Adverse event reports
8
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00650862854206), received between May 11, 2016 and May 5, 2021 — classified as 5 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Material Rupture” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K050814); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Material Rupture2
  • Break1
  • Burst Container or Vessel1
  • Hole In Material1
  • Insufficient Device Problem Information1
  • Leak/Splash1
  • Material Deformation1
  • Separation Problem1

Patient problems

  • No Information2
  • Hemorrhage/Blood Loss/Bleeding1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1
  • Obstruction/Occlusion1
  • Occlusion1
  • Stenosis1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2021Material Rupture; Material Rupture

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11781084 ↗
  • MalfunctionSep 3, 2020Stretched; Material Deformation

    THE PHYSICIAN WAS DOING A STENTING BILATERAL TO THE PROXIMAL COMMON ILIAC ARTERIES. REPORT RECEIVED STATED THAT WHERE THE STENT AND BALLOON ATTACHED TO THE SHAFT, IT APPEARED STRETCHED OUT UPON REMOVAL AND THERE WAS CONCERN THE BALLOON COULD HAVE DETACHED IN PATIENT. THE PHYSICIAN WAS ABLE TO REMOVE BALLOON WITH NO HARM TO PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10492219 ↗
  • InjuryNov 12, 2019Material Rupture; Separation Problem

    DURING A BILATERAL ILIAC ARTERY ANGIOPLASTY AND STENTING PROCEDURE, THE BALLOON RUPTURED AND THEN SEPARATED FROM THE CATHETER.

    Patient problems: Occlusion

    MAUDE report 9310189 ↗
  • InjuryAug 1, 2019Break

    IT WAS REPORTED THAT THE PHYSICIAN PLACED THE STENT IN THE RIGHT COMMON ILIAC ARTERY. AFTER SUCCESSFULLY DEPLOYING THE STENT, WHEN REMOVING THE CATHETER FROM THE PATIENT, THE BALLOON SHEARED OFF THE CATHETER AND REMAINED ON THE WIRE AND PART OF THE BALLOON WAS IN THE SHEATH. THE CATHETER WHEN OFF THE WIRE, DIDN'T HAVE THE BALLOON ON IT. ACCESS WAS GAINED ON THE CONTRALATERAL SIDE AND A SNARE WAS ATTEMPTED TO CAPTURE THE BALLOON WHICH WAS UNSUCCESSFUL. THE PHYSICIAN REMOVED THE 7 FR SHEATH AND THE BALLOON CAME OUT WITH THE SHEATH.

    Patient problems: Obstruction/Occlusion

    MAUDE report 8850357 ↗
  • InjurySep 26, 2017Insufficient Device Problem Information

    IT WAS REPORTED THAT DURING AN INTERVENTION ON A LEFT SUBCLAVIAN ARTERY 95% STENOSIS THE STENT DID NOT CROSS THE LESION.

    Patient problems: Stenosis

    MAUDE report 6894776 ↗
  • MalfunctionSep 6, 2017Burst Container or Vessel

    THE PHYSICIAN USED A STENT FOR A RIGHT COMMON ILIAC ARTERY TARGET IN A FENESTRATED ENDOVASCULAR ANEURYSM REPAIR. THE VASCULATURE WAS HEAVILY CALCIFIED. THE STENT DEPLOYED CORRECTLY BUT THE BALLOON BURST AT 10 ATM.

    Patient problems: No Information

    MAUDE report 6844413 ↗
  • InjuryMay 1, 2017Hole In Material

    IT WAS REPORTED THAT THE STENT BALLOON HAD A HOLE IN IT AND THE PHYSICIAN HAD TO USE ANOTHER BALLOON TO TOUCH IT UP.

    Patient problems: No Information

    MAUDE report 6534709 ↗
  • InjuryMay 11, 2016Leak/Splash

    AS PART OF THE (B)(6) CLINICAL STUDY BEING CONDUCTED BY (B)(6) THE FOLLOWING EVENT WAS REPORTED: ON TWELVE MONTH FOLLOW UP CT SCAN, A DISTAL TYPE I ENDOLEAK WAS NOTED INVOLVING THE STENT. THERE WAS LOSS OF THE DISTAL SEAL IN THE RIGHT INTERNAL ILIAC ARTERY. AN ADDITIONAL STENT WAS PLACED INSIDE THE ORIGINAL STENT IN THE RIGHT INTERNAL ILIAC ARTERY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 5647934 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K050814. One submission often covers many models, so check the product and lot details in each record.

  • Z-1077-2022 · Mar 3, 2022 · Open, Classified
    iCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 85404 85405 85406 85407 85408 85409 85410 85411 85412 85413 85414 85415 85416 8541
  • Z-2761-2017 · May 25, 2017 · Terminated
    Atrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S. only) Atrium Advanta V12 Balloon Expandable Covered Stent; 7mm x 22mm

Other Prosthesis, tracheal, expandable devices

All Prosthesis, tracheal, expandable devices →

Sources and updates