Visia AF MRI™ S VR SureScan™ DVFC3D4 (UDI 00643169988644)Recalls & reported problems
Medtronic, Inc. · ICD DVFC3D4 VISIA AF S VR DF4 MRI US · Class 3 · Implant
All models of Visia AF MRI S VR SureScan (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169988644), received between Mar 8, 2019 and May 31, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Sensing Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Sensing Problem3
- Inappropriate/Inadequate Shock/Stimulation2
- Pacing Problem1
Patient problems
- Shock from Patient Lead(s)2
- Chest Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 31, 2023Pacing Problem; Device Sensing Problem; Pacing Problem; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT DUE TO CHEST PAIN. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS INTERROGATED AND SHOWED THAT THE PATIENT RECEIVED ANTI-TACHYCARDIA PACING (ATP) FOR A DEVICE-CLASSIFIED EPISODE IN THE VENTRICULAR FIBRILLATION (VF) ZONE. IT WAS ALSO NOTED THAT A DISCRIMINATOR DETERMINED THE RHYTHM TO ACTUALLY HAVE BEEN SUPRA VENTRICULAR TACHYCARDIA (SVT), HOWEVER THE DISCRIMINATOR WAS PROGRAMMED TO MONITOR ONLY AND THE ATP WAS NOT WITHHELD AS A RESULT. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Chest Pain
MAUDE report 17029959 ↗InjuryAug 2, 2022Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) DETECTED AN ATRIAL ARRHYTHMIA FOLLOWED BY RAPID VENTRICULAR RESPONSE AS FAST VENTRICULAR TACHYCARDIA AND DELIVERED MULTIPLE INAPPROPRIATE THERAPIES. THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 15147827 ↗InjuryMar 8, 2019Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED BY THE PATIENT THAT THEIR IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS ¿BEEPING¿. IN ADDITION, THE PATIENT REPORTED THEY RECEIVED A SHOCK AND SENT A WIRELESS TRANSMISSION THEIR PHYSICIAN. THE PATIENT REPORTED THE PHYSICIAN STATED THAT THE PATIENT WAS IN ATRIAL FIBRILLATION (AF), AND ¿IT SOMEHOW TRICKED MY DEVICE TO THINK MY HEART RATE WAS GOING VERY FAST.¿ THE ICD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 8404049 ↗
Other Implantable cardioverter defibrillator (non-CRT) devices
- EMBLEM™ MRI S-ICD A219Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 recalls36 deaths15,404 reports
- EMBLEM™ S-ICD A209Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 recall7 deaths6,788 reports
- EMBLEM™ S-ICD 3501Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)7 deaths4,936 reports
- Cobalt™ XT DR MRI SureScan™ DDPA2D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)3 recalls30 deaths1,603 reports
- ENDOTAK RELIANCE® G 0185Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,632 reports
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- Evera MRI™ XT DR SureScan® DDMB1D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)1 recall35 deaths1,426 reports
- Durata™ 7122St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)27 deaths1,377 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page