iCAST 85441Recalls & reported problems
Atrium Medical Corporation · iCAST COVERED STENT, 5MMX22MMX80CM · Class 2 · Implant
All models of iCAST (29)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00650862854411), received between May 31, 2016 and Sep 16, 2020 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Component Falling” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K050814); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Component Falling1
- Detachment Of Device Component1
- Device Dislodged or Dislocated1
- Failure to Advance1
- Hole In Material1
- Loose or Intermittent Connection1
Patient problems
- No Consequences Or Impact To Patient2
- No Information2
- No Patient Involvement1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2020Loose or Intermittent Connection; Failure to Advance
THE ICAST STENT THAT WOULD NOT CROSS THE LESION WAS NOT IMPLANTED. THE PROCEDURE WAS COMPLETED WITH ANOTHER STENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10544504 ↗MalfunctionOct 20, 2017Hole In Material
USED PRODUCT RETURNED WITH PIN HOLE IN BALLOON.
Patient problems: No Information
MAUDE report 6964297 ↗MalfunctionSep 6, 2017Detachment Of Device Component; Component Falling
MEDWATCH RECEIVED FROM FDA. THE STENT ON THE DELIVERY CATHETER FELL OFF OR SLIPPED OFF THE BALLOON AFTER THE DEVICE WAS OPENED ONTO THE STERILE FIELD. THIS OCCURRED PRIOR TO TRYING TO IMPLANT THE DEVICE IN THE PATIENT.
Patient problems: No Consequences Or Impact To Patient; No Patient Involvement
MAUDE report 6844419 ↗InjuryMay 31, 2016Device Dislodged or Dislocated
DURING AN INTERVENTION ON AN ILIAC STENOSIS THE STENT BECAME DISLODGED FROM THE BALLOON IN THE SHEATH.
Patient problems: No Information
MAUDE report 5688928 ↗
FDA clearance and classification
Cleared through Special 510(k) K050814 on May 6, 2005 (decision: Substantially Equivalent). Submitted by Atrium Medical Corp. as “Atrium Icast Covered Stent”. The FDA review took 36 days. FDA record ↗
The FDA regulates product code JCT (prosthesis, tracheal, expandable) under 21 CFR 878.3720 as a Class II device. 21 CFR 878.3720 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K050814. One submission often covers many models, so check the product and lot details in each record.
- Z-1077-2022 · Mar 3, 2022 · Open, ClassifiediCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 85404 85405 85406 85407 85408 85409 85410 85411 85412 85413 85414 85415 85416 85417 85418 85419 85420
- Z-2761-2017 · May 25, 2017 · TerminatedAtrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S. only) Atrium Advanta V12 Balloon Expandable Covered Stent; 7mm x 22mm
Common questions about the iCAST 85441
Has the iCAST 85441 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00650862854411), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the iCAST 85441?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between May 31, 2016 and Sep 16, 2020. In those reports, the most common specific device problems are “Component Falling” (1), “Detachment Of Device Component” (1) and “Device Dislodged or Dislocated” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the iCAST 85441?
None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the iCAST 85441 FDA approved?
It was cleared through Special 510(k) K050814 on May 6, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the iCAST 85441 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the iCAST 85441, and is it still sold?
The FDA's device database lists Atrium Medical Corporation as the company responsible for it (brand iCAST, model 85441). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Prosthesis, tracheal, expandable devices
- iCAST 85453Atrium Medical Corporation · Prosthesis, tracheal, expandable43 reports
- WALLSTENT™ BILIARY H965430500Boston Scientific Corporation · Prosthesis, tracheal, expandable13 reports
- iCAST 85419Atrium Medical Corporation · Prosthesis, tracheal, expandable13 reports
- iCAST 85416Atrium Medical Corporation · Prosthesis, tracheal, expandable12 reports
- iCAST 85445Atrium Medical Corporation · Prosthesis, tracheal, expandable12 reports
- iCAST 85455Atrium Medical Corporation · Prosthesis, tracheal, expandable11 reports
- WALLSTENT® Endoprosthesis H965404130Boston Scientific Corporation · Prosthesis, tracheal, expandable11 reports
- iCAST 85414Atrium Medical Corporation · Prosthesis, tracheal, expandable10 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page