St. Jude Medical Infinity™ DBS System (REF 6172)Recalls & reported problems

St. Jude Medical, Inc. · 8CH Directional Lead, 40 cm, 0.5 · Class 3 · Implant

All models of St. Jude Medical Infinity (13)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This St. Jude Medical Infinity is a brain stimulation system with leads placed in the brain. The 307 reports mostly describe weak or failed stimulation, leads that migrate, and infections, with some reports of involuntary movements. There are 10 deaths and 297 injuries reported. A closed recall from October 2018 covers an electrode that may use the wrong material. Reports rose from 40 in the year before to 60 in the latest year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
306
Deaths reported
10
Injuries reported
296
Recalls naming it
1

Summary

The FDA's MAUDE database contains 306 adverse event reports that list this exact device (DI 05415067030320), received between Apr 8, 2019 and Sep 30, 2026 — classified as 10 deaths, 296 injuries.

The most frequently reported specific device problem is “Therapeutic or Diagnostic Output Failure” (64 reports).

1 FDA recall record name this device's identifier.

5 other recalls cover devices cleared under the same FDA submission (P140009); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 9 reportsDec 2023: 3 reportsJan 2024: 6 reportsFeb 2024: 5 reportsMar 2024: 8 reportsApr 2024: 6 reportsMay 2024: 3 reportsJun 2024: 0 reportsJul 2024: 4 reportsAug 2024: 2 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 3 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 6 reportsMay 2025: 5 reportsJun 2025: 4 reportsJul 2025: 4 reportsAug 2025: 5 reportsSep 2025: 3 reportsOct 2025: 2 reportsNov 2025: 4 reportsDec 2025: 12 reportsJan 2026: 10 reportsFeb 2026: 9 reportsMar 2026: 8 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 4 reportsJul 2026: 3 reportsAug 2026: 2 reportsSep 2026: 3 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 11, 2018Class IITerminated

    The most proximal unsegmented electrode of the Deep Brain Stimulation leads, may be constructed with MP35N instead of the required 90/10 platinum-iridium.

    Z-0593-2019 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem167
  • Therapeutic or Diagnostic Output Failure64
  • Migration24
  • High impedance13
  • Malposition of Device8
  • Use of Device Problem8
  • Insufficient Device Problem Information5
  • Fracture4

Patient problems

  • Movement Disorder61
  • Unspecified Infection56
  • Inadequate Pain Relief50
  • No Clinical Signs, Symptoms or Conditions31
  • Post Operative Wound Infection29
  • Erosion20
  • Wound Dehiscence14
  • Skin Erosion8

Recent adverse event reports

The 25 most recent of 306 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50952989 ↗
  • InjurySep 22, 2026Therapeutic or Diagnostic Output Failure

    It was reported that the patient experienced returning symptoms. As a result, surgical intervention was undertaken on (b)(6) 2026 wherein the lead was replaced to address the issue. Therapy was restored post-operatively.

    Patient problems: Movement Disorder

    MAUDE report 50860270 ↗
  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    It was reported patient had infection at the burr hole site. Investigation was unable to identify which burr hole cap attributed to the issue. Surgical intervention was undertaken wherein the left lead and burr hole were removed. Patient was prescribed anti-biotics to address the issue.

    Patient problems: Unspecified Infection

    MAUDE report 50828391 ↗
  • InjuryAug 31, 2026Adverse Event Without Identified Device or Use Problem

    It was reported the patient experienced worsening anxiety, suicidal thoughts, and depression. The patient was admitted for inpatient treatment.

    Patient problems: Anxiety; Depression; Suicidal Ideation

    MAUDE report 50530146 ↗
  • InjuryAug 3, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that following having the second lead implanted the patient couldn't walk and experienced speech issues. Suspected potential stroke post op. Additionally, it was reported that the patient was unable to place their system into MRI mode. Diagnostics revealed high impedances on the lead. Reprogramming was done to provide further symptom relief / speech improvement.

    Patient problems: Ambulation or Postural Difficulties; Speech Disorder; Stroke/CVA

    MAUDE report 50122648 ↗
  • InjuryJul 30, 2026Migration

    It was reported that imaging indicated the lead migrated after original placement. As a result, surgical intervention was undertaken on (b)(6) 2026 where the lead was replaced to address the issue. Therapy was not affected.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50077865 ↗
  • InjuryJul 29, 2026Adverse Event Without Identified Device or Use Problem

    it was reported that patient experienced reduced wound healing in the right lead and burr hole cap site. As result, surgical intervention was undertaken wherein the lead and burr hole cap site was cleaned and sutured to address the issue.

    Patient problems: Impaired Healing

    MAUDE report 50072066 ↗
  • InjuryJul 8, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT¿S BURR HOLE COVER WAS EXPOSED. THE SITE APPEARED INFECTED. THE PATIENT REPORTS A HARD FALL SEVERAL MONTHS PRIOR. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHERE THE LEAD AND BURR HOLE COVER WERE EXPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 25752210 ↗
  • InjuryJun 30, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING SURGERY ON (B)(6) 2026 [RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-07960, RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-07961, RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-07962, RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-08731, RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-08732] INFECTION WAS FOUND ONCE INCISION WAS MADE. THE PHYSICIAN WASHED THE SITE OUT WITH ANTIBIOTICS AND RECLOSED. IMPLANT CASE WAS ABANDONED.

    Patient problems: Unspecified Infection

    MAUDE report 25669474 ↗
  • InjuryJun 30, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING SURGERY ON (B)(6) 2026 [RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-07960, RELATED MANUFACTURER REFERENCE NUMBER:1627487-2026-07961, RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-07962, RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-08731, RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-08732] INFECTION WAS FOUND ONCE INCISION WAS MADE. THE PHYSICIAN WASHED THE SITE OUT WITH ANTIBIOTICS AND RECLOSED. IMPLANT CASE WAS ABANDONED.

    Patient problems: Unspecified Infection

    MAUDE report 25669454 ↗
  • InjuryJun 24, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT LEAD WAS VISUALLY CONFIRMED TO BE FRACTURED AND WAS STICKING OUT OF THE SKIN. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE PATIENT'S RIGHT SIDED SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Skin Erosion

    MAUDE report 25608454 ↗
  • InjuryJun 24, 2026Fracture

    IT WAS REPORTED THAT THE PATIENT'S RIGHT LEAD WAS VISUALLY CONFIRMED TO BE FRACTURED AND WAS STICKING OUT OF THE SKIN. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE PATIENT'S RIGHT SIDED SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Movement Disorder

    MAUDE report 25608451 ↗
  • InjuryMay 4, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT EXPERIENCED AN OPEN WOUND AT THE BURR HOLE COVER. SURGICAL INTERVENTION TOOK PLACE WHEREIN THE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Erosion

    MAUDE report 25071430 ↗
  • InjuryApr 17, 2026Use of Device Problem

    IT WAS REPORTED THAT THE PATIENT'S LEAD WAS EXPLANTED AND REPLACED DUE TO POOR INITIAL PLACEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24906259 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT EXPERIENCED AN INFECTION AT THE LEAD SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED ON (B)(6) 2026 TO ADDRESS THE ISSUE.

    Patient problems: Post Operative Wound Infection

    MAUDE report 24867115 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT EXPERIENCED AN INFECTION AT THE LEAD SITE. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED ON (B)(6) 2026 TO ADDRESS THE ISSUE.

    Patient problems: Post Operative Wound Infection

    MAUDE report 24867114 ↗
  • InjuryMar 31, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED RETURNING SYMPTOMS. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEAD WAS REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Movement Disorder

    MAUDE report 24746265 ↗
  • InjuryMar 25, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED SPEECH ISSUES FOLLOWING IMPLANT OF THE DBS SYSTEM. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEADS WERE EXPLANTED AND REPLACED. IT IS UNKNOWN WHICH LEAD WAS AT FAULT.

    Patient problems: Speech Disorder

    MAUDE report 24692128 ↗
  • InjuryMar 16, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENTS RIGHT LEAD ERODED THROUGH THE SKIN. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE ENTIRE SYSTEM WAS EXPLANTED TO ADDRESS THE ISSUE.

    Patient problems: Erosion

    MAUDE report 24612898 ↗
  • InjuryMar 16, 2026High impedance

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED INEFFECTIVE THERAPY DUE TO HIGH IMPEDANCES ON THE RIGHT LEAD. SURGICAL INTERVENTION WAS UNDERTAKEN WHEREIN THE LEAD WAS EXPLANTED AND REPLACED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24611160 ↗
  • InjuryMar 13, 2026Malposition of Device

    ADDITIONAL INFORMATION RECEIVED INDICATES THAT EFFECTIVE THERAPY WAS CONFIRMED DURING POST OP PROGRAMMING.

    Patient problems: Movement Disorder

    MAUDE report 24594224 ↗
  • InjuryMar 13, 2026Malposition of Device

    ADDITIONAL INFORMATION RECEIVED INDICATES THAT EFFECTIVE THERAPY WAS CONFIRMED DURING POST OP PROGRAMMING.

    Patient problems: Movement Disorder

    MAUDE report 24594204 ↗
  • InjuryMar 11, 2026Malposition of Device

    ADDITIONAL INFORMATION INDICATES REPLACEMENT LEAD WAS IMPLANTED ON (B)(6) 2026.

    Patient problems: Movement Disorder

    MAUDE report 24570175 ↗
  • InjuryMar 5, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT EXPERIENCED AN INFECTION AT THE RIGHT LEAD AND BURR HOLE CAP SITE AND WAS HOSPITALIZED. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE RIGHT LEAD AND BURR HOLE CAP WAS EXPLANTED TO ADDRESS THE ISSUE. THE PATIENT WAS ADMINISTERED IV ANTIBIOTICS TO ADDRESS THE ISSUE.

    Patient problems: Post Operative Wound Infection; Wound Dehiscence

    MAUDE report 24519982 ↗
  • InjuryFeb 18, 2026High impedance

    IT WAS REPORTED THE PATIENT UNDERWENT EXPLORATORY SURGERY ON 26JAN2026 WHERE IT WAS FOUND THE PATIENT'S LEAD AND EXTENSION EXHIBITED HIGH IMPEDANCES. AS A RESULT, SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2026 WHEREIN THE LEAD AND EXTENSION WERE EXPLANTED AND REPLACED TO ADDRESS THE ISSUE.

    Patient problems: Inadequate Pain Relief

    MAUDE report 24381504 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140009. One submission often covers many models, so check the product and lot details in each record.

  • Z-0459-2026 · Oct 6, 2025 · Open, Classified
    Liberta RC DBS IPG (Implantable Pulse Generator), Model Number 62400
  • Z-2696-2017 · Jun 2, 2017 · Terminated
    Infinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
  • Z-2693-2017 · Jun 2, 2017 · Terminated
    Infinity 5, Implantable Pulse Generator, REF 6660, Rx only, STERILE EO, Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions.
  • Z-2695-2017 · Jun 2, 2017 · Terminated
    Infinity 7, Implantable Pulse Generator, REF 6662, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I
  • Z-2694-2017 · Jun 2, 2017 · Terminated
    Infinity 5, Implantable Pulse Generator, REF 6661, Rx only, STERILE EO Product Usage: The Infinity IPG is an electronic device designed to be connected to one or two extensions. I

Other STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR devices

All STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR devices →