Skip to content

iCast Covered Stent System, 6mmX16mmX120cm 42616Recalls & reported problems

Atrium Medical Corporation · iCast Covered Stent System, 6mmX16mmX120cm · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
15
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00650862426168), received between Jul 20, 2024 and May 20, 2026 — classified as 1 injury, 14 malfunctions.

The most frequently reported specific device problem is “Expiration Date Error” (12 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (P120003); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 12 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202612
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 9, 2024Class IIOpen, Classified

    There is an error in the Unique Device Identifier (UDI) printed on the product's distribution label.

    Z-0024-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Expiration Date Error12
  • Inaccurate Information12
  • Activation, Positioning or Separation Problem1
  • Difficult or Delayed Positioning1
  • Positioning Problem1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2026Premature Separation

    IT WAS REPORTED THAT WHILE PULLING BACK THE ICAST STENT DEVICE INTO THE GORE VBX, THE STENT PARTIALLY DISLODGED FROM THE BALLOON. THE PHYSICIAN BALLOONED THE STENT UP, AND IT WAS SUCCESSFULLY IMPLANTED WITH NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25228758 ↗
  • MalfunctionNov 21, 2025Difficult or Delayed Positioning; Positioning Problem; Difficult or Delayed Positioning; Positioning Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23615107 ↗
  • InjuryJun 13, 2025Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22201234 ↗
  • MalfunctionJul 22, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    REPORT RECEIVED FROM DISTRIBUTOR STATED THAT THE PRODUCT LABELING WAS FOUND TO HAVE AN ERROR. THE PRODUCT DATE OF MANUFACTURE WAS LISTED IN THE DISTRIBUTOR UDI BARCODE LABEL INSTEAD OF THE EXPIRATION DATE OF THE PRODUCT. THE EXPIRATION DATE ON THE PRIMARY MANUFACTURER PRODUCT LABEL WAS CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19802132 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787634 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    REPORT RECEIVED FROM DISTRIBUTOR STATED THAT THE PRODUCT LABELING WAS FOUND TO HAVE AN ERROR. THE PRODUCT DATE OF MANUFACTURE WAS LISTED IN THE DISTRIBUTOR UDI BARCODE LABEL INSTEAD OF THE EXPIRATION DATE OF THE PRODUCT. THE EXPIRATION DATE ON THE PRIMARY MANUFACTURER PRODUCT LABEL WAS CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787633 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    REPORT RECEIVED FROM DISTRIBUTOR STATED THAT THE PRODUCT LABELING WAS FOUND TO HAVE AN ERROR. THE PRODUCT DATE OF MANUFACTURE WAS LISTED IN THE DISTRIBUTOR UDI BARCODE LABEL INSTEAD OF THE EXPIRATION DATE OF THE PRODUCT. THE EXPIRATION DATE ON THE PRIMARY MANUFACTURER PRODUCT LABEL WAS CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787632 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787631 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787630 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    REPORT RECEIVED FROM DISTRIBUTOR STATED THAT THE PRODUCT LABELING WAS FOUND TO HAVE AN ERROR. THE PRODUCT DATE OF MANUFACTURE WAS LISTED IN THE DISTRIBUTOR UDI BARCODE LABEL INSTEAD OF THE EXPIRATION DATE OF THE PRODUCT. THE EXPIRATION DATE ON THE PRIMARY MANUFACTURER PRODUCT LABEL WAS CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787629 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787628 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    REPORT RECEIVED FROM DISTRIBUTOR STATED THAT THE PRODUCT LABELING WAS FOUND TO HAVE AN ERROR. THE PRODUCT DATE OF MANUFACTURE WAS LISTED IN THE DISTRIBUTOR UDI BARCODE LABEL INSTEAD OF THE EXPIRATION DATE OF THE PRODUCT. THE EXPIRATION DATE ON THE PRIMARY MANUFACTURER PRODUCT LABEL WAS CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787627 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    N/A

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787626 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    REPORT RECEIVED FROM DISTRIBUTOR STATED THAT THE PRODUCT LABELING WAS FOUND TO HAVE AN ERROR. THE PRODUCT DATE OF MANUFACTURE WAS LISTED IN THE DISTRIBUTOR UDI BARCODE LABEL INSTEAD OF THE EXPIRATION DATE OF THE PRODUCT. THE EXPIRATION DATE ON THE PRIMARY MANUFACTURER PRODUCT LABEL WAS CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787625 ↗
  • MalfunctionJul 20, 2024Expiration Date Error; Inaccurate Information; Expiration Date Error; Inaccurate Information

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19787624 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P120003. One submission often covers many models, so check the product and lot details in each record.

  • Z-0291-2024 · Sep 21, 2023 · Open, Classified
    Atrium Advanta V12 Covered Stent System, 10mm x 59mm x 80cm (OUS)-Indicated for restoring and improving the patency of the iliac andrenal arteries. Renal approval includes 5mm, 6m

Other Iliac Covered Stent, Arterial devices

All Iliac Covered Stent, Arterial devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page