Evera S DRRecalls & reported problems across all models
Medtronic, Inc. · 8 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
3 open recalls name at least one Evera S DR model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 384
- Deaths reported
- 16
- Injuries reported
- 261
- Recalls
- 4
We found 8 models of Evera S DR by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 384 adverse event reports in the FDA's MAUDE database, received between Dec 9, 2016 and Sep 25, 2026 — classified as 16 deaths, 261 injuries, 107 malfunctions.
The most frequently reported specific device problem is “Inappropriate/Inadequate Shock/Stimulation” (98 reports).
4 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Evera S DR recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 10, 2023Open, Classified
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Z-1710-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- May 10, 2023Open, Classified
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Z-1709-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Feb 3, 2021Open, Classified
A small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation
Z-1305-2021 · Root cause: Device Design · names 4 of these models · Recall details · FDA record ↗
- Dec 19, 2017Terminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
Z-0587-2018 · Root cause: Process design · names 2 of these models · Recall details · FDA record ↗
Most reported Evera S DR problems
Device problems
- Inappropriate/Inadequate Shock/Stimulation98
- Device Sensing Problem93
- Adverse Event Without Identified Device or Use Problem85
- Premature Discharge of Battery36
- Pacing Problem35
- Over-Sensing33
- Electromagnetic Interference26
- Invalid Sensing21
Patient problems
- Shock from Patient Lead(s)100
- No Known Impact Or Consequence To Patient73
- No Clinical Signs, Symptoms or Conditions71
- Unspecified Infection60
- Sepsis11
- Death10
- Bradycardia9
- No Patient Involvement9
Evera S DR models
8 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Evera™ S DR DDBC3D1Implantable cardioverter defibrillator (non-CRT)2 recalls7 deaths143 reports
- Evera™ S DR DDBC3D4Implantable cardioverter defibrillator (non-CRT)2 recalls7 deaths123 reports
- Evera™ S DR DDBC3D1Implantable cardioverter defibrillator (non-CRT)1 recall1 death53 reports
- Evera™ S DR DDBC3D1Implantable cardioverter defibrillator (non-CRT)1 recall1 death25 reports
- Evera™ S DR DDBC3D4Implantable cardioverter defibrillator (non-CRT)1 recall13 reports
- Evera™ S DR DDBC3D4Implantable cardioverter defibrillator (non-CRT)1 recall7 reports
- Evera™ S DR DDBC3D1Implantable cardioverter defibrillator (non-CRT)16 reports
- Evera™ S DR DDBC3D4Implantable cardioverter defibrillator (non-CRT)4 reports
Recent Evera S DR adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem
Patient problems: Medical device site infection; Sepsis
MAUDE report 50903231 ↗InjurySep 14, 2026Device Sensing Problem
It was reported by the patient that their Implantable Cardioverter Defibrillator (ICD) "went off during the night" and they "felt funny" and was "coming in and out of it". The patient also stated that they "fainted and fell" and remembered "coming back" and "gasping for air". The ICD remains in use. No further patient complications have been reported as a result of this event.
Patient problems: Dyspnea; Fall; Shock from Patient Lead(s); Syncope/Fainting
MAUDE report 50757021 ↗InjuryJun 29, 2026Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED DUE TO SYNCOPE. A DEVICE INTERROGATION INDICATED THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) TREATED AN EPISODE DETECTED AS VENTRICULAR FIBRILLATION (VF) THAT MAY HAVE BEEN VENTRICULAR TACHYCARDIA (VT). THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING DURING THE EPISODE THOUGH THERE APPEARED TO BE NO IMPACT TO DEVICE FUNCTION. THE ICD AND RV LEAD REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; Syncope/Fainting
MAUDE report 25649637 ↗MalfunctionMay 20, 2026Electromagnetic Interference; Device Sensing Problem
IT WAS REPORTED THAT ELECTROMAGNETIC INTERFERENCE RESULTED IN THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) MISCLASSIFYING AN ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EPISODE. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25228120 ↗InjuryFeb 20, 2026Adverse Event Without Identified Device or Use Problem
IT WAS FURTHER REPORTED THAT THE INFECTION WAS IDENTIFIED AS METHICILLIN-RESISTANT STAPHYLOCOCCUS AUREUS AND SEPSIS WAS NOTED.
Patient problems: Drug Resistant Bacterial Infection; Sepsis; Unspecified Infection
MAUDE report 24401279 ↗InjuryFeb 3, 2026Electromagnetic Interference; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED BY THE PATIENT THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXPERIENCED ELECTROMAGNETIC INTERFERENCE FROM AN ITEM PURCHASED BY THE PATIENT AND THE PATIENT WAS SHOCKED SEVERAL TIMES. THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 24251156 ↗InjuryDec 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS FURTHER REPORTED THAT THE ICD WAS EXPLANTED.
Patient problems: Unspecified Infection
MAUDE report 23908361 ↗InjuryDec 16, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT TWO DAYS POST IMPLANT OF THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) A SET SCREW REPOSITIONING PROCEDURE WAS PERFORMED. THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23821169 ↗MalfunctionDec 16, 2025Electromagnetic Interference; Over-Sensing; Electromagnetic Interference; Over-Sensing
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXPERIENCED ELECTROMAGNETIC INTERFERENCE THAT RESULTED IN OVERSENSING ON THE RIGHT ATRIAL (RA) CHANNEL AND MISCLASSIFICATION OF EPISODES. THE ICD WAS EXPLANTED APPROXIMATELY A WEEK LATER DUE TO NORMAL BATTERY DEPLETION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23821077 ↗MalfunctionDec 11, 2025Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing; Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing
Approximately nine months later, EMI/noise was noted on stored episodes triggering episode misclassification.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23780021 ↗MalfunctionDec 10, 2025Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing; Signal Artifact/Noise; Electromagnetic Interference; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED STEADILY INCREASING PACING IMPEDANCE, AND TRENDS SHOWED A SIGNIFICANT JUMP IN IMPEDANCE VALUE FOLLOWED BY A NEW HIGH BASE LINE. IN ADDITION, THERE WERE HIGH RV CAPTURE THRESHOLDS. IT WAS FURTHER REPORTED THAT THE GRADUAL RISE IN IMPEDANCE AND THRESHOLDS WERE SUSPECTED TO BE RELATED TO POSSIBLE MINERALIZATION OF THE TIP ELECTRODE. IN ADDITION, ON THE DAY THAT THE PHYSICIAN WAS PLACING AN ATRIOVENTRICULAR FISTULA, THE RV LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) FOR HIGH RATE NON-SUSTAINED EPISODES AND AN ELEVATED SENSING INTEGRITY COUNTER (SIC). OVERSENSING AND NOISE/ ELECTROMAGNETIC INTERFERENCE (EMI) WERE ALSO OBSERVED DURING THE PROCEDURE. THE LIA, OVERSENSING, AND NOISE/EMI WERE ALL ATTRIBUTED TO THE CAUTERY USED AT THE TIME. THE LEAD AND DEVICE REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23769748 ↗MalfunctionNov 10, 2025Premature Discharge of Battery; Premature Discharge of Battery
THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS RETURNED TO THE MANUFACTURER AFTER BEING EXPLANTED DUE TO REACHING ELECTIVE REPLACEMENT INDICATOR (ERI) AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER¿S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23514613 ↗InjuryOct 15, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) DETECTED IN THE VENTRICULAR TACHYCARDIA (VT) ZONE.
Patient problems: Shock from Patient Lead(s)
MAUDE report 23290039 ↗InjuryOct 15, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPIES WERE DELIVERED FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE (RVR) DETECTED IN THE VENTRICULAR FIBRILLATION (VF) ZONE. THE DATA ALSO SHOWED FOLLOWING THE SECOND INAPPROPRIATE SHOCK, A TRUE VF ARRHYTHMIA WAS INDUCED AND WAS SUBSEQUENTLY TREATED WITH AN ADDITIONAL HIGH-VOLTAGE DEFIBRILLATION SHOCK. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s); Ventricular Fibrillation
MAUDE report 23289447 ↗InjuryOct 15, 2025Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED VIA INTERNAL REVIEW OF HISTORICAL DATA TRANSMITTED FROM THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) THAT INAPPROPRIATE HIGH-VOLTAGE THERAPY WAS DELIVERED FOR ATRIAL FIBRILLATION (AF) WITH RAPID VENTRICULAR RESPONSE (RVR) DETECTED IN THE VENTRICULAR TACHYCARDIA (VT) ZONE.
Patient problems: Shock from Patient Lead(s)
MAUDE report 23289442 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.