iCAST 85418Recalls & reported problems
Atrium Medical Corporation · iCAST COVERED STENT, 9MMX38MMX120CM · Class 2 · Implant
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00650862854183), received between May 31, 2016 and Feb 28, 2022 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K050814); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Activation, Positioning or Separation Problem2
- Material Rupture2
- Deflation Problem1
- Entrapment of Device1
- Inflation Problem1
- Material Split, Cut or Torn1
- Premature Activation1
- Premature Separation1
Patient problems
- No Consequences Or Impact To Patient3
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 28, 2022Material Rupture
REPORT RECEIVED STATED DURING A PROCEDURE FOR AN ILLIAC OCCLUSION, WITH SHOCKWAVE USED PRIOR FOR PRE-DILATION. THE ICAST WAS INSERTED AFTERWARDS ALONG WITH ANOTHER BALLOON ON THE CONTRA SIDE. BOTH WERE EXPANDED TOGETHER. ICAST BALLOON RUPTURED BEFORE GETTING TO ATM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13619638 ↗MalfunctionOct 5, 2021Premature Activation; Premature Separation; Premature Activation; Activation, Positioning or Separation Problem; Premature Separation
MalfunctionAug 11, 2020Material Split, Cut or Torn
InjuryMay 18, 2017Inflation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED THAT THE PROCEDURE WAS A THORACIC AORTIC ANEURYSM VASCULAR REPAIR (TAVR). THE PHYSICIAN PLANNED TO STENT THE LEFT COMMON FEMORAL ARTERY FROM A RIGHT GROIN APPROACH USING A 25 CM 7FR BOSTON SCIENTIFIC SHEATH. THE STENT WAS PLACED AND ON INFLATION OF THE BALLOON BLOOD WAS SEEN IN THE INSUFFLATOR. ONLY THE DISTAL END OF THE STENT BALLOON DEPLOYED. WHEN REMOVING THE BALLOON THE STENT MOVED BACK TO THE LEVEL OF THE COMMON ILIAC ARTERY. AFTER MULTIPLE INFLATION ATTEMPTS THE BALLOON WAS REMOVED AND THE STENT WAS DEPLOYED IN THE LEFT ILIAC. AN ADDITIONAL STENT WAS THEN SUCCESSFULLY PLACED IN THE LEFT COMMON FEMORAL ARTERY. THE PATIENT IS DOING FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6575745 ↗InjuryJun 28, 2016Deflation Problem
IT WAS REPORTED THAT AFTER DEPLOYING THE STENT THE BALLOON DID NOT DEFLATE AS EXPECTED. THE BALLOON TOOK TOO LONG TO DEFLATE.
Patient problems: No Information
MAUDE report 5754837 ↗InjuryMay 31, 2016Entrapment of Device; Material Rupture
IT WAS REPORTED THAT AFTER DEPLOYMENT OF THE STENT IN THE ILIAC ARTERY, THE BALLOON GOT CAUGHT ON A CALCIUM DEPOSIT AND RUPTURED. NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5688922 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K050814. One submission often covers many models, so check the product and lot details in each record.
- Z-1077-2022 · Mar 3, 2022 · Open, ClassifiediCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 85404 85405 85406 85407 85408 85409 85410 85411 85412 85413 85414 85415 85416 8541
- Z-2761-2017 · May 25, 2017 · TerminatedAtrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S. only) Atrium Advanta V12 Balloon Expandable Covered Stent; 7mm x 22mm
Other Prosthesis, tracheal, expandable devices
- iCAST 85453Atrium Medical Corporation · Prosthesis, tracheal, expandable43 reports
- WALLSTENT™ BILIARY H965430500Boston Scientific Corporation · Prosthesis, tracheal, expandable13 reports
- iCAST 85419Atrium Medical Corporation · Prosthesis, tracheal, expandable13 reports
- iCAST 85445Atrium Medical Corporation · Prosthesis, tracheal, expandable12 reports
- iCAST 85416Atrium Medical Corporation · Prosthesis, tracheal, expandable12 reports
- iCAST 85455Atrium Medical Corporation · Prosthesis, tracheal, expandable11 reports
- WALLSTENT® Endoprosthesis H965404130Boston Scientific Corporation · Prosthesis, tracheal, expandable11 reports
- iCAST 85414Atrium Medical Corporation · Prosthesis, tracheal, expandable10 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.