR CHANGE IN RECHARGE RESPONSIBILITY. PT REPORTED HAVING A FALL IN 2024 AS A POTENTIAL CONTRIBUTING FACTOR. IMPEDANCES RECHECKED ON AUTO AND VARIOUS AMPLITUDES. PT TO RETURN TO CLINIC IN 3 MOS FOR INTERROGATION. MANUFACTURER REPRESENTATIVE (REP) INDICATES THE ISSUE HAS NOT BEEN RESOLVED AT THE TIME OF REPORT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25128581 ↗
  • MalfunctionMay 6, 2026Improper or Incorrect Procedure or Method; Battery Problem

    2026-MAY-05 RTG1041741 (REP): IT WAS REPORTED THAT INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE STATING THE PATIENT RESIDED IN A NURSING FACILITY WHERE STAFF HAD REPORTEDLY REFUSED TO CHARGE THE IMPLANTED NEUROSTIMULATOR. THE REPRESENTATIVE INITIALLY BELIEVED THE BATTERY WAS DEPLETED BUT LATER REPORTED THAT THE DEVICE HAD ENTERED AN OVERDISCHARGE STATE. THE REPRESENTATIVE ALSO REPORTED OBSERVING A MESSAGE INDICATING THE RECHARGER WAS MISSING THE TARGET. ATTEMPTS TO CONNECT USING THE PATIENT PROGRAMMER AND A CLINICIAN TABLET WERE UNSUCCESSFUL. TECHNICAL SERVICES REVIEWED INSTRUCTIONS FOR INITIATING PHYSICIAN RECHARGE MODE. 2026-MAY-07 E1 (REP): THE MANUFACTURER REPRESENTATIVE PROVIDED ADDITIONAL INFORMATION INDICATING THAT THE OVERDISCHARGE CONDITION WAS ADDRESSED BY FOLLOWING THE OVERDISCHARGE RECOVERY PROCEDURE AND CHARGING THE BATTERY TO 50%. STAFF AT THE REHABILITATION FACILITY WERE ALSO INSTRUCTED ON PROPER CHARGING PROCEDURES. THE PATIENT EXPERIENCED STIFFNESS ASSOCIATED WITH THE OVERDISCHARGE. THE DATE THE CONDITION WAS FIRST OBSERVED BY THE PHYSICIAN IS UNKNOWN. THE INFORMATION WAS CONFIRMED AND PROVIDED BY THE PHYSICIAN.

    Patient problems: Insufficient Information; Muscular Rigidity

    MAUDE report 25107379 ↗
  • MalfunctionMay 6, 2026Failure to Deliver Energy

    IT WAS REPORTED THAT PATIENT (PT) THINKS THEY MIGHT HAVE DONE SOMETHING WRONG WITH THE CONTROLLER OR SOMETHING HAPPENED BECAUSE THEY WERE TRYING TO CHANGE THE SETTINGS ON THE CONTROLLER, BUT THEY MIGHT HAVE PRESSED SOMETHING ELSE. AGENT WALKED PATIENT THROUGH CONNECTING TO THEIR IMPLANTABLE NEUROSTIMULATOR (INS). PATIENT STATED THEY WERE SEEING A FLASHING OFF SIGN ON THE PROGRAMMER. AGENT ASKED IF THEY WANTED THEIR THERAPY ON. PATIENT SAID YES, SO AGENT WALKED PATIENT THROUGH TURNING THEIR THERAPY ON AND PATIENT WAS ABLE TO TURN THEIR THERAPY ON. EITHER WAY, AGENT FLAGGED CASE AS STAGING RECORD TO BE CONSERVATIVE. DUE TO NATURE OF CALL AND PATIENT HAVING A HARD TIME FOLLOWING DIRECTIONS ON HOW TO FIND SERIAL NUMBERS, SERIAL NUMBERS OF EXTERNAL DEVICES AND EVENT DATE WERE NOT GATHERED DURING THE CALL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25096952 ↗
  • MalfunctionApr 21, 2026Delayed Charge Time

    ADDITIONAL INFORMATION WAS PROVIDED INDICATING THAT THE PATIENT CALLED BACK, REITERATING PRIOR CONCERNS, AND ADDED THAT WHILE HOLDING THE RECHARGER, THE ANTENNA WAS MIGRATING AWAY FROM THE INS, MAKING CHARGING DIFFICULT. THE CALLER REPORTED DIFFICULTY SEEING SYMBOLS AND BUTTONS ON THE PROGRAMMER AND RECHARGER, AND STATED THAT AFTER HOLDING THE ANTENNA OVER THE INS FOR APPROXIMATELY ONE HOUR, THE BATTERY LEVEL REMAINED AT 50%, UNCHANGED FROM THE START OF CHARGING. THE CALLER ALSO NOTED NOT RECEIVING AN UPDATE AFTER ONE WEEK REGARDING THE WIRELESS RECHARGER THAT WAS EXPECTED TO BE SENT. THE AGENT REVIEWED THE TRANSITION PROCESS FROM THE CURRENT RECHARGER TO THE WIRELESS RECHARGER AND SENT AN EMAIL TO THE FIELD TEAM FOR VISIBILITY AND FOLLOW-UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24940020 ↗
  • MalfunctionApr 21, 2026Charging Problem; Connection Problem

    IT WAS REPORTED THAT THE WIRELESS RECHARGER (ACTIVA RC) FOR A DEEP BRAIN STIMULATION IMPLANTABLE PULSE GENERATOR (IPG) WAS UNRESPONSIVE, THE DOCK WAS NOT STAYING LIT, AND THE RECHARGER WAS NOT CHARGING FROM THE DOCK. THE GREEN LIGHT ON THE DOCK WOULD NOT COME ON UNLESS THE CORD WAS POSITIONED IN A CERTAIN WAY, AND THE GREEN LIGHTS ON THE RECHARGER WERE OBSERVED TO SCROLL RAPIDLY. THERE WAS A POSSIBLE CONNECTIONISSUE WITH THE CORD AND DOCK. THERE WAS CONCERN FOR THE IMPLANTABLE NEUROSTIMULATOR BATTERY DEPLETING, AS THE PATIENT WAS AT 50% CHARGE. THE PATIENT REPRESENTATIVE UNPLUGGED THE AC POWER SUPPLY AND PLUGGED IT INTO A DIFFERENT OUTLET, BUT THE SAME ISSUE OCCURRED. PATIENT SERVICES REVIEWED BEST PRACTICES FOR RECHARGING WITH THE CALLER AND ADVISED TO KEEP THE DOCK PLUGGED IN AND THE RECHARGER ON THE DOCK WHEN NOT IN USE. A REQUEST WAS SENT TO THE REPAIR DEPARTMENT TO REPLACE THE DOCK, RECHARGER, CHARGING CABLE, AND WALL ADAPTOR DUE TO THE URGENCY OF CHARGING THE IMPLANT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24939354 ↗
  • MalfunctionApr 17, 2026Failure to Interrogate; Migration or Expulsion of Device; Communication or Transmission Problem

    IT WAS REPORTED THAT THE PATIENT WOULD TYPICALLY CHARGE THEIR IMPLANTED NEUROSTIMULATOR FOR ABOUT 90 MINUTES AND TODAY (ON (B)(6) 2026, AND LAST WEEK TOO) THEY PUT THE RECHARGER ON THE PATIENT AND IT WENT FOR LIKE 25 MINUTES AND THEN IT STARTED BEEPING AGAIN. THE CALLER STATED THEY WOULD HAVE TO REPOSITION THE RECHARGING ANTENNA TO GET THE SENSORS MATCHED UP ON THE SENSOR UNDER THE PATIENT'S SKIN AND IT WOULD WORK FOR LIKE A MINUTE AND THEN THE BEEPING WOULD START AGAIN. CALLER CONFIRMED THAT THE IMPLANTABLE NEUROSTIMULATOR (INS) WAS SHIFTING UNDER THE PATIENT'S SKIN. PATIENT WAS IN THE BACKGROUND OF THE CALL AND SAID THE INS HAD BEEN MOVING AROUND FOR A LONG TIME AND THAT THE MOVING AROUND HAD STARTED BEFORE THE RECENT ISSUE WITH THE RECHARGER NOT STAYING CONNECTED HAD BEGUN. CALLER SAID THE RECHARGER NOT CONNECTING APPEARED TO HAVE HAPPENED ALL OF SUDDEN AND THEY NEVER HAD AN ISSUE BEFORE. THEY ALSO MENTIONED THEY WERE NOT WITH THE PATIENT EVERY DAY SO THEY WOULD ASK THE OTHER CAREGIVERS IF THEY'D NOTICED AN ISSUE. PATIENT COULD NOT SAY FOR HOW LONG THE INS HAD BEEN MOVING AROUND. PATIENT SERVICES REVIEWED RECHARGING CONSIDERATIONS WITH THE CALLER. THE CALLER SAID THE PATIENT WAS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24911659 ↗
  • MalfunctionApr 17, 2026Connection Problem; Material Integrity Problem

    IT WAS REPORTED THAT THEY HAD A BROKEN DESKTOP CHARGER CONNECTOR PIN. A MESSAGE WAS SENT TO REPAIR TO REPLACE THE DESKTOP CHARGER. AGENT ASKED WHEN THEY NOTICED THE DEVICE DAMAGED AND THEY SAID MAYBE LAST WEEK. PATIENT WAS ABLE TO GET THE BROKEN METAL PIN OUT OF THE RECHARGER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24908117 ↗
  • MalfunctionApr 14, 2026Connection Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE METAL TIP OF THE ADAPTER DETACHED. THE ADAPTER WAS REPLACED, AND THE ISSUE WAS REPORTED AS RESOLVED. THE PATIENT WAS REPORTED TO BE ALIVE WITH NO INJURY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24874514 ↗
  • MalfunctionApr 3, 2026Connection Problem; Material Integrity Problem

    IT WAS REPORTED THAT INFORMATION WAS RECEIVED FROM A PATIENT REGARDING AN EXTERNAL DEVICE. THE CALLER HAD A BROKEN DESKTOP CHARGER CONNECTOR PIN, PT SAID THE CASING IS BROKEN AND THE WIRES ARE EXPOSED WHEN THEY TRIED TO USE IT THE RECHARGER SCREEN WAS DARK AND BLANK. A REQUEST WAS SENT TO REPAIR TO REPLACE THE DESKTOP CHARGER. THE PATIENT MENTIONED UNRELATED MEDICAL HISTORY. THIS INCLUDED PATIENT SAID THEY WERE SUCCESSFUL TO CHARGE THEIR IMPLANTED NEUROSTIMULATOR BEFORE IT WENT DEAD YESTERDAY. PT WILL CALL BACK IF THE REPLACEMENT DOES NOT RESOLVE THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24773355 ↗
  • MalfunctionMar 31, 2026Failure to Deliver Energy

    IT WAS REPORTED THAT THE PATIENT HAS BEEN TREMBLING SINCE LAST NIGHT. REVIEWED HOW TO CHECK THERAPY STATUS WITH THE PATIENT PROGRAMMER AND CALLER REPORTED THERAPY WAS OFF. REVIEWED HOW TO TURN THERAPY BACK ON AGAIN AND CALLER WAS ABLE TO DO SO SUCCESSFULLY.

    Patient problems: Shaking/Tremors

    MAUDE report 24747535 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P960009 on Jul 31, 1997. Submitted by Medtronic, Inc. as “Medtronic Activa Tremor Control System”. The FDA review took 456 days. The FDA has since approved 532 supplements (changes to the device or its labeling), most recently on Nov 13, 2026. FDA record ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Recalls under the same FDA submission

These recalls involve devices cleared under P960009. One submission often covers many models, so check the product and lot details in each record.

  • Z-2240-2026 · May 13, 2026 · Open, Classified
    Medtronic 1x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64001; Stimulator, electrical, implanted, for parkinsonian tremor
  • Z-2241-2026 · May 13, 2026 · Open, Classified
    Medtronic 2x4 Pocket Adaptor Kit for Deep Brain Stimulation, Model 64002; Stimulator, electrical, implanted, for parkinsonian tremor
  • Z-0206-2025 · Sep 24, 2024 · Open, Classified
    Medtronic Percept RC Implantable Neurostimulators (INS), Model B35300.
  • Z-0372-2025 · Sep 18, 2024 · Open, Classified
    Medtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
  • Z-0217-2024 · May 27, 2022 · Completed
    LEAD 3387S-40 STIMLOC DBS, Model Number 3387S-40. Electrical implantable device.
  • Z-1128-2022 · Mar 30, 2022 · Open, Classified
    Percept BrainSense Implantable Neurostimulator (INS), Model B35200
  • Z-0418-2022 · Oct 19, 2021 · Open, Classified
    Model B35200 - Percept PC BrainSense Implantable Neurostimulator
  • Z-0201-2022 · Sep 23, 2021 · Open, Classified
    Clinician Programmer Application (CPA) model A610 Clinician Software Application
  • Z-1368-2021 · Mar 9, 2021 · Open, Classified
    A610 Clinician Programmer Application for Deep Brain Stimulation Clinician Programming Guide Percept PC Model B35200 Neurostimulator Application, version 2.0.4584, 2.0.4594, 2.0.46
  • Z-3069-2017 · Mar 17, 2017 · Terminated
    DBS Lead Depth Stop contained in Medtronic Deep Brain Stimulation (DBS) lead kits, models 3387/3387S, 3389/3389S, and 3391/3391S, The lead kit contains one DBS lead in addition to
  • Z-2645-2016 · Jun 23, 2016 · Terminated
    DBS Pocket Adaptor, Model 64001 and Model 64002. The Medtronic Models 64001 (1x4) and 64002 (2x4) Pocket Adaptors can be used as a part of a neurostimulation system for deep brain
  • Z-2602-2016 · Jun 10, 2016 · Terminated
    Medtronic, Lead Kit for DBS Stimulation. Model Numbers 3387S-40 (p/n 3387S0007V) & 3389S-40 (p/n 3389S0008V).
  • Z-1617-2015 · Apr 16, 2015 · Terminated
    Medtronic DBS Extension Kit for Deep Brain Stimulation, Models 37085 and 37086. Sterile
  • Z-0279-2014 · Sep 30, 2013 · Terminated
    Medtronic Activa PC, model 37601, and Activa RC, model 37612, Activa SC, models 37602 and 37603. Multi-program Rechargeable Neurostimulator for Deep Brain Stimulation.
  • Z-1190-2013 · Feb 8, 2013 · Terminated
    Medtronic DBS Lead Kit for Deep Brain Stimulation, models 3387, 3387S, 3389, 3389S, 3391, 3391S. Sterile and Non-Pyrogenic. Product Usage: The deep brain stimulation system del

…and 5 more.

Common questions about the Activa® RC 37612

Has the Activa® RC 37612 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00643169864238), based on the FDA recall file dated Nov 27, 2026. However, 20 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the Activa® RC 37612?

The FDA's MAUDE database lists 1,846 adverse event reports for this device, received between May 10, 2018 and Sep 17, 2026. In those reports, the most common specific device problems are “Connection Problem” (637), “Material Integrity Problem” (573) and “Charging Problem” (455); the most common patient problems are “Shaking/Tremors” (177), “Electric Shock” (88) and “Ambulation or Postural Difficulties” (66). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Activa® RC 37612?

2 reports are classified as deaths, 277 as injuries and 1,567 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Activa® RC 37612 FDA approved?

It was approved through premarket approval (PMA) P960009 on Jul 31, 1997.

Is the Activa® RC 37612 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Activa® RC 37612, and is it still sold?

The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Activa® RC, model 37612). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR devices

All STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR devices →