Skip to content

Visualase 9735571 (UDI 00643169933415)Recalls & reported problems

Medtronic Navigation, Inc. · ACCESSORY KIT 9735571 VISUALASE NEURO · Class 1

All models of Visualase (25)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169933415), received between Oct 22, 2019 and Jul 21, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation2
  • Mechanical Problem2
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • Unspecified Tissue Injury1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 21, 2025Leak/Splash; Mechanical Problem; Leak/Splash; Mechanical Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A THERMAL THERAPY SYSTEM BEING USED OUTSIDE OF A PROCEDURE. WHILE WALKING THROUGH A DEMONSTRATION, IT WAS NOTICED THAT THERE WAS A SALINE LEAK/DRIP AT THE BLUE STRAIN RELIEF SHEATH WHILE IN THE MAGNETIC RESONANCE IMAGING (MRI) MACHINE. TOWELS WERE USED TO CONTAIN THE DRIPPAGE AND CLEAN IT UP WITH NO APPARENT DAMAGE CAUSED. AFTER THE DRY RUN, IT WAS NOTICED THAT THE LOCAL REPRESENTATIVE (REP) COULD JUST PULL THE OUTER CANULA OF THE INNER CANULA/LUMEN AS IF THE ADHESION SPOT INSIDE THE BLUE STRAIN RELIEF WAS NOT HOLDING. THE REP HAD ALSO NOTICED WHEN THEY SET UP THAT THE STIFFENING STYLET WAS SLIGHTLY DIFFICULT TO INSERT, BUT IT DID PASS WITH SOME FRICTION. IT WAS NOTED THAT THE ACCESSORY KIT WAS "WAY" PAST EXPIRATION DATE. IT HAD EXPIRED ON 2020-OCT-25. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22556422 ↗
  • MalfunctionDec 2, 2020Mechanical Problem; Mechanical Problem

    ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THE PATIENT WAS DISCHARGED WITHOUT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury

    MAUDE report 10935885 ↗
  • MalfunctionNov 14, 2019Material Fragmentation; Material Fragmentation

    MEDTRONIC RECEIVED INFORMATION THAT, WHILE IN A SOFT NEURO TISSUE ABLATION PROCEDURE, THE BONE ANCHOR BOLT WAS DAMAGED WHEN BEING REMOVED FROM THE ANATOMY. IT WAS REPORTED THAT THE WINGS ON THE BOLT BROKE AWAY FROM THE STEM OF THE UNIT. IT WAS REPORTED THAT THE BOLT WAS REMOVED WITH A TOOL IN THE OPERATING ROOM (OR). ADDITIONALLY, THE PATIENT WAS BEING BROUGHT BACK TO THE OR TO WAKE UP AND EXUBATE AS PART OF THE PROCEDURE WORKFLOW WHEN THE REPORTED ISSUE OCCURRED. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO REPORTED IMPACT ON PATIENT OUTCOME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9323503 ↗
  • MalfunctionOct 22, 2019Material Fragmentation; Material Fragmentation

    MEDTRONIC RECEIVED INFORMATION THAT, WHILE IN A SOFT NEURO TISSUE ABLATION, THE PLASTIC WINGS ON THE TITANIUM BONE ANCHOR DETACHED FROM THE METAL WHILE REMOVING THE BONE ANCHOR. IT WAS REPORTED THAT THE SITE WAS ABLE TO REMOVE THE REST OF THE BOLT WITH ANOTHER TOOL. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO REPORTED IMPACT ON PATIENT OUTCOME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9217935 ↗

FDA clearance and classification

The FDA regulates product code MDM (instrument, manual, surgical, general use) under 21 CFR 878.4800 as a Class I device. 21 CFR 878.4800 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Visualase 9735571

Has the Visualase 9735571 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00643169933415), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Visualase 9735571?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Oct 22, 2019 and Jul 21, 2025. In those reports, the most common specific device problems are “Material Fragmentation” (2), “Mechanical Problem” (2) and “Leak/Splash” (1); the most common patient problem is “Unspecified Tissue Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Visualase 9735571?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Visualase 9735571 FDA approved?

Devices of this type (product code MDM) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Medtronic Navigation, Inc.

Is the Visualase 9735571 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the Visualase 9735571, and is it still sold?

The FDA's device database lists Medtronic Navigation, Inc. as the company responsible for it (brand Visualase, model 9735571). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices

All INSTRUMENT, MANUAL, SURGICAL, GENERAL USE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page