Visualase 9735561 (UDI 00643169933378)Recalls & reported problems
Medtronic Navigation, Inc. · VCLAS 9735561 15MM TIP/1.85MM CATHETER · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169933378), received between Aug 31, 2020 and Aug 31, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Insufficient Heating” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Mar 23, 2018Terminated
Medtronic has become aware that certain lots of the VCLAS 3MM, 10MM, and 15MM laser diffusing fibers were not tested at intended laser power levels, due to an issue related to calibration of the test equipment, which can lead to the potential for low laser energy output from fiber or fiber burnout at lower wattage.
Z-2501-2018 · Root cause: Equipment maintenance · FDA record ↗
Most reported problems
Device problems
- Insufficient Heating1
Patient problems
- No Patient Involvement1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2020Insufficient Heating; Insufficient Heating
MEDTRONIC RECEIVED INFORMATION THAT, WHILE TESTING WITH THE THERMAL THERAPY SYSTEM, THERE WAS AN ISSUE WITH THE DISPOSABLE COOLING CATHETER SYSTEM (CCS) KIT. IT WAS REPORTED THAT LESIONS FORMED NON-UNIFORMLY AND ABLATED TISSUE WAS ONE SIDED AS THE UNIT ABLATED IN A 180 DEGREE REGION INSTEAD OF 360 DEGREES AROUND THE LASER DIFFUSING FIBER (LDF). IT WAS REPORTED THAT TISSUE TESTING WAS BEING CONDUCTED WHERE UNDER SMALLER ABLATION TIMES, THE ABLATED TISSUE WAS ON ONE SIDE OF THE LDF AND CATHETER. IT WAS NOTED THAT SWITCHING LDF'S DID NOT RESTORE FUNCTIONALITY, LEADING TO THE COOLING CATHETER BEING SUSPECTED TO BE THE CAUSE OF THE ISSUE. IT WAS NOTED THAT A 15W LASER WITH 8ML/MIN FLOW AND 16 SECONDS OF EXPOSURE TIME WERE BEING USED. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.
Patient problems: No Patient Involvement
MAUDE report 10471849 ↗
Other Powered laser surgical instrument devices
- Moxy 0010-2400Boston Scientific Corporation · Powered Laser Surgical Instrument4,572 reports
- SOLTIVE TFL-PLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls450 reports
- Flexiva TracTip M0068403960Boston Scientific Corporation · Powered laser surgical instrument562 reports
- Moses™ AC-10030100Lumenis Ltd. · Powered Laser Surgical Instrument532 reports
- Flexiva M0068403910Boston Scientific Corporation · Powered laser surgical instrument376 reports
- Visualase 9735542Medtronic Navigation, Inc. · Powered laser surgical instrument1 recall2 deaths311 reports
- SOLTIVE TFL-SLSGyrus ACMI, LLC · Powered laser surgical instrument6 recalls33 reports
- NeuroBlate® System 20980Monteris Medical Inc · Powered laser surgical instrument7 deaths296 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page