Evera™ S DR DDBC3D1 (UDI 00643169929159)Recalls & reported problems
Medtronic, Inc. · ICD-DR DDBC3D1 EVERA S US DF1 · Class 3 · Implant
All models of Evera S DR (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 9
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00643169929159), received between Apr 10, 2018 and May 20, 2025 — classified as 9 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Device Sensing Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Sensing Problem6
- Inappropriate/Inadequate Shock/Stimulation5
- Pacing Problem4
- Appropriate Device Problem Term/Code Not Available2
- Adverse Event Without Identified Device or Use Problem1
- Device Operates Differently Than Expected1
- Electrical /Electronic Property Problem1
- Electromagnetic Interference1
Patient problems
- Shock from Patient Lead(s)5
- No Clinical Signs, Symptoms or Conditions4
- Bradycardia2
- Cardiac Arrest1
- Chest Pain1
- Fall1
- Impaired Healing1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 20, 2025Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD HAD HIGH THRESHOLDS AND NO CAPTURE. ADDITIONALLY, TRUE ATRIAL EVENTS WERE REPORTED AS IN THE REFRACTORY PERIOD, RESULTING IN A PACING ISSUE. THE LEAD AND DEVICE REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22054070 ↗InjuryJun 3, 2024Electromagnetic Interference; Inappropriate/Inadequate Shock/Stimulation; Electromagnetic Interference; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED BY THE PATIENT THAT THEY WERE INAPPROPRIATELY SHOCKED BY THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WHEN THEY WERE NEAR AN "IMPROPERLY WIRED POOL PUMP MOTOR AND THERE WAS ELECTRICAL INTERFERENCE". THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 19453859 ↗InjuryMay 7, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT WITH STROKE SYMPTOMS AND A HEART RATE IN THE 40 BEATS PER MINUTE (BPM) RANGE. THE DEVICE WAS PROGRAMMED WITH LOWER RATE AT 60 BPM. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Stroke/CVA
MAUDE report 19259987 ↗InjuryOct 30, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) APPEARED TO TAKE A LONGER AMOUNT OF TIME TO INITIATE THERAPY. THE ICD WAS REPROGRAMMED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18035896 ↗MalfunctionSep 20, 2023Pacing Problem; Device Sensing Problem; Pacing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DETECTED VENTRICULAR TACHYCARDIA (VT) AND PROVIDED ANTI-TACHYCARDIA PACING. IT WAS SUSPECTED THAT THE RHYTHM WAS ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE, MEANING THE PACING WAS INAPPROPRIATE. THE DEVICE REMAINS IN USE AND APPEARS TO BE CURRENTLY FUNCTIONING AS PROGRAMMED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17781068 ↗InjuryMay 7, 2022Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DELIVERED MULTIPLE INAPPROPRIATE THERAPIES AND MAY HAVE CONTRIBUTED TO A FALL. THE PATIENT WAS ADMITTED TO THE HOSPITAL AND THE ICD WAS "TURNED OFF" ACCORDING TO THE CALLER. THE ICD REMAINS IMPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Fall; Shock from Patient Lead(s)
MAUDE report 14317803 ↗MalfunctionOct 29, 2021Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT AN ALERT WAS TRIGGERED FOR PACING IMPEDANCE VARIATION AND SHORT V-V INTERVALS. THE NEXT DAY ANOTHER ALERT WAS TRIGGERED FOR HIGH PACING IMPEDANCE. OVERSENSING/NOISE WAS INDICATED IN STORED EPISODES. THE RIGHT VENTRICULAR (RV) LEAD WAS SUSPECTED TO BE FRACTURED AND REPROGRAMMING WAS PERFORMED TO TURN OFF DETECTIONS AND THE PATIENT WAS FITTED WITH A LIFEVEST. HIGH ATRIAL THRESHOLD AND OVERSENSING WAS NOTED. THE RIGHT ATRIAL (RA) LEAD REMAINS IN USE. THE LONGEVITY ON THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS NOTED TO HAVE DROPPED SIGNIFICANTLY OVER A SHORT PERIOD OF TIME. THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12720368 ↗MalfunctionMay 20, 2021Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT HAD A HEART RATE LOWER THAN THE PROGRAMMED LOWER LIMIT OF THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). ADDITIONALLY, UNDERSENSING ON THE RIGHT ATRIAL (RA) LEAD IS OBSERVED, POSSIBLY TRIGGERING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) DEVICE CLASSIFIED TERMINATION OF AT/AF EVENTS. THE ICD AND RA LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 11855709 ↗InjuryMay 7, 2021Failure to Deliver Shock/Stimulation; Device Sensing Problem; Failure to Deliver Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT WAS BROUGHT INTO THE EMERGENCY ROOM DUE TO CARDIAC ARREST. THE NEXT DAY, THE PATIENT WAS UNRESPONSIVE AND INTUBATED. INTERROGATION DETERMINED THAT THE PATIENT APPEARED TO BE IN VENTRICULAR TACHYCARDIA (VT) APPROXIMATELY EIGHT MINUTES, OF WHICH THERAPY WAS WITHHELD DUE TO THE ONSET DISCRIMINATOR. THE DEVICE WAS REPROGRAMMED TO TURN THE DISCRIMINATOR OFF. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Cardiac Arrest
MAUDE report 11790387 ↗InjuryDec 17, 2020Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DETECTED FAST VENTRICULAR TACHYCARDIA (FVT) AND DELIVERED A SHOCK. IT WAS SUSPECTED THAT THE SHOCK WAS DUE TO ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE (RVR). THE DEVICE REMAINS IN USE AND CURRENTLY APPEARS TO BE FUNCTIONING AS PROGRAMMED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 11029122 ↗InjuryJul 22, 2020Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT CAME TO THE EMERGENCY DEPARTMENT SUSPECTING THEY HAD RECEIVED A SHOCK. A REMOTE MONITORING TRANSMISSION SHOWED NO THERAPY HAD BEEN PROVIDED. HOWEVER, REVIEW OF STORED EPISODES INDICATED THE PATIENT HAD BEEN TREATED FOR VENTRICULAR FIBRILLATION BUT THE RHYTHM WAS SUSPECTED TO BE ATRIAL FIBRILLATION WITH RAPID VENTRICULAR RESPONSE. THE SHOCK WAS SUSPECTED TO BE INAPPROPRIATE. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 10307269 ↗MalfunctionNov 26, 2019Device Sensing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD DETECTED A VENTRICULAR TACHYCARDIA NON-SUSTAINED (VTNS) EPISODE AS A SUPRA VENTRICULAR TACHYCARDIA (SVT) EPISODE. FOLLOW UP WAS CONDUCTED AND NO FURTHER INFORMATION WAS ASCERTAINED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9378790 ↗InjuryApr 10, 2019Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED INAPPROPRIATE THERAPY FOR AN ATRIAL ARRHYTHMIA WITH RAPID VENTRICULAR RESPONSE THAT WAS DETECTED AS FAST SUPRA VENTRICULAR TACHYCARDIA (SVT). THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 8499522 ↗MalfunctionMar 7, 2019Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT PRIOR TO USE, THE IMPLANTABLE CARDIOVERTER DEFIBRILLATORS (ICD) EXHIBITED GRAY BAR ON INTERROGATION, INDICATING LOW TELEMETRY VOLTAGE. THE ICDS WERE NOT USED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8401772 ↗InjuryAug 7, 2018Entrapment of Device; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available
THE PATIENT REPORTED DEVELOPING AN INFECTION AND THAT THE DEVICE IS "STUCK TO THE SKIN AND DOES NOT MOVE." A REQUEST FOR ADDITIONAL INFORMATION WAS MADE BUT NOT RECEIVED. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR REMAINS IN USE AND NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Impaired Healing; Skin Erosion; Unspecified Infection
MAUDE report 7759994 ↗MalfunctionApr 10, 2018Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED BY THE PATIENT THAT THEIR CELLPHONE WAS AFFECTING THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD). PATIENT REPORTED EXPERIENCING CHEST PAIN UNTIL THE PHONE WAS MOVED AWAY FROM THE DEVICE. FOLLOW-UP WITH THE PATIENT¿S PHYSICIAN FOR FURTHER INFORMATION WAS ATTEMPTED AND MULTIPLE ATTEMPTS WERE UNSUCCESSFUL. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Chest Pain
MAUDE report 7418385 ↗
Other Implantable cardioverter defibrillator (non-CRT) devices
- EMBLEM™ MRI S-ICD A219Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 recalls36 deaths15,404 reports
- EMBLEM™ S-ICD A209Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 recall7 deaths6,788 reports
- EMBLEM™ S-ICD 3501Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)7 deaths4,936 reports
- Cobalt™ XT DR MRI SureScan™ DDPA2D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)3 recalls30 deaths1,603 reports
- ENDOTAK RELIANCE® G 0185Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,632 reports
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- Evera MRI™ XT DR SureScan® DDMB1D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)1 recall35 deaths1,426 reports
- Durata™ 7122St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)27 deaths1,377 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page