Skip to content

GORE VIABAHN VBX Balloon Expandable Endo BXB095902ARecalls & reported problems

W. L. Gore & Associates, Inc. · BX2 HEP REDUCED PROFILE BX2 9MMX59MM 7Fr 135CM Cath · Class 3 · Implant

All models of GORE VIABAHN VBX Balloon Expandable Endo (111)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
0
Injuries reported
6
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00733132658824), received between Dec 17, 2025 and Sep 29, 2026 — classified as 6 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (P160021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 1, 2024Class IIOpen, Classified

    Due to packaging defects, sterility assurance and heparin activity may be compromised.

    Z-2945-2024 · Root cause: Package design/selection · Recall details · FDA record ↗

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Complete Blockage1
  • Detachment of Device or Device Component1
  • Device Endoleak1
  • Device Stenosis1
  • Obstruction of Flow1
  • Partial Blockage1
  • Patient-Device Incompatibility1

Patient problems

  • Clinical Endoleak2
  • Embolism/Embolus1
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • Obstruction/Occlusion1
  • Reduced Blood Flow1
  • Stenosis1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Detachment of Device or Device Component

    Patient problems: Foreign Body In Patient

    MAUDE report 50937996 ↗
  • InjuryJun 11, 2026Device Endoleak

    THE FOLLOWING INFORMATION WAS REPORTED THROUGH THE AAA 24-01 TAMBE POST APPROVAL STUDY: ON (B)(6), 2026, THE PATIENT UNDERWENT ENDOVASCULAR PROCEDURE TO TREAT A THORACOABDOMINAL AORTIC ANEURYSM USING A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) DEVICE SYSTEM AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE). ON MAY 18, 2026, A TYPE 1C ENDOLEAK WAS IDENTIFIED IN THE SUPERIOR MESENTERIC ARTERY (SMA). ON (B)(6), 2026, AND A REINTERVENTION WAS PERFORMED EXTENDING THE SMA VBX DEVICE WITH A 7MM X 59MM VBX DEVICE.

    Patient problems: Clinical Endoleak

    MAUDE report 25479830 ↗
  • MalfunctionApr 17, 2026Device Dislodged or Dislocated

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2026, DURING AN ILIAC BRANCH ENDOPROSTHESIS (IBE) PROCEDURE A 12FR X 26CM GORE® DRYSEAL FLEX INTRODUCER SHEATH (DSF) WAS USED TO ADVANCE A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX STENT) TO THE LEFT HYPOGASTRIC ARTERY. AS REPORTED, THERE WAS A PRE-EXISTING ENDURANT STENT AT THE INTENDED TREATMENT SITE. IT WAS REPORTED THE VBX STENT HAD TO BE RETRACTED BECAUSE THE WIRE AND SHEATH PROLAPSED INTO THE EXISTING EVAR DEVICE TO THE POINT THAT VBX STENT PUSHABILITY WAS LOST. AS THE VBX STENT WAS RETRACTED BACK INTO THE DSF, THE VBX STENT DISLODGED FROM THE BALLOON IN THE TRUNK OF THE EVAR DEVICE. THE PHYSICIAN WAS ABLE TO SNARE THE CONSTRAINED VBX STENT AND REMOVE IT FROM THE PATIENT WITH NO ISSUES. A NEW VBX DEVICE WAS ADVANCED AND DEPLOYED. THE PROCEDURE ENDED WITH GOOD OUTCOME. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24918031 ↗
  • InjuryApr 3, 2026Obstruction of Flow

    THE FOLLOWING WAS REPORTED TO GORE VIA AAA 2401 TAMBE STUDY: ON (B)(6) 2025, A PATIENT UNDERWENT A THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) PROCEDURE TO TREAT A DEGENERATIVE ANEURYSM UTILIZING GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS, GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS (AORTIC EXTENDER CONFORMABLE), AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE). A VBX DEVICE WAS PLACED IN THE SUPERIOR MESENTERIC ARTERY, TWO IN THE RIGHT RENAL ARTERY, ONE IN THE CELIAC TRUNK, AND ONE IN THE LEFT RENAL ARTERY. A BOSTON SCIENTIFIC INNOVA¿ VASCULAR SELF-EXPANDING STENT SYSTEM WAS PLACED IN THE RIGHT RENAL ARTERY AS WELL. THERE WAS KNOWN HEAVY SCARRING/CALCIFICATION OF THE RIGHT GROIN PRIOR TO PROCEDURE WHICH WAS NEAR OCCLUSIVE (RIGHT FEMORAL ARTERY OCCLUSION) AND THE SUTURE BROKE DURING THE ACCESS CLOSURE AFTER DEVICE DEPLOYMENTS IN THE RIGHT GROIN REQUIRING THROMBECTOMY, ENDARTERECTOMY AND BOVINE PATCH WITH LARGE AMOUNT OF THROMBUS REMOVED. PHYSICIAN STATED THIS ADVERSE EVENT WAS RELATED TO DISEASE CONDITION. ON (B)(6) 2026, THE FOLLOW UP CTA SCAN SHOWED THERE WAS A SMALL LEFT SPLENIC INFARCT AND SMALL RIGHT RENAL INFARCT WHICH WERE RELATED TO TH

    Patient problems: Embolism/Embolus

    MAUDE report 24780150 ↗
  • InjuryMar 13, 2026Complete Blockage; Patient-Device Incompatibility

    THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2025, A PATIENT UNDERWENT A THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) PROCEDURE. DURING THIS PROCEDURE, GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) DEVICES WERE IMPLANTED AS BRANCH DEVICES IN THE VISCERAL ARTERIES. ON (B)(6) 2026, A REINTERVENTION WAS PERFORMED DUE TO STENOSIS AT THE DISTAL END OF THE VBX DEVICE IN THE SUPERIOR MESENTERIC ARTERY (SMA). A BARE METAL STENT (MEDTRONIC PROTEGE) WAS IMPLANTED. THE PHYSICIAN ATTRIBUTED THE STENOSIS TO VASCULITIS.

    Patient problems: Stenosis

    MAUDE report 24593566 ↗
  • InjuryMar 6, 2026Partial Blockage

    THE FOLLOWING WAS OBTAINED FROM AAA 24-01 STUDY: ON (B)(6) 2026, A STUDY PATIENT UNDERWENT A THORACOABDOMINAL AORTA ANEURYSM (TAMBE) PROCEDURE. CUT-DOWN ACCESS WAS FROM THE LEFT SIDE. DURING THIS PROCEDURE, GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) DEVICES WERE IMPLANTED IN THE VISCERAL ARTERIES AS BRANCH DEVICES. ON (B)(6) 2026, AS THERE WAS CONCERN OF POSSIBLE STENT THROMBOSIS, AN ANGIOGRAM WAS PERFORMED. ON (B)(6) 2026, DUE TO A ENDOLEAK TYPE 1C, A THROMBECTOMY AND STENT EXTENSION OF THE SUPERIOR MESENTERIC ARTERY WAS COMPLETED. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Obstruction/Occlusion

    MAUDE report 24533522 ↗
  • InjuryJan 9, 2026Device Stenosis

    ON (B)(6) 2025, THE PATIENT UNDERWENT A GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE) PROCEDURE WHERE CONFORMABLE GORE® TAG® THORACIC STENT GRAFT (CTAG) AND FOUR GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) DEVICES WERE USED DURING THE PROCEDURE. IT WAS REPORTED DURING THE CASE, THE PHYSICIAN WAS UNABLE TO WIRE AND CANNULATE THE CELIAC ARTERY DUE TO TORTUOUS ANATOMY. AFTER AN EXTENSIVE AMOUNT OF TIME, THE PHYSICIAN DECIDED TO WIRE INSTEAD AND PLACE A VBX DEVICE IN THE SUPERIOR MESENTERIC ARTERY, RIGHT RENAL, AND THE LEFT RENAL. THE CELIAC PORTAL WAS EXTENDED WITH A VBX DEVICE AND AN 14MM AMPLATZ PLUG WAS PLACED INSIDE OF THE VBX DEVICE. ITS BELIEVED DURING THE CASE, POSSIBLY DUE TO POOR IMAGING AND VISUALIZATION OF THE CELIAC ARTERY, THE CTAG MAY HAVE BEEN COVERING THE ORIGIN. POSTOPERATIVELY THE PATIENT WAS DOING WELL.

    Patient problems: Reduced Blood Flow

    MAUDE report 24030200 ↗
  • InjuryDec 17, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2025, PATIENT UNDERWENT AN ENDOVASCULAR PROCEDURE TO TREAT AN ANEURYSM UTILIZING GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS (TAMBE DEVICE) AND GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS (VBX DEVICE) DEVICES. ON (B)(6) 2025, PATIENT UNDERWENT A REINTERVENTION PROCEDURE DUE TO AN ACCESS SITE HEMATOMA IN THE LEFT ARM. DURING THE INITIAL PROCEDURE THERE WAS A SURGICAL CUT DOWN PERFORMED TO DELIVER THE VBX DEVICES TO THEIR TARGET LOCATIONS. IT IS UNKNOWN WHAT CAUSED THE HEMATOMA, HOWEVER, IT IS THOUGHT TO BE UNRELATED TO THE GORE GRAFTS. DURING THE PROCEDURE THERE WAS IMAGING THAT SHOWED AN INDISCERNIBLE LEAK (TYPE AND LOCATION UNKNOWN) BUT PHYSICIAN THOUGHT IT IN THE SUPERIOR MESENTERIC ARTERY (SMA), SO WHILE THEY HAD ACCESS DURING THE HEMATOMA PROCEDURE, PHYSICIAN DECIDED TO ELECTIVELY IMPLANT AN ADDITIONAL VBX DEVICE AND EXTENDED THE SMA PORTAL OF THE TAMBE DEVICE. THE PATIENT TOLERATED PROCEDURE.

    Patient problems: Clinical Endoleak

    MAUDE report 23827695 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P160021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0077-2022 · Sep 7, 2021 · Open, Classified
    Heparin, GORE VIABAHN VBX, BALLOON EXPANDABLE ENDOPROTHESIS, REF BXAL085902E, 8 MM X 59 MM 8 Fr, LARGE BALLOON EXPANDLBE, Nominal Stent Diameter 8 mm, Max Post-Dilated Stent Diame

Other Iliac covered stent, arterial devices

All Iliac covered stent, arterial devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page